The sensitivity of bayard—alpert gauges

Redhead, P. A.·1969·Journal of Vacuum Science and Technology

Abstract

The sensitivity of Bayard–Alpert gauges has been measured as a function of the grid-filament spacing and found to vary by as much as a factor of 2.5. It is suggested that the large variation of sensitivity from one gauge to another results from variability of filament positioning. An hypothesis is proposed, and substantiated by test, which relates the sensitivity to the angular momentum, about the axis of the grid, with which electrons are launched into the grid. It is shown that electrons with small initial angular momentum have the longest orbits in the useful ionizing volume. A new type of Bayard–Alpert gauge has been developed, based on this hypothesis, whose sensitivity is independent of filament position and of the potential of the envelope. The sensitivity factor of this new design is about 55 Torr−1 for nitrogen.

Integrity signals

Registry

Crossref · 10.1116/1.1492719

Cites

Statistics

Recomputing p-values needs the paper’s full text.

Claims the paper makes

This paper’s full text isn’t openly available, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Groszkowski found a weak increase of 20% in sensitivity when cathode-anode distance was varied from 1 mm to 15 mm, different from Schütze and in contradiction of Redhead.

    “For the cathodeanode distance Groszkowski found a weak increase of 20% in sensitivity, when this distance was varied from 1 mm to 15 mm, which is somewhat different from the results of Schütze (he used a longitudinal wire, not hairpin as cathode) described above and in contradiction of a finding by Redhead [21], see below.”
  • Redhead found an increase in sensitivity by a factor of 2.5 when the hairpin cathode was moved radially away from the anode grid of a BAG from 0.5 mm to 6 mm.

    “Redhead [21] found an increase in sensitivity by a factor of 2.5, when the hairpin cathode was moved radially away from the anode grid of a BAG from 0.5 mm to 6 mm distance.”
  • A shield around the cathode opposite to the anode reduced the dependence on grid-cathode distance significantly for Lgf > 3 mm.

    “A shield around the cathode opposite to the anode reduced this dependence on grid-cathode distance L gf significantly for L gf > 3 mm [21].”

Cited by

Sorted by citations

Authors
Redhead, P. A.
Affiliations
Radio & Electrical Engineering Division, National Research Council, Ottawa, Canada
Venue
Journal of Vacuum Science and Technology
Published
1969-09-01
Volume
6
Issue
5
Pages
848-854
ISSN
0022-5355
Cited by
22
References
0
Language
en

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review