Relationship Between Catheter Forces, Lesion Characteristics, “Popping,” and Char Formation: Experience with Robotic Navigation System

DI BIASE, LUIGI, NATALE, ANDREA, BARRETT, CONOR, TAN, CARMELA, ELAYI, CLAUDE S., CHING, CHI KEONG, WANG, PAUL, AL‐AHMAD, AMIN, ARRUDA, MAURICIO, BURKHARDT, J. DAVID, WISNOSKEY, BRIAN J., CHOWDHURY, PUNAM, DE MARCO, SHARI, ARMAGANIJAN, LUCIANA, LITWAK, KENNETH N., SCHWEIKERT, ROBERT A., CUMMINGS, JENNIFER E. · 2009 · Journal of Cardiovascular Electrophysiology

Abstract

Introduction: Popping, char and perforation are complications that can occur following catheter ablation. We measured the amount of grams (g) applied to the endocardium during ablation using a sensor incorporated in the long sheath of a robotic system. We evaluated the relationship between lesion formation, pressure, and the development of complications. Methods: Using a robotic navigation system, lesions were placed in the left atrium (LA) at six settings, using a constant duration (40 seconds) and flow rate of either 17 cc/min or 30 cc/min with an open irrigated catheter (OIC). Evidence of complications was noted and lesion location recorded for later analysis at necropsy. Results: Lesions using 30 Watts (W) were more likely to be transmural at higher (>40 g) than lower (<30 g) pressures (75% vs 25%, P < 0.001). Significantly higher number of lesions using >40 g of pressure demonstrated “popping” and crater formation as compared with lesions with 20–30 g of pressure (41% vs 15%, P = 0.008). A majority of lesions placed using higher power (45 W) with higher pressures (>40 g) were associated with char and crater formation (66.7%). No lesions using 10 g of pressure were transmural, regardless of the power. Lesions placed with a power setting less than 35 W were more likely to result in “relative” sparing of the endocardial surface than lesions at a power setting higher than 35 W (62% vs 33.3%, P = 0.02) regardless of the pressure. Conclusions: When using an OIC, lower power settings (≤35 W) and lower/medium contact pressure were more likely to show a “relative” spared endocardial surface. Overall, contact pressure between 20 g and 30 g and a power setting of 40 W appeared to achieve transmurality by preserving safety.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1111/j.1540-8167.2008.01355.x

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 163

Metadata

Authors
DI BIASE, L., NATALE, A., BARRETT, C., TAN, C., ELAYI, C. S., CHING, C. K., WANG, P., AL‐AHMAD, A., ARRUDA, M., BURKHARDT, J. D., WISNOSKEY, B. J., CHOWDHURY, P., DE MARCO, S., ARMAGANIJAN, L., LITWAK, K. N., SCHWEIKERT, R. A., & CUMMINGS, J. E.
Venue
Journal of Cardiovascular Electrophysiology
Publisher
Wiley
Published
2009-03-23
Volume
20
Issue
4
Pages
436-440
ISSN
  • 1045-3873 (ISSN-L)
  • 1540-8167
OpenAlex
W2022157658
Cited by
163
References
0
FWCI
4.62
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Relationship Between Catheter Forces, Lesion Characteristics, “Popping,” and Char Formation: Experience with Robotic Navigation System. https://app.cochat.ai/papers/10.1111/j.1540-8167.2008.01355.x
The paper
DI BIASE, L., NATALE, A., BARRETT, C., TAN, C., ELAYI, C. S., CHING, C. K., WANG, P., AL‐AHMAD, A., ARRUDA, M., BURKHARDT, J. D., WISNOSKEY, B. J., CHOWDHURY, P., DE MARCO, S., ARMAGANIJAN, L., LITWAK, K. N., SCHWEIKERT, R. A., & CUMMINGS, J. E. (2009). Relationship Between Catheter Forces, Lesion Characteristics, “Popping,” and Char Formation: Experience with Robotic Navigation System. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/j.1540-8167.2008.01355.x

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.