Performance Characteristics of a New LSO PET/CT Scanner With Extended Axial Field-of-View and PSF Reconstruction
The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
A new combined lutetium oxyorthosilicate (LSO) PET/CT scanner with an extended axial field-of-view (FOV) of 21.8 cm has been developed (Biograph TruePoint PET/CT with TrueV; Siemens Molecular Imaging) and introduced into clinical practice. The scanner includes the recently announced point spread function (PSF) reconstruction algorithm. The PET components incorporate four rings of 48 detector blocks, 5.4 cm times 5.4 cm in cross-section. Each block comprises a 13 times 13 matrix of 4 times 4 times 20 mm3elements. Data are acquired with a 4.5 ns coincidence time window and an energy window of 425-650 keV. The physical performance of the new scanner has been evaluated according to the recently revised National Electrical Manufacturers Association (NEMA) NU 2-2007 standard and the results have been compared with a previous PET/CT design that incorporates three rings of block detectors with an axial coverage of 16.2 cm (Biograph TruePoint PET/CT; Siemens Molecular Imaging). In addition to the phantom measurements, patient Noise Equivalent Count Rates (NECRs) have been estimated for a range of patients with different body weights (42-154 kg). The average spatial resolution is the same for both scanners: 4.4 mm (FWHM) and 5.0 mm (FWHM) at 1 cm and 10 cm respectively from the center of the transverse FOV. The scatter fractions of the Biograph TruePoint and Biograph TruePoint TrueV are comparable at 32%. Compared to the three ring design, the system sensitivity and peak NECR with smoothed randoms correction (1R) increase by 82% and 73%, respectively. The increase in sensitivity from the extended axial coverage of the Biograph TruePoint PET/CT with TrueV should allow a decrease in either scan time or injected dose without compromising diagnostic image quality. The contrast improvement with the PSF reconstruction potentially offers enhanced detectability for small lesions.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Fully 3-D PET reconstruction with system matrix derived from point source measurements
Panin, V. Y., Kehren, F., Michel, C., & Casey, M. (2006) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2006.876171 · Cited by 536 · cited 3×
“[4] the incorporation of the PSF model into the reconstruction results in a uniform spatial resolution throughout the FOV.”
MethodResultInfluential
Cited by
Drzezga, A., Souvatzoglou, M., Eiber, M., Beer, A. J., Fürst, S., Martinez-Möller, A., Nekolla, S. G., Ziegler, S., Ganter, C., Rummeny, E. J., et al. (2012) · Journal of Nuclear Medicine · doi:10.2967/jnumed.111.098608 · Cited by 508
Physical and clinical performance of the mCT time-of-flight PET/CT scanner
Jakoby, B. W., Bercier, Y., Conti, M., Casey, M. E., Bendriem, B., & Townsend, D. W. (2011) · Physics in Medicine and Biology · doi:10.1088/0031-9155/56/8/004 · Cited by 492
Self-citationPerformance Characteristics of the Digital Biograph Vision PET/CT System
van Sluis, J., de Jong, J., Schaar, J., Noordzij, W., van Snick, P., Dierckx, R., Borra, R., Willemsen, A., & Boellaard, R. (2019) · Journal of Nuclear Medicine · doi:10.2967/jnumed.118.215418 · Cited by 484
“view (FOV) increased volume sensitivity (6).”
Prenosil, G. A., Sari, H., Fürstner, M., Afshar-Oromieh, A., Shi, K., Rominger, A., & Hentschel, M. (2022) · Journal of Nuclear Medicine · doi:10.2967/jnumed.121.261972 · Cited by 338
An Assessment of the Impact of Incorporating Time-of-Flight Information into Clinical PET/CT Imaging
Lois, C., Jakoby, B. W., Long, M. J., Hubner, K. F., Barker, D. W., Casey, M. E., Conti, M., Panin, V. Y., Kadrmas, D. J., & Townsend, D. W. (2010) · Journal of Nuclear Medicine · doi:10.2967/jnumed.109.068098 · Cited by 265
Self-citationPerformance Evaluation of the Vereos PET/CT System According to the NEMA NU2-2012 Standard
Rausch, I., Ruiz, A., Valverde-Pascual, I., Cal-González, J., Beyer, T., & Carrio, I. (2019) · Journal of Nuclear Medicine · doi:10.2967/jnumed.118.215541 · Cited by 168
“2-cm axial FOV mCT Flow are not available but are expected to be similar to those of extended-axial-FOV systems (25,35).”
BackgroundRausch, I., Cal-González, J., Dapra, D., Gallowitsch, H. J., Lind, P., Beyer, T., & Minear, G. (2015) · EJNMMI Physics · doi:10.1186/s40658-015-0132-1 · Cited by 133
“Plots of the trues, randoms, and scatter event rates as well as the NECR and the scatter fraction curves as a function of activity are shown in Fig.”
BackgroundMethodResultInfluentialAkamatsu, G., Mitsumoto, K., Taniguchi, T., Tsutsui, Y., Baba, S., & Sasaki, M. (2014) · European Journal of Radiology · doi:10.1016/j.ejrad.2013.09.030 · Cited by 96
Performance Characteristics of the Whole-Body Discovery IQ PET/CT System
Reynés-Llompart, G., Gámez-Cenzano, C., Romero-Zayas, I., Rodríguez-Bel, L., Vercher-Conejero, J. L., & Martí-Climent, J. M. (2017) · Journal of Nuclear Medicine · doi:10.2967/jnumed.116.185561 · Cited by 78
“An additional method to improve sensitivity is to extend the axial field of view (FOV), hence an 81% in sensitivity gain has been reported (11).”
MethodClinical evaluation of PET image reconstruction using a spatial resolution model
Andersen, F. L., Klausen, T. L., Loft, A., Beyer, T., & Holm, S. (2013) · European Journal of Radiology · doi:10.1016/j.ejrad.2012.11.015 · Cited by 62
Reynés‐Llompart, G., Gámez‐Cenzano, C., Vercher‐Conejero, J. L., Sabaté‐Llobera, A., Calvo‐Malvar, N., & Martí‐Climent, J. M. (2018) · Medical Physics · doi:10.1002/mp.12986 · Cited by 55
Martersteck, A., Ayala, I., Ohm, D. T., Spencer, C., Coventry, C., Weintraub, S., Bigio, E. H., Mesulam, M. M., Geula, C., & Rogalski, E. (2022) · Acta Neuropathologica Communications · doi:10.1186/s40478-022-01412-w · Cited by 9
“The approximate point spread function was set at 4.5mm3, based on prior studies of the Biograph TruePoint/TrueV [22, 26].”
Method
Metadata
- Authors
- Jakoby, B. W., Bercier, Y., Watson, C. C., Bendriem, B., & Townsend, D. W.
- Venue
- IEEE Transactions on Nuclear Science
- Published
- 2009-06
- Volume
- 56
- Issue
- 3
- Pages
- 633-639
- ISSN
- 0018-9499 (ISSN-L)
- OpenAlex
- W2122964209
- Cited by
- 110
- References
- 1
- FWCI
- 4.93
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.