Student or educator? Sign up with your school email and get 6 months of Pro free.
Cited by 242

The DOI matches its Crossref record.

Electrocardiographically Documented Unnecessary, Spontaneous Shocks in 241 Patients With Implantable Cardioverter Defibrillators

GRIMM, WOLFRAM, FLORES, BELINDA F., MARCHLINSKI, FRANCIS E. · 1992 · Pacing and Clinical Electrophysiology

Abstract

The incidence and cause of electrocardiographically documented spontaneous implantuble Cardioverter defibrillator (ICD) discharges for a rhythm other than ventricular tachycardia (VT) or fibrillation (VF) (unnecessary shocks) were determined in 241 patients who underwent ICD implantation between March 1983 and November 1991. During follow‐up of 24 ± 20 months, 54 of 241 patients (22%) received a total of 132 unnecessary ICD shocks confirmed by Holler or telemetry monitoring or stored electrograms (Egs) from the ICD. The rhythm preceding these unnecessary ICD shocks was atrial fibrillation in 30 patients, sinus or supraventricular tachycardia (SVT) in 11 patients, antitachycardia pacing triggered by atrial fibrillation or SVT resulting in VT in 5 patients, nonsustained VT in 3 patients, and normal sinus or pacemaker rhythm in 10 patients. Unnecessary ICD discharges occurred most frequently during the first week after implantation or generator replacement (18 of 54 patients [33%]). Unnecessary ICD discharges could be documented more often by stored Egs in patients with devices with Eg storage capability (Ventritex Cadence(tm), 19 of 54 patients [35%]) than by Holter or telemetry monitoring in patients with devices without Eg storage capabilities (34 of 193 patients [18%], P < 0.01), despite a shorter mean follow‐up duration of 14 ± 9 months versus 26 ± 21 months, respectively. Only six of 54 patients (11%) in whom unnecessary ICD discharges occurred had recurrent unnecessary shocks during 22 ± 20 months of follow‐up after treatment directed at the cause of the first episode or device reprogramming to preclude non‐VT rhythm detection. In conclusion, unnecessary ICD shocks are a frequent complication of ICD therapy occurring in at least 22% of patients. The cause of these shocks is most frequently atrial fibrillation with a rapid ventricular response. The availability of Eg storage capabilities facilitates the diagnosis of the electrical event precipitating inappropriate ICD shocks. Diagnosis of the cause of unnecessary shocks allows for the institution of therapy that may reduce the risk for subsequent events.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1111/j.1540-8159.1992.tb02953.x

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 242

Metadata

Authors
GRIMM, W., FLORES, B. F., & MARCHLINSKI, F. E.
Venue
Pacing and Clinical Electrophysiology
Publisher
Wiley
Published
1992-11
Volume
15
Issue
11
Pages
1667-1673
ISSN
  • 0147-8389 (ISSN-L)
  • 1540-8159
OpenAlex
W2041007877
Cited by
242
References
0
FWCI
7.12
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Electrocardiographically Documented Unnecessary, Spontaneous Shocks in 241 Patients With Implantable Cardioverter Defibrillators. https://app.cochat.ai/papers/10.1111/j.1540-8159.1992.tb02953.x
The paper
GRIMM, W., FLORES, B. F., & MARCHLINSKI, F. E. (1992). Electrocardiographically Documented Unnecessary, Spontaneous Shocks in 241 Patients With Implantable Cardioverter Defibrillators. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/j.1540-8159.1992.tb02953.x

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.