Blood pressure monitoring: Automated oscillometric devices
Abstract
An understanding of the principles of blood pressure measurement with automated oscillometric devices (e.g., Dinamap) allows a user to realize the purpose and level of accuracy for which the equipment is designed. Good technique cells for various actions: (1) using a cuff of proper size (too large a cuff will generally work acceptably, but too small a cuff will yield erroneously high readings); (2) squeezing all of the residual air out of the cuff before applying it to the arm or leg; (3) wrapping the cuff snugly around the arm or leg; (4) instructing the patient to refrain from talking or moving and letting nothing press against the cuff during measurement; (5) keeping the cuff and heart at the same horizontal level. (6) When instrument malfunction is suspected, first checking the status of the patient (not the status of the instrument) to be sure that a clinical emergency is not causing the suspected "malfunction"; (7) repeating a reading several times when it is suspicious and, if necessary, modifying the measurement conditions by using, for example, different limbs or different cuffs until the question is resolved; and (8) remembering that a leak in the cuff, hoses, or connection can cause an unexpected malfunction.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Automated BPMDs use proprietary algorithms to estimate BP by analysis of cuff pressure waveform signals, and this approach has remained mostly unchanged for decades.
“For automated BPMDs, these processes are undertaken electronically and based on proprietary algorithms designed to estimate BP by analysis of cuff pressure waveform signals detected because they are transmitted from the cuff into the tubing system (and onwards to the pressure transducer), and ultimately processed by microcomputer. This approach has remained mostly unchanged for decades [13, 44, 45”
From Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure (2022) · cites this paper as [13]
Cites
Arm position as a source of error in blood pressure measurement
Mariotti, G., Alli, C., Avanzini, F., Canciani, C., Di Tullio, M., Manzini, M., Salmoirago, E., Taioli, E., Zussino, A., & Radice, M. (1987) · Clinical Cardiology · doi:10.1002/clc.4960101016 · Cited by 34 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
ATS/ACCP statement on cardiopulmonary exercise testing
(2003) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.167.2.211 · Cited by 3,026 · see citing works
Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure
Sharman, J. E., Tan, I., Stergiou, G. S., Lombardi, C., Saladini, F., Butlin, M., Padwal, R., Asayama, K., Avolio, A., Brady, T. M., et al. (2022) · Journal of Human Hypertension · doi:10.1038/s41371-022-00693-x · Cited by 84 · see citing works
Optimism and Cardiovascular Health: Multi-Ethnic Study of Atherosclerosis (MESA)
Hernandez, R., Kershaw, K. N., Siddique, J., Boehm, J. K., Kubzansky, L. D., Diez-Roux, A., Ning, H., & Lloyd-Jones, D. M. (2015) · Health Behavior and Policy Review · doi:10.14485/hbpr.2.1.6 · Cited by 63 · see citing works
Garg, P. K., Jorgensen, N. W., McClelland, R. L., Leigh, J. A., Greenland, P., Blaha, M. J., Yoon, A. J., Wong, N. D., Yeboah, J., & Budoff, M. J. (2018) · Journal of Cardiovascular Computed Tomography · doi:10.1016/j.jcct.2018.10.001 · Cited by 26 · see citing works
- Authors
- Ramsey, M.
- Corresponding author
- Ramsey, Maynard
- Venue
- Journal of Clinical Monitoring
- Published
- 1991-01
- Volume
- 7
- Issue
- 1
- Pages
- 56-67
- ISSN
- 0748-1977 (ISSN-L)
- OpenAlex
- W2049364112
- Cited by
- 155
- References
- 1
- Citations by year
- 2026: 1 · 2024: 4 · 2023: 3 · 2022: 3 · 2021: 5 · 2020: 8 · 2019: 2 · 2018: 5 · 2017: 1 · 2016: 7 · 2015: 3 · 2014: 2 · 2013: 3 · 2012: 3
- FWCI
- 5.91
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.springer.com/tdm (tdm)
- Locations
- The Journal of Clinical Monitoring (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Keywords
- blood pressure measurement, cuff size, measurement methods, cuff leak, instrument malfunction, measurement accuracy, patient positioning
- MeSH terms
- Equipment Failure, Humans, Microcomputers, Oscillometry, Blood Pressure Monitors
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review