Cited by 29Blood Pressure and Hypertension StudiesNon-Invasive Vital Sign MonitoringCardiovascular Health and Disease Prevention

Blood pressure measurement in special populations and circumstances

Open access
Stergiou, George S., Dolan, Eamon, Kollias, Anastasios, Poulter, Neil R., Shennan, Andrew, Staessen, Jan A., Zhang, Zhen‐Yu, Weber, Michael A.·2018·The Journal of Clinical Hypertension

Abstract

According to the established validation protocols, a typical validation study of a blood pressure (BP) monitor includes general population adults with normal or elevated BP. It is recognized, however, that the automated (oscillometric) BP monitors may have different accuracy or uses in some special populations compared with adults in the general population. Thus, an automated BP monitor with proven accuracy in a general population of adults may not be accurate in a special population, and therefore separate validation is needed. Recognized special populations deserving separate validation are those for which there is theoretical, and also clinical evidence, that the accuracy of BP monitors in these groups differs from that in the general population. Young children, pregnant women (including those with preeclampsia), individuals with arm circumference >42 cm, and patients with atrial fibrillation are regarded as special populations. Adolescents, individuals older than 80 years, and patients with end‐stage renal disease or diabetes mellitus have also been considered as possible special groups, but there is still inadequate evidence of altered accuracy of BP monitors in these subjects. Validation studies should be performed in special populations and evaluated separately after the BP‐measuring device has successfully undergone a validation study in a general population (unless the test device is intended only for a special population). This article discusses issues relating to the measurement of BP and the diagnosis of hypertension in selected special populations, as well as in low‐resource settings, where a simplified yet efficient evaluation strategy is necessary.

Integrity signals

Registry

Crossref · 10.1111/jch.13296

Cites

9 of 9 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Claims the paper makes

We couldn’t retrieve this paper’s full text, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Stergiou, G. S., Dolan, E., Kollias, A., Poulter, N. R., Shennan, A., Staessen, J. A., Zhang, Z., & Weber, M. A.
Affiliations
  • National and Kapodistrian University of Athens (GR)
  • Connolly Hospital Blanchardstown (IE)
  • Hypertension Center STRIDE‐7 School of Medicine Third Department of Medicine Sotiria Hospital National and Kapodistrian University of Athens Greece
  • Imperial College London (GB)
  • St Thomas' Hospital (GB)
  • Maastricht University (NL)
  • KU Leuven (BE)
  • State University of New York (US)
Corresponding author
Stergiou, George S.
Venue
The Journal of Clinical Hypertension
Publisher
Wiley
Published
2018-07
Volume
20
Issue
7
Pages
1122-1127
ISSN
  • 1524-6175 (ISSN-L)
  • 1751-7176
OpenAlex
W2884316892
Cited by
29
References
9
Citations by year
2026: 3 · 2025: 4 · 2024: 2 · 2023: 4 · 2022: 4 · 2021: 4 · 2020: 1 · 2019: 4 · 2018: 3
FWCI
2.27
Type
journal article
Language
en
Open access
Locations
  • Journal of Clinical Hypertension (publishedVersion)
  • Research Publications (Maastricht University) (publishedVersion, other-oa)
  • PubMed Central (publishedVersion)
  • PubMed (publishedVersion)
  • Lirias (acceptedVersion, other-oa)
  • Research Portal (King's College London) (submittedVersion)
  • Research Publications (Maastricht University) (submittedVersion, other-oa)
Topics
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
  • Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
Keywords
blood pressure measurement, automated blood pressure monitors, special populations, children, pregnant women, preeclampsia, arm circumference, atrial fibrillation, hypertension diagnosis, resource-limited settings
MeSH terms
Adult, Aged, Aged, 80 and over, Atrial Fibrillation, Blood Pressure, Blood Pressure Determination, Child, Child, Preschool, Equipment Design, Female, Humans, Hypertension, Kidney Failure, Chronic, Middle Aged, Oscillometry, Pre-Eclampsia, Pregnancy, Reproducibility of Results, Blood Pressure Monitors, Prevalence, Sphygmomanometers

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review