Fatalities Due to Failure of Continuous Subcutaneous Insulin Infusion Devices: A Report of Six Cases
Abstract
Diabetes mellitus type 1 and type 2 are diseases characterized by impaired regulation of blood glucose due to decreased insulin production and insulin resistance, respectively. Management of diabetes mellitus often requires injection of exogenous insulin. Continuous subcutaneous insulin infusion (CSII or insulin pump) is a diabetes treatment modality utilizing a device to aid in regulation of glycemic control. Malfunctions in device components can have rare fatal consequences. Described in this report are six fatalities due to one such malfunction, the failure of plastic cannulas of CSII devices to penetrate the skin and deliver insulin, resulting in fatal diabetic ketoacidosis (DKA). The cases derive from four different death investigation systems. For each case, scene and autopsy findings are presented, as well as selected toxicology and histology findings. These cases illustrate the importance of careful examination of CSII devices in death investigations and introduce a discussion on discrepant manner of death classifications.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Without recognition and replacement, IIS failures lead to inadequate insulin delivery, hyperglycemia, and increased risk for diabetes-related ketoacidosis (DKA).
“Without recognition and replacement, IIS failures lead to inadequate insulin delivery, hyperglycemia, and increased risk for diabetes-related ketoacidosis (DKA).2,5,8,9”
From Frequency and Detection of Insulin Infusion Site Failure in the Type 1 Diabetes Exchange Online Community (2023) · cites this paper as [8]
Infusion failures are the leading cause of hospitalizations for diabetes-related ketoacidosis among people with diabetes who use CSII therapy, in some cases resulting in fatal outcomes.
“Notably, infusion failures are the leading cause of hospitalizations for diabetes-related ketoacidosis (DKA) hospitalization among PWD who use CSII therapy, in some cases resulting in fatal outcomes.8,9”
From Development of a Real-time Force-based Algorithm for Infusion Failure Detection (2024) · cites this paper as [9]
Cites
Nonmetabolic Complications of Continuous Subcutaneous Insulin Infusion: A Patient Survey
Pickup, J. C., Yemane, N., Brackenridge, A., & Pender, S. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0192 · Cited by 123 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Hughes, M. S., Douvas, J. L., Layfield-Bryan, M., Blanco, L. E., Gray, J. C., Zapotoczny, G., Espinoza, J., Wilcox, J. H., & Lal, R. A. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0005 · Cited by 16 · see citing works
Development of a Real-time Force-based Algorithm for Infusion Failure Detection
Blanco, L. E., Wilcox, J. H., Hughes, M. S., & Lal, R. A. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241247530 · Cited by 2 · see citing works
- Authors
- Ziegler, A., Williams, T., Yarid, N., Schultz, D. L., & Bundock, E. A.
- Affiliations
- Boston University (US)
- King County Medical Examiner's Office (US)
- Erie County Medical Examiner 501 Kensington Avenue Buffalo NY 14214
- District 12 Medical Examiner Office 2001 Siesta Drive #302 Sarasota FL 34231
- Office of Chief Medical Examiner (US)
- Corresponding author
- Ziegler, Andrew
- ORCID
- Yarid, Nicole
- Venue
- Journal of Forensic Sciences
- Publisher
- Wiley
- Published
- 2018-06-25
- Volume
- 64
- Issue
- 1
- Pages
- 275-280
- ISSN
- 0022-1198 (ISSN-L)
- 1556-4029
- OpenAlex
- W2810990973
- Cited by
- 14
- References
- 1
- Citations by year
- 2025: 1 · 2024: 3 · 2023: 3 · 2022: 2 · 2021: 4 · 2020: 1
- FWCI
- 0.99
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Journal of Forensic Sciences (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Autopsy Techniques and Outcomes
- Life Sciences › Pharmacology, Toxicology and Pharmaceutics › Toxicology › Forensic Toxicology and Drug Analysis
- Keywords
- insulin pump therapy, diabetic ketoacidosis, death investigation, forensic toxicology
- MeSH terms
- Adult, Aged, Blood Glucose, Diabetes Mellitus, Equipment Failure, Female, Glycosuria, Humans, Insulin Infusion Systems, Male, Middle Aged, Diabetic Ketoacidosis, Fatal Outcome, Infusions, Subcutaneous
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review