The DOI matches its Crossref record; 5 of the 5 references it cites were found in a registry or an index.
Towards task-based assessment of CT performance: System and object MTF across different reconstruction algorithms
Abstract
PURPOSE: To investigate a measurement method for evaluating the resolution properties of CT imaging systems across reconstruction algorithms, dose, and contrast. METHODS: An algorithm was developed to extract the task-based modulation transfer function (MTF) from disk images generated from the rod inserts in the ACR phantom (model 464 Gammex, WI). These inserts are conventionally employed for HU accuracy assessment. The edge of the disk objects was analyzed to determine the edge-spread function, which was differentiated to yield the line-spread function and Fourier-transformed to generate the object-specific MTF for task-based assessment, denoted MTF(Task). The proposed MTF measurement method was validated against the conventional wire technique and further applied to measure the MTF of CT images reconstructed with an adaptive statistical iterative algorithm (ASIR) and a model-based iterative (MBIR) algorithm. Results were further compared to the standard filtered back projection (FBP) algorithm. Measurements were performed and compared across different doses and contrast levels to ascertain the MTF(Task) dependencies on those factors. RESULTS: For the FBP reconstructed images, the MTF(Task) measured with the inserts were the same as the MTF measured from the wire-based method. For the ASIR and MBIR data, the MTF(Task) using the high contrast insert was similar to the wire-based MTF and equal or superior to that of FBP. However, results for the MTF(Task) measured using the low-contrast inserts, the MTF(Task) for ASIR and MBIR data was lower than for the FBP, which was constant throughout all measurements. Similarly, as a function of mA, the MTF(Task) for ASIR and MBIR varied as a function of noise--with MTF(Task) being proportional to mA. Overall greater variability of MTF(Task) across dose and contrast was observed for MBIR than for ASIR. CONCLUSIONS: This approach provides a method for assessing the task-based MTF of a CT system using conventional and iterative reconstructions. Results demonstrated that the object-specific MTF can vary as a function of dose and contrast. The analysis highlighted the paradigm shift for iterative reconstructions when compared to FBP, where iterative reconstructions generally offer superior noise performance but with varying resolution as a function of dose and contrast. The MTF(Task) generated by this method is expected to provide a more comprehensive assessment of image resolution across different reconstruction algorithms and imaging tasks.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Prescribing, recording, and reporting photon beam therapy
Icru (1993) · Medical Entomology and Zoology · Cited by 1,996
A three‐dimensional statistical approach to improved image quality for multislice helical CT
Thibault, J., Sauer, K. D., Bouman, C. A., & Hsieh, J. (2007) · Medical Physics · doi:10.1118/1.2789499 · Cited by 937
Assessment of display performance for medical imaging systems: Executive summary of AAPM TG18 report
Samei, E., Badano, A., Chakraborty, D., Compton, K., Cornelius, C., Corrigan, K., Flynn, M. J., Hemminger, B., Hangiandreou, N., Johnson, J., et al. (2005) · Medical Physics · doi:10.1118/1.1861159 · Cited by 410
Self-citationHandbook of Medical Imaging, Volume 1. Physics and Psychophysics
(2000) · SPIE eBooks · doi:10.1117/3.832716 · Cited by 175
Deriving the modulation transfer function of CT from extremely noisy edge profiles
Mori, I., & Machida, Y. (2008) · Radiological Physics and Technology · doi:10.1007/s12194-008-0039-9 · Cited by 41
Cited by
Image quality in CT: From physical measurements to model observers
Verdun, F. R., Racine, D., Ott, J. G., Tapiovaara, M. J., Toroi, P., Bochud, F. O., Veldkamp, W. J. H., Schegerer, A., Bouwman, R. W., Giron, I. H., et al. (2015) · Physica Medica · doi:10.1016/j.ejmp.2015.08.007 · Cited by 300
Optimisation of Radiological Protection in Digital Radiology Techniques for Medical Imaging
154, I. P. (2023) · Annals of the ICRP · doi:10.1177/01466453231210646 · Cited by 37
Metadata
- Authors
- Richard, S., Husarik, D. B., Yadava, G., Murphy, S. N., & Samei, E.
- Venue
- Medical Physics
- Publisher
- Wiley
- Published
- 2012-06-11
- Volume
- 39
- Issue
- 7Part1
- Pages
- 4115-4122
- ISSN
- 0094-2405 (ISSN-L)
- OpenAlex
- W1975265909
- Cited by
- 419
- References
- 5
- FWCI
- 15.58
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Towards task-based assessment of CT performance: System and object MTF across different reconstruction algorithms. https://app.cochat.ai/papers/10.1118/1.4725171
- The paper
- Richard, S., Husarik, D. B., Yadava, G., Murphy, S. N., & Samei, E. (2012). Towards task-based assessment of CT performance: System and object MTF across different reconstruction algorithms. Medical Physics. https://doi.org/10.1118/1.4725171
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.