Assessment of display performance for medical imaging systems: Executive summary of AAPM TG18 report
Abstract
Digital imaging provides an effective means to electronically acquire, archive, distribute, and view medical images. Medical imaging display stations are an integral part of these operations. Therefore, it is vitally important to assure that electronic display devices do not compromise image quality and ultimately patient care. The AAPM Task Group 18 (TG18) recently published guidelines and acceptance criteria for acceptance testing and quality control of medical display devices. This paper is an executive summary of the TG18 report. TG18 guidelines include visual, quantitative, and advanced testing methodologies for primary and secondary class display devices. The characteristics, tested in conjunction with specially designed test patterns (i.e., TG18 patterns), include reflection, geometric distortion, luminance, the spatial and angular dependencies of luminance, resolution, noise, glare, chromaticity, and display artifacts. Geometric distortions are evaluated by linear measurements of the TG18‐QC test pattern, which should render distortion coefficients less than 2%/5% for primary/secondary displays, respectively. Reflection measurements include specular and diffuse reflection coefficients from which the maximum allowable ambient lighting is determined such that contrast degradation due to display reflection remains below a 20% limit and the level of ambient luminance does not unduly compromise luminance ratio (LR) and contrast at low luminance levels. Luminance evaluation relies on visual assessment of low contrast features in the TG18‐CT and TG18‐MP test patterns, or quantitative measurements at 18 distinct luminance levels of the TG18‐LN test patterns. The major acceptable criteria for primary/secondary displays are maximum luminance of greater than , LR of greater than , and contrast conformance to that of the grayscale standard display function (GSDF) of better than , respectively. The angular response is tested to ascertain the viewing cone within which contrast conformance to the GSDF is better than and LR is greater than for primary/secondary displays, or alternatively, within which the on‐axis contrast thresholds of the TG18‐CT test pattern remain discernible. The evaluation of luminance spatial uniformity at two distinct luminance levels across the display faceplate using TG18‐UNL test patterns should yield nonuniformity coefficients smaller than 30%. The resolution evaluation includes the visual scoring of the CX test target in the TG18‐QC or TG18‐CX test patterns, which should yield scores greater than for primary/secondary displays. Noise evaluation includes visual evaluation of the contrast threshold in the TG18‐AFC test pattern, which should yield a minimum of targets visible for primary/secondary displays. The guidelines also include methodologies for more quantitative resolution and noise measurements based on MTF and NPS analyses. The display glare test, based on the visibility of the low‐contrast targets of the TG18‐GV test pattern or the measurement of the glare ratio (GR), is expected to yield scores greater than and GRs greater than for primary/secondary displays. Chromaticity, measured across a display faceplate or between two display devices, is expected to render a , color separation of less than 0.01 for primary displays. The report offers further descriptions of prior standardization efforts, current display technologies, testing prerequisites, streamlined procedures and timelines, and TG18 test patterns.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Organic electroluminescent diodes
Tang, C. W., & VanSlyke, S. A. (1987) · Applied Physics Letters · doi:10.1063/1.98799 · Cited by 14,363
Contrast Sensitivity of the Human Eye and Its Effects on Image Quality
Barten, P. G. (1999) · SPIE eBooks · doi:10.1117/3.353254 · Cited by 808
Rose, A., & Biberman, L. (1975) · Physics Today · doi:10.1063/1.3069240 · Cited by 378
High performance organic polymer light-emitting heterostructure devices
He, Y., Gong, S. C., Hattori, R., & Kanicki, J. (1999) · Applied Physics Letters · doi:10.1063/1.123862 · Cited by 128
Subtle orthopedic fractures: teleradiology workstation versus film interpretation.
Scott, W. W., Rosenbaum, J. E., Ackerman, S. J., Reichle, R., Magid, D., Weller, J. C., & Gitlin, J. N. (1993) · Radiology · doi:10.1148/radiology.187.3.8497636 · Cited by 83
Evaluation of video gray‐scale display
Nawfel, R. D., Chan, K. H., Wagenaar, D. J., & Judy, P. F. (1992) · Medical Physics · doi:10.1118/1.596846 · Cited by 45
Badano, A. G., & Kanicki, J. (2001) · Journal of Applied Physics · doi:10.1063/1.1385571 · Cited by 34
Self-citationHorii, S. C. (1995) · Journal of Imaging Informatics in Medicine · doi:10.1007/bf03168049 · Cited by 5
Cited by
European guidelines for quality assurance in breast cancer screening and diagnosis
Amendoeira, I., Perry, N., Broeders, M. J. M., de Wolf, C., Törnberg, S., Holland, R. D., & von Karsa, L. (2013) · Zurich Open Repository and Archive (University of Zurich) · doi:10.2772/13196 · Cited by 1,556
Richard, S., Husarik, D. B., Yadava, G., Murphy, S. N., & Samei, E. (2012) · Medical Physics · doi:10.1118/1.4725171 · Cited by 419
Self-citationImage quality in CT: From physical measurements to model observers
Verdun, F. R., Racine, D., Ott, J. G., Tapiovaara, M. J., Toroi, P., Bochud, F. O., Veldkamp, W. J. H., Schegerer, A., Bouwman, R. W., Giron, I. H., et al. (2015) · Physica Medica · doi:10.1016/j.ejmp.2015.08.007 · Cited by 300
Perceptual and interpretive error in diagnostic radiology—causes and potential solutions
Degnan, A. J., Ghobadi, E. H., Hardy, P., Krupinski, E., Scali, E. P., Stratchko, L., Ulano, A., Walker, E., Wasnik, A. P., & Auffermann, W. F. (2019) · Academic Radiology · doi:10.1016/j.acra.2018.11.006 · Cited by 121
Metadata
- Authors
- Samei, E., Badano, A., Chakraborty, D., Compton, K., Cornelius, C., Corrigan, K., Flynn, M. J., Hemminger, B., Hangiandreou, N., Johnson, J., Moxley‐Stevens, D. M., Pavlicek, W., Roehrig, H., Rutz, L., Samei, E., Shepard, J., Uzenoff, R. A., Wang, J., & Willis, C. E.
- Venue
- Medical Physics
- Publisher
- Wiley
- Published
- 2005-04
- Volume
- 32
- Issue
- 4
- Pages
- 1205-1225
- ISSN
- 0094-2405 (ISSN-L)
- 2473-4209
- OpenAlex
- W2088216094
- Cited by
- 410
- References
- 8
- FWCI
- 19.28
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Assessment of display performance for medical imaging systems: Executive summary of AAPM TG18 report. https://app.cochat.ai/papers/10.1118/1.1861159
- The paper
- Samei, E., Badano, A., Chakraborty, D., Compton, K., Cornelius, C., Corrigan, K., Flynn, M. J., Hemminger, B., Hangiandreou, N., Johnson, J., Moxley‐Stevens, D. M., Pavlicek, W., Roehrig, H., Rutz, L., Samei, E., Shepard, J., Uzenoff, R. A., Wang, J., & Willis, C. E. (2005). Assessment of display performance for medical imaging systems: Executive summary of AAPM TG18 report. Medical Physics. https://doi.org/10.1118/1.1861159
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.