The DOI matches its Crossref record; 26 of the 26 references it cites were found in a registry or an index.
Design, Implementation and Clinical Tests of a Wire-Based Robot for Neurorehabilitation
Open accessAbstract
This paper presents the development of and clinical tests on NeReBot (NEuroREhabilitation roBOT): a three degrees-of-freedom (DoF), wire-driven robot for poststroke upper-limb rehabilitation. Basically, the robot consists of a set of three wires independently driven by three electric motors. The wires are connected to the patient's upper limb by means of a splint and are supported by a transportable frame, located above the patient. By controlling wire length, rehabilitation treatment (based on the passive or active-assistive spatial motion of the limb) can be delivered over a wide working space. The arm trajectory is set by the therapist through a very simple teaching-by-showing procedure, enabling most common "hands on" therapy exercises to be reproduced by the robot. Compared to other rehabilitation robots, NeReBot offers the advantages of a low-cost mechanical structure, intrinsically safe treatment thanks to the use of wires, high acceptability by the patient, who does not feel constrained by an "industrial-like" robot, transportability (it can be easily placed aside a hospital bed and/or a wheelchair), and a good trade-off between low number of DoF and spatial performance. These features and the very encouraging results of the first clinical trials make the NeReBot a good candidate for adoption in the rehabilitation treatment of subacute stroke survivors. Clinical trials were performed with a 12-patient experimental group and a 12-patient control group. Resulted that the patients who received robotic therapy in addition to conventional therapy showed greater reductions in motor impairment (in terms of Medical Research Council score, the upper limb subsection of the Fugl-Meyer score, and the Motor Status Score) and improvements in functional abilities (as measured by the Functional Independence Measure and its motor component). No adverse effects occurred and the robotic approach was very well accepted. According to these results, the NeReBot therapy may efficaciously complement standard poststroke multidisciplinary rehabilitation and offer novel therapeutic strategies for neurological rehabilitation.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Biomechanics and motor control of human movement
(1990) · Choice Reviews Online · doi:10.5860/choice.28-2135 · Cited by 8,451
The post-stroke hemiplegic patient. 1. a method for evaluation of physical performance
Fugl-Meyer, A., Jääskö, L., Leyman, I., Olsson, S., & Steglind, S. (1975) · Journal of Rehabilitation Medicine · doi:10.2340/1650197771331 · Cited by 5,564
Robot-aided neurorehabilitation
Krebs, H. I., Hogan, N., Aisen, M. L., & Volpe, B. T. (1998) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.662623 · Cited by 1,425
Treatment-Induced Cortical Reorganization After Stroke in Humans
Liepert, J., Bauder, H., Miltner, W. H. R., Taub, E., & Weiller, C. (2000) · Stroke · doi:10.1161/01.str.31.6.1210 · Cited by 1,404
Lum, P. S., Burgar, C. G., Shor, P. C., Majmundar, M., & Van der Loos, M. (2002) · Archives of Physical Medicine and Rehabilitation · doi:10.1053/apmr.2001.33101 · Cited by 1,159 · cited 1×
“The integration of robotic therapy into current practice may increase therapists’ efficiency and effectiveness by alleviating labor-intensive aspects of physical rehabilitation [2], [3], i.”
BackgroundReliability of the Fugl-Meyer Assessment of Sensorimotor Recovery Following Cerebrovascular Accident
Duncan, P. W., Propst, M., & Nelson, S. G. (1983) · Physical Therapy · doi:10.1093/ptj/63.10.1606 · Cited by 962
Intensity of leg and arm training after primary middle-cerebral-artery stroke: a randomised trial
Kwakkel, G., Wagenaar, R. C., Twisk, J. W., Lankhorst, G. J., & Koetsier, J. C. (1999) · The Lancet · doi:10.1016/s0140-6736(98)09477-x · Cited by 938
Prange, G. B., Jannink, M. J. A., Groothuis-Oudshoorn, C. G. M., Hermens, H. J., & IJzerman, M. J. (2006) · The Journal of Rehabilitation Research and Development · doi:10.1682/jrrd.2005.04.0076 · Cited by 917
Bütefisch, C., Hummelsheim, H., Denzler, P., & Mauritz, K.-H. (1995) · Journal of the Neurological Sciences · doi:10.1016/0022-510x(95)00003-k · Cited by 840
Cited by
A survey on robotic devices for upper limb rehabilitation
Maciejasz, P., Eschweiler, J., Gerlach-Hahn, K., Jansen-Troy, A., & Leonhardt, S. (2014) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-11-3 · Cited by 1,210
Review of control strategies for robotic movement training after neurologic injury
Marchal-Crespo, L., & Reinkensmeyer, D. J. (2009) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-6-20 · Cited by 1,182
Robotic neurorehabilitation: a computational motor learning perspective
Huang, V. S., & Krakauer, J. W. (2009) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-6-5 · Cited by 425
Leeb, R., Perdikis, S., Tonin, L., Biasiucci, A., Tavella, M., Creatura, M., Molina, A., Al-Khodairy, A., Carlson, T., & Millán, J. D. R. (2013) · Artificial Intelligence in Medicine · doi:10.1016/j.artmed.2013.08.004 · Cited by 166
Metadata
- Authors
- Rosati, G., Gallina, P., & Masiero, S.
- Venue
- IEEE Transactions on Neural Systems and Rehabilitation Engineering
- Published
- 2007-12
- Volume
- 15
- Issue
- 4
- Pages
- 560-569
- ISSN
- 1534-4320 (ISSN-L)
- 1558-0210
- OpenAlex
- W2120557068
- Cited by
- 267
- References
- 26
- FWCI
- 8.34
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Design, Implementation and Clinical Tests of a Wire-Based Robot for Neurorehabilitation. https://app.cochat.ai/papers/10.1109/tnsre.2007.908560
- The paper
- Rosati, G., Gallina, P., & Masiero, S. (2007). Design, Implementation and Clinical Tests of a Wire-Based Robot for Neurorehabilitation. IEEE Transactions on Neural Systems and Rehabilitation Engineering. https://doi.org/10.1109/tnsre.2007.908560
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.