Data-driven detection and registration of spine surgery instrumentation in intraoperative images
Open accessAbstract
Purpose. Conventional model-based 3D-2D registration algorithms can be challenged by limited capture range, model validity, and stringent intraoperative runtime requirements. In this work, a deep convolutional neural network was used to provide robust initialization of a registration algorithm (known-component registration, KC-Reg) for 3D localization of spine surgery implants, combining the speed and global support of data-driven approaches with the previously demonstrated accuracy of model-based registration. Methods. The approach uses a Faster R-CNN architecture to detect and localize a broad variety and orientation of spinal pedicle screws in clinical images. Training data were generated using projections from 17 clinical cone-beam CT scans and a library of screw models to simulate implants. Network output was processed to provide screw count and 2D poses. The network was tested on two test datasets of 2,000 images, each depicting real anatomy and realistic spine surgery instrumentation – one dataset involving the same patient data as in the training set (but with different screws, poses, image noise, and affine transformations) and one dataset with five patients unseen in the test data. Assessment of device detection was quantified in terms of accuracy and specificity, and localization accuracy was evaluated in terms of intersection-overunion (IOU) and distance between true and predicted bounding box coordinates. Results. The overall accuracy of pedicle screw detection was ~86.6% (85.3% for the same-patient dataset and 87.8% for the many-patient dataset), suggesting that the screw detection network performed reasonably well irrespective of disparate, complex anatomical backgrounds. The precision of screw detection was ~92.6% (95.0% and 90.2% for the respective same-patient and many-patient datasets). The accuracy of screw localization was within 1.5 mm (median difference of bounding box coordinates), and median IOU exceeded 0.85. For purposes of initializing a 3D-2D registration algorithm, the accuracy was observed to be well within the typical capture range of KC-Reg.1 Conclusions. Initial evaluation of network performance indicates sufficient accuracy to integrate with algorithms for implant registration, guidance, and verification in spine surgery. Such capability is of potential use in surgical navigation, robotic assistance, and data-intensive analysis of implant placement in large retrospective datasets. Future work includes correspondence of multiple views, 3D localization, screw classification, and expansion of the training dataset to a broader variety of anatomical sites, number of screws, and types of implants.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Each contextualization study that considered multiple objects did not explicitly isolate parameters for each object, but used a single NN with output layer indexed by object.
“However, we found that each of the contextualization studies which considered multiple objects (Grupp et al., 2020c; Doerr et al., 2020; Wang et al., 2020; Esfandiari et al., 2021) did not explicitly isolate the parameters corresponding to each object, but instead used a single NN with data in the output layer indexed according to the appropriate object.”
From The Impact of Machine Learning on 2D/3D Registration for Image-Guided Interventions: A Systematic Review and Perspective (2021) · cites this paper as Doerr et al., 2020
The Methods section (page unavailable) describes one Faster R-CNN that handles multiple regions of interest, and the Discussion says the training images contained two screws. That supports the shared-network, multiple-object part of the claim. The Methods describe outputs per region of interest and bounding-box coordinates by class, however, not an output layer explicitly indexed by individual object. (LLM, conf=0.83)
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Object Detection With Deep Learning: A Review
Zhao, Z.-Q., Zheng, P., Xu, S.-T., & Wu, X. (2019) · IEEE Transactions on Neural Networks and Learning Systems · doi:10.1109/tnnls.2018.2876865 · Cited by 5,545 · see citing works
Lehman, R. A., Lenke, L. G., Keeler, K. A., Kim, Y. J., & Cheh, G. (2007) · Spine · doi:10.1097/brs.0b013e31815a7f13 · Cited by 131 · see citing works
Impact of spine surgery complications on costs associated with management of adult spinal deformity
Yeramaneni, S., Robinson, C., & Hostin, R. A. (2016) · Current Reviews in Musculoskeletal Medicine · doi:10.1007/s12178-016-9352-9 · Cited by 128 · see citing works
Uneri, A., Silva, T. D., Goerres, J., Jacobson, M. W., Ketcha, M. D., Reaungamornrat, S., Kleinszig, G., Vogt, S. K., Khanna, A. J., Osgood, G. M., et al. (2017) · Physics in Medicine and Biology · doi:10.1088/1361-6560/aa62c5 · Cited by 42 · see citing works
Self-citationi3PosNet: instrument pose estimation from X-ray in temporal bone surgery
Kügler, D., Sehring, J., Stefanov, A., Stenin, I., Kristin, J., Klenzner, T., Schipper, J., & Mukhopadhyay, A. (2020) · International Journal of Computer Assisted Radiology and Surgery · doi:10.1007/s11548-020-02157-4 · Cited by 39 · see citing works
Low-dose preview for patient-specific, task-specific technique selection in cone-beam CT
Wang, A., Stayman, J. W., Otake, Y., Vogt, S. K., Kleinszig, G., Khanna, A. J., Gallia, G. L., & Siewerdsen, J. H. (2014) · Medical Physics · doi:10.1118/1.4884039 · Cited by 35 · see citing works
Self-citationKnown-component metal artifact reduction (KC-MAR) for cone-beam CT
Uneri, A., Zhang, X., Yi, T. Y., Stayman, J. W., Helm, P. A., Osgood, G. M., Theodore, N., & Siewerdsen, J. H. (2019) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ab3036 · Cited by 25 · see citing works
Self-citationZhang, X., Uneri, A., Stayman, J. W., Zygourakis, C. C., Lo, S.-F. L., Theodore, N., & Siewerdsen, J. H. (2019) · Medical Physics · doi:10.1002/mp.13652 · Cited by 15 · see citing works
Self-citation
Flagged first, then by citations
Research works printed but not linked: 3
- [5]Suzani A, Seitel A, Liu Y, Fels S, Roholing RN. Fast automatic vertebrae detection and localization in pathological CT scans – a deep learning approach. In: MICCAI.; 2015.
- [7]Kügler D, Stefanov A, Mukhopadhyay A. i3PosNet: Instrument Pose Estimation from X-Ray. IEEE Trans Med Imaging. 2018 http://arxiv.org/abs/1802.09575. PMC7316684 32440956 · arXiv
- [11]Ren S, He K, Girshick R, Sun J. Faster r-cnn:Towards real-time object detection with region proposal networks. In: Advances in Neural Information Processing Systems.; 2015: 91– 99.
Cited by
Sorted by citations
Unberath, M., Gao, C., Hu, Y., Judish, M., Taylor, R. H., Armand, M., & Grupp, R. (2021) · Frontiers in Robotics and AI · doi:10.3389/frobt.2021.716007 · Cited by 102 · see citing works
Doerr, S. A., Weber-Levine, C., Hersh, A. M., Awosika, T., Judy, B. F., Jin, Y., Raj, D., Liu, A., Lubelski, D., Jones, C. K., et al. (2022) · Neurosurgical FOCUS · doi:10.3171/2022.1.focus21745 · Cited by 35 · see citing works
Self-citationZhang, X., Uneri, A., Wu, P., Ketcha, M. D., Jones, C. K., Huang, Y., Lo, S.-F. L., Helm, P. A., & Siewerdsen, J. H. (2021) · Physics in Medicine and Biology · doi:10.1088/1361-6560/abde96 · Cited by 17 · see citing works
Self-citationUneri, A., Wu, P., Jones, C. K., Vagdargi, P., Han, R., Helm, P. A., Luciano, M. G., Anderson, W. S., & Siewerdsen, J. H. (2021) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ac2f89 · Cited by 6 · see citing works
Self-citationSurgical navigation for guidewire placement from intraoperative fluoroscopy in orthopaedic surgery
Mekki, L., Sheth, N., Vijayan, R. C., Rohleder, M., Sisniega, A., Kleinszig, G., Vogt, S. K., Kunze, H., Osgood, G. M., Siewerdsen, J. H., et al. (2023) · Physics in Medicine and Biology · doi:10.1088/1361-6560/acfec4 · Cited by 5 · see citing works
Self-citationUneri, A., Wu, P., Jones, C. K., Ketcha, M. D., Vagdargi, P., Han, R., Helm, P. A., Luciano, M. G., Anderson, W. S., & Siewerdsen, J. H. (2021) · SPIE Medical Imaging: Image-Guided Procedures, Robotic Interventions, and Modeling · doi:10.1117/12.2582160 · Cited by 4 · see citing works
Self-citationPark, J., Yeom, J. S., Kim, Y.-H., Hwang, Y., Kim, N., & Park, S. (2024) · Medicina · doi:10.3390/medicina60122040 · Cited by 1 · see citing works
Tuoheti, A., Gufuding, Mijiti, M., Yang, J., Habulihan, H., Abuliken, A., Mamuti, Y., & Rexiti, P. (2026) · Global Spine Journal · doi:10.1177/21925682261485860 · Cited by 1 · see citing works
Yang, E., Howell, H. J., Schonfeld, E., Fuller, J., Khan, F., Ayloo, B., Izima, C., Sampath, S. G., Tang, A. J., Rolfe, N. W., et al. (2026) · Artificial Intelligence Surgery · doi:10.20517/ais.2025.98
Uchiyama, T., Sakata, Y., Hirai, R., Ishikawa, H., & Mori, S. (2026) · IEEE Winter Conference on Applications of Computer Vision · doi:10.1109/wacv61042.2026.00262
- Authors
- Doerr, S. A., Uneri, A., Huang, Y., Jones, C. K., Zhang, X., Ketcha, M. D., Helm, P., & Siewerdsen, J. H.
- Affiliations
- Johns Hopkins University (US)
- Medtronic (United States) (US)
- Corresponding author
- Siewerdsen, Jeffrey H.
- ORCID
- Venue
- Medical Imaging 2020: Image-Guided Procedures, Robotic Interventions, and Modeling
- Publisher
- SPIE
- Published
- 2020-03-16
- Volume
- 11315
- Pages
- 91
- PMC
- PMC9450103
- OpenAlex
- W3010790897
- Cited by
- 10
- References
- 8
- Citations by year
- 2026: 3 · 2024: 1 · 2023: 1 · 2022: 1 · 2021: 4
- FWCI
- 2.72
- Type
- conference paper
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- SPIE Medical Imaging: Image-Guided Procedures, Robotic Interventions, and Modeling (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- National Institute of Biomedical Imaging and Bioengineering (R01EB017226)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Medical Imaging and Analysis
- Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Medical Image Segmentation Techniques
- Health Sciences › Medicine › Surgery › Spinal Fractures and Fixation Techniques
- Keywords
- spinal surgery, 2D-3D registration, implant localization, intraoperative imaging, Faster R-CNN, cone-beam computed tomography, surgical navigation, robotic assistance
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review