Automatic blood pressure monitors A clinical evaluation of five models in adults
The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
Summary: Five automatic blood pressure monitors were evaluated by comparing their values with almost simultaneous blood pressure readings from radial artery catheters. A total of 583 comparisons were made on 48 patients. Statistical analysis was performed on the pooled comparisons of systolic and diastolic pressure for each monitor. Agreement between the pressure measurements was variable between both patients and machines. For systolic pressure, two machines (Copal and Sentron) had correlation coefficients of over 0.9 while for the other machines (Dinamap 845XT, Narco and Vitastat 9001S) the values lay between 0.7 and 0.8. The steepest regression lines were also found with the Copal (0.93) and Sentron (0.86) data, but the other machines had considerably flatter slopes (0.55 to 0.67). The intercepts on the y axes ranged from + 14.1 (Copa) to + 50.1 (Dinamap). In clinical use, popularity of the different machines rejected the degree to which the machine reproduced the behaviour of the direct pressure measurement.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Noninvasive automatic determination of mean arterial pressure
Ramsey, M. (1979) · Medical and Biological Engineering and Computing · doi:10.1007/bf02440948 · Cited by 287
Cited by
Blood pressure monitoring: Automated oscillometric devices
Ramsey, M. (1991) · Journal of Clinical Monitoring · doi:10.1007/bf01617900 · Cited by 155
Effect of four combined oral contraceptives on blood pressure in the pill-free interval
Nichols, M., Robinson, G., Bounds, W., Newman, B., & Guillebaud, J. (1993) · Contraception · doi:10.1016/0010-7824(93)90034-5 · Cited by 57
The Impact of Arm Position on the Measurement of Orthostatic Blood Pressure
Guss, D. A., Abdelnur, D., & Hemingway, T. J. (2008) · The Journal of Emergency Medicine · doi:10.1016/j.jemermed.2007.05.049 · Cited by 15
Metadata
- Authors
- JOHNSON, C. J. H., & KERR, J. H.
- Venue
- Anaesthesia
- Publisher
- Wiley
- Published
- 1985-05
- Volume
- 40
- Issue
- 5
- Pages
- 471-478
- ISSN
- 0003-2409 (ISSN-L)
- 1365-2044
- OpenAlex
- W2013685366
- Cited by
- 38
- References
- 1
- FWCI
- 7.18
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Automatic blood pressure monitors A clinical evaluation of five models in adults. https://app.cochat.ai/papers/10.1111/j.1365-2044.1985.tb10851.x
- The paper
- JOHNSON, C. J. H., & KERR, J. H. (1985). Automatic blood pressure monitors A clinical evaluation of five models in adults. Anaesthesia. https://doi.org/10.1111/j.1365-2044.1985.tb10851.x
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.