CoChat

Can automated blood film analysis replace the manual differential? An evaluation of the CellaVision DM96 automated image analysis system

BRIGGS, C., LONGAIR, I., SLAVIK, M., THWAITE, K., MILLS, R., THAVARAJA, V., FOSTER, A., ROMANIN, D., MACHIN, S. J. · 2009 · International Journal of Laboratory Hematology
Cited by 154

The DOI matches its Crossref record.

Abstract

Summary: Automation of differentials is desirable for economic and time‐saving reasons. Over the last 20 years, automated imaging processes have started to be introduced where stained blood films are scanned by a computer‐driven microscope and leucocytes classified; however, early methods were slow and had difficulty in classifying abnormal cells. More recently the CellaVision™ DM96 (CellaVision AB, Lund, Sweden) has been introduced with added features such as continuous loading of slides and a faster throughput than previous instruments. The accuracy of CellaVision TM DM96 has been evaluated by comparing results to reference manual differentials. Results from different operators using the DM96 were compared with their own manual differential and to a 400‐cell reference manual differential. Precision of the instrument was compared to the manual differential. The preclassification accuracy of the DM96 was 89.2%. Precision was similar to that of the 100‐cell manual differential. The DM96 was faster than the manual method, even after reclassification by a laboratory scientist of any cells wrongly categorized by the instrument. The DM96 accuracy in morphological classification of leucocytes and red blood cells; depends upon both blood pathology and experience of the laboratory scientist using the instrument. For some cell types and operators, DM96 accuracy was better than the individual’s 100 cell manual differential.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1111/j.1751-553x.2007.01002.x

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 154

Metadata

Authors
BRIGGS, C., LONGAIR, I., SLAVIK, M., THWAITE, K., MILLS, R., THAVARAJA, V., FOSTER, A., ROMANIN, D., & MACHIN, S. J.
Venue
International Journal of Laboratory Hematology
Publisher
Wiley
Published
2009-01-11
Volume
31
Issue
1
Pages
48-60
ISSN
  • 1751-5521 (ISSN-L)
  • 1751-553X
OpenAlex
W1590583116
Cited by
154
References
0
FWCI
4.33
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Can automated blood film analysis replace the manual differential? An evaluation of the CellaVision DM96 automated image analysis system. https://app.cochat.ai/papers/10.1111/j.1751-553x.2007.01002.x
The paper
BRIGGS, C., LONGAIR, I., SLAVIK, M., THWAITE, K., MILLS, R., THAVARAJA, V., FOSTER, A., ROMANIN, D., & MACHIN, S. J. (2009). Can automated blood film analysis replace the manual differential? An evaluation of the CellaVision DM96 automated image analysis system. International Journal of Laboratory Hematology. https://doi.org/10.1111/j.1751-553x.2007.01002.x

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.