Evaluation and performance characteristics of the coagulation system: ACL TOP analyzer – HemosIL reagents
The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.
Abstract
Summary: ACL TOP is a fully automated coagulation analyzer, designed for simultaneous measurement of routine and special coagulation parameters. We evaluated analytical and technical performance characteristics of the coagulation system composed of the ACL TOP analyzer and HemosIL reagent group for the determination of routine clotting (PT, APTT, fibrinogen, FVII, and FVIII), chromogenic (protein C) and immunological assays (FXIII antigen). Within run and between run CVs ranged from 0.9% to 7.7% and from 2.0% to 14.8% respectively. The obtained CVs for imprecision of calibration curves were <5% for PT and <7% for fibrinogen. The method comparison study showed good correlation between results obtained on the ACL TOP and BCS/BCT analyzers, with correlation co‐efficients ranging from 0.709 to 0.955, but with significantly different results for PT INR, APTT, fibrinogen and protein C, and wide dispersion of differences observed in difference plots for most assays. Despite good correlation and agreement for FVIII, problems in measuring FVIII<10% were encountered. The effective throughput for the ACL TOP and BCS was 151 and 212 PT/APTT/fibrinogen tests per hour, respectively. Although the ACL TOP is designed to run multiple assays on a large number of samples, software limitations make the instrument suitable rather for mid‐sized laboratories.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Current databases on biological variation: pros, cons and progress
C. Ricós, V. A. (1999) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365519950185229 · Cited by 987 · cited 2×
“For FXIII, there is until now no published data concerning biological variation (Ricós et al., 1999).”
Background
Cited by
Diagnostic testing and interpretation of tests for autoimmunity
Castro, C., & Gourley, M. (2010) · Journal of Allergy and Clinical Immunology · doi:10.1016/j.jaci.2009.09.041 · Cited by 213
Moore, G. W., Jones, P. O., Platton, S., Hussain, N., White, D., Thomas, W., Rigano, J., Pouplard, C., Gray, E., & Devreese, K. M. J. (2021) · Journal of Thrombosis and Haemostasis · doi:10.1111/jth.15438 · Cited by 47
Effects of vigorous mixing of blood vacuum tubes on laboratory test results
Lima-Oliveira, G., Lippi, G., Salvagno, G. L., Montagnana, M., Gelati, M., Volanski, W., Boritiza, K. C., Picheth, G., & Guidi, G. C. (2013) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2012.10.033 · Cited by 33
Influence of posture on routine hemostasis testing
Lippi, G., Salvagno, G. L., Lima-Oliveira, G., Danese, E., Favaloro, E. J., & Guidi, G. C. (2015) · Blood Coagulation & Fibrinolysis · doi:10.1097/mbc.0000000000000326 · Cited by 27
Lima-Oliveira, G., Lippi, G., Salvagno, G. L., Brocco, G., Gaino, S., Dima, F., Volaski, W., Rego, F. G., Picheth, G., & Guidi, G. C. (2014) · Biopreservation and Biobanking · doi:10.1089/bio.2013.0043 · Cited by 22
Metadata
- Authors
- MILOS, M., HERAK, D., KURIC, L., HORVAT, I., & ZADRO, R.
- Venue
- International Journal of Laboratory Hematology
- Publisher
- Wiley
- Published
- 2009-01-11
- Volume
- 31
- Issue
- 1
- Pages
- 26-35
- ISSN
- 1751-5521 (ISSN-L)
- 1751-553X
- OpenAlex
- W1926470235
- Cited by
- 40
- References
- 2
- FWCI
- 1.38
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Evaluation and performance characteristics of the coagulation system: ACL TOP analyzer – HemosIL reagents. https://app.cochat.ai/papers/10.1111/j.1751-553x.2007.00999.x
- The paper
- MILOS, M., HERAK, D., KURIC, L., HORVAT, I., & ZADRO, R. (2009). Evaluation and performance characteristics of the coagulation system: ACL TOP analyzer – HemosIL reagents. International Journal of Laboratory Hematology. https://doi.org/10.1111/j.1751-553x.2007.00999.x
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.