A comparative study of limited‐angle cone‐beam reconstruction methods for breast tomosynthesis

Open access
Zhang, Yiheng, Chan, Heang‐Ping, Sahiner, Berkman, Wei, Jun, Goodsitt, Mitchell M., Hadjiiski, Lubomir M., Ge, Jun, Zhou, Chuan·2006·Medical Physics

Abstract

Digital tomosynthesis mammography (DTM) is a promising new modality for breast cancer detection. In DTM, projection‐view images are acquired at a limited number of angles over a limited angular range and the imaged volume is reconstructed from the two‐dimensional projections, thus providing three‐dimensional structural information of the breast tissue. In this work, we investigated three representative reconstruction methods for this limited‐angle cone‐beam tomographic problem, including the backprojection (BP) method, the simultaneous algebraic reconstruction technique (SART) and the maximum likelihood method with the convex algorithm (ML‐convex). The SART and ML‐convex methods were both initialized with BP results to achieve efficient reconstruction. A second generation GE prototype tomosynthesis mammography system with a stationary digital detector was used for image acquisition. Projection‐view images were acquired from 21 angles in increments over a angular range. We used an American College of Radiology phantom and designed three additional phantoms to evaluate the image quality and reconstruction artifacts. In addition to visual comparison of the reconstructed images of different phantom sets, we employed the contrast‐to‐noise ratio (CNR), a line profile of features, an artifact spread function (ASF), a relative noise power spectrum (NPS), and a line object spread function (LOSF) to quantitatively evaluate the reconstruction results. It was found that for the phantoms with homogeneous background, the BP method resulted in less noisy tomosynthesized images and higher CNR values for masses than the SART and ML‐convex methods. However, the two iterative methods provided greater contrast enhancement for both masses and calcification, sharper LOSF, and reduced interplane blurring and artifacts with better ASF behaviors for masses. For a contrast‐detail phantom with heterogeneous tissue‐mimicking background, the BP method had strong blurring artifacts along the x‐ray source motion direction that obscured the contrast‐detail objects, while the other two methods can remove the superimposed breast structures and significantly improve object conspicuity. With a properly selected relaxation parameter, the SART method with one iteration can provide tomosynthesized images comparable to those obtained from the ML‐convex method with seven iterations, when BP results were used as initialization for both methods.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1118/1.2237543

Cites

16 of 16 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 16 references

Cited by

See all 285

Metadata

Authors
Zhang, Y., Chan, H., Sahiner, B., Wei, J., Goodsitt, M. M., Hadjiiski, L. M., Ge, J., & Zhou, C.
Venue
Medical Physics
Publisher
Wiley
Published
2006-09-25
Volume
33
Issue
10
Pages
3781-3795
ISSN
  • 0094-2405 (ISSN-L)
  • 2473-4209
OpenAlex
W2045563668
Cited by
285
References
16
FWCI
15.03
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: A comparative study of limited‐angle cone‐beam reconstruction methods for breast tomosynthesis. https://app.cochat.ai/papers/10.1118/1.2237543
The paper
Zhang, Y., Chan, H., Sahiner, B., Wei, J., Goodsitt, M. M., Hadjiiski, L. M., Ge, J., & Zhou, C. (2006). A comparative study of limited‐angle cone‐beam reconstruction methods for breast tomosynthesis. Medical Physics. https://doi.org/10.1118/1.2237543

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.