Ion desorption by electron bombardment; relation to total and partial pressure measurement
Abstract
When electrons strike certain adsorbed gases, positive ions, neutrals, and/or excited neutrals are released. This electron impact desorption (E.I.D.) effect can cause significant errors in total pressure measurement with hot-cathode gauges and can change the striking characteristics of magnetic, cold-cathode gauges. In certain mass spectrometers, “surface ion” peaks appear in the mass spectrum resulting from E.I.D. of gases sorbed at the ion source electrodes. The general characteristics of E.I.D. and the results of measurements of E.I.D. from metal surfaces are briefly reviewed. The effects of E.I.D. on the characteristics of hot- and cold-cathode gauges and mass spectrometers are considered in some detail. Methods of minimizing errors caused by E.I.D. are discussed.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Mo, W and Pt-Ir show similar behaviour in terms of electron stimulated desorption, whilst Pt-clad Mo seems to have better properties and should be preferable as the grid material.
“Mo, W and Pt-Ir show similar behaviour in terms of ESD, whilst Pt-clad Mo seems to have better properties and should be preferable as the grid material.”
From A review on hot cathode ionisation gauges with focus on a suitable design for measurement accuracy and stability (2020) · cites this paper as [121]
Cited by
Sorted by citations
Jousten, K., Boineau, F., Bundaleski, N., Illgen, C., Setina, J., Teodoro, O. M. N. D., Vicar, M., & Wüest, M. (2020) · Vacuum · doi:10.1016/j.vacuum.2020.109545 · Cited by 34 · see citing works
- Authors
- Redhead, P. A.
- Venue
- Journal of Vacuum Science and Technology
- Publisher
- American Vacuum Society
- Published
- 1970-01
- Volume
- 7
- Issue
- 1
- Pages
- 182-187
- ISSN
- 0022-5355
- Cited by
- 35
- References
- 0
- Language
- en
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review