Accuracy of two continuous glucose monitoring systems: a head‐to‐head comparison under clinical research centre and daily life conditions
Open accessThe DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.
Abstract
Aims: To assess the accuracy and reliability of the two most widely used continuous glucose monitoring ( CGM ) systems. Methods: We studied the Dexcom® G4 Platinum ( DG4P ; Dexcom, San Diego, CA , USA ) and Medtronic Paradigm Veo Enlite system ( ENL ; Medtronic, Northridge, CA , USA ) CGM systems, in 24 patients with type 1 diabetes. The CGM systems were tested during 6‐day home use and a nested 6‐h clinical research centre ( CRC ) visit. During the CRC visit, frequent venous blood glucose samples were used as reference while patients received a meal with an increased insulin bolus to induce an aggravated postprandial glucose nadir. At home, patients performed at least six reference capillary blood measurements per day. A Wilcoxon signed‐rank test was performed using all data points ≥15 min apart. Results: The overall mean absolute relative difference ( MARD ) value [standard deviation (s.d.)] measured at the CRC was 13.6 (11.0)% for the DG4P and 16.6 (13.5)% for the ENL [p < 0.0002, confidence interval of difference ( CI Δ) 1.7–4.3%, n = 530]. The overall MARD assessed at home was 12.2 (12.0)% for the DG4P and 19.9 (20.5)% for the ENL (p < 0.0001, CI Δ = 5.8–8.7%, n = 839). During the CRC visit, the MARD in the hypoglycaemic range [≤3.9 mmol/l (70 mg/dl)], was 17.6 (12.2)% for the DG4P and 24.6 (18.8)% for the ENL (p = 0.005, CI Δ 3.1–10.7%, n = 117). Both sensors showed higher MARD values during hypoglycaemia than during euglycaemia [3.9–10 mmol/l (70–180 mg/dl)]: for the DG4P 17.6 versus 13.0% and for the ENL 24.6 versus 14.2%. Conclusions: During circumstances of intended use, including both a CRC and home phase, the ENL was noticeably less accurate than the DG4P sensor. Both sensors showed lower accuracy in the hypoglycaemic range. The DG4P was less affected by this negative effect of hypoglycaemia on sensor accuracy than was the ENL .
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1177/096228029900800204 · Cited by 9,619 · cited 2×
“Cumulative distribution plots show individual sensor performance (Figure 3A and B), while the Bland–Altman plot (Figure 1) provides information on variations in sensor accuracy over the range of measured glucose concentrations [5], indicating lower accuracy in the hypoglycaemic range for both CGM systems.”
BackgroundMethodMeasuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1191/096228099673819272 · Cited by 9,619
Cited by
Physical activity/exercise and diabetes: A position statement of the American Diabetes Association
Colberg, S. R., Sigal, R. J., Yardley, J. E., Riddell, M. C., Dunstan, D. W., Dempsey, P. C., Horton, E. S., Castorino, K., & Tate, D. F. (2016) · Diabetes Care · doi:10.2337/dc16-1728 · Cited by 2,991
Continuous Glucose Monitoring: A Review of Successes, Challenges, and Opportunities
Rodbard, D. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0417 · Cited by 737
Sensors for daily life: A review
Javaid, M., Haleem, A., Rab, S., Pratap Singh, R., & Suman, R. (2021) · Sensors International · doi:10.1016/j.sintl.2021.100121 · Cited by 430
van Beers, C. A. J., DeVries, J. H., Kleijer, S. J., Smits, M. M., Geelhoed-Duijvestijn, P. H., Kramer, M. H. H., Diamant, M., Snoek, F. J., & Serné, E. H. (2016) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(16)30193-0 · Cited by 386
Self-citationAssessing Sensor Accuracy for Non-Adjunct Use of Continuous Glucose Monitoring
Kovatchev, B. P., Patek, S. D., Ortiz, E. A., & Breton, M. D. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0272 · Cited by 234
Metadata
- Authors
- Kropff, J., Bruttomesso, D., Doll, W., Farret, A., Galasso, S., Luijf, Y. M., Mader, J. K., Place, J., Boscari, F., Pieber, T. R., Renard, E., & DeVries, J. H.
- Venue
- Diabetes, Obesity and Metabolism
- Publisher
- Wiley
- Published
- 2014-09-10
- Volume
- 17
- Issue
- 4
- Pages
- 343-349
- ISSN
- 1462-8902 (ISSN-L)
- 1463-1326
- OpenAlex
- W1996641990
- Cited by
- 132
- References
- 2
- FWCI
- 16.04
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Accuracy of two continuous glucose monitoring systems: a head‐to‐head comparison under clinical research centre and daily life conditions. https://app.cochat.ai/papers/10.1111/dom.12378
- The paper
- Kropff, J., Bruttomesso, D., Doll, W., Farret, A., Galasso, S., Luijf, Y. M., Mader, J. K., Place, J., Boscari, F., Pieber, T. R., Renard, E., & DeVries, J. H. (2015). Accuracy of two continuous glucose monitoring systems: a head‐to‐head comparison under clinical research centre and daily life conditions. Diabetes, Obesity and Metabolism. https://doi.org/10.1111/dom.12378
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.