The DOI matches its Crossref record.
Dual‐source spiral CT with pitch up to 3.2 and 75 ms temporal resolution: Image reconstruction and assessment of image quality
Abstract
Purpose: To present the theory for image reconstruction of a high‐pitch, high‐temporal‐resolution spiral scan mode for dual‐source CT (DSCT) and evaluate its image quality and dose. Methods: With the use of two x‐ray sources and two data acquisition systems, spiral CT exams having a nominal temporal resolution per image of up to one‐quarter of the gantry rotation time can be acquired using pitch values up to 3.2. The scan field of view (SFOV) for this mode, however, is limited to the SFOV of the second detector as a maximum, depending on the pitch. Spatial and low contrast resolution, image uniformity and noise, CT number accuracy and linearity, and radiation dose were assessed using the ACR CT accreditation phantom, a 30 cm diameter cylindrical water phantom or a 32 cm diameter cylindrical PMMA CTDI phantom. Slice sensitivity profiles (SSPs) were measured for different nominal slice thicknesses, and an anthropomorphic phantom was used to assess image artifacts. Results were compared between single‐source scans at and dual‐source scans at . In addition, image quality and temporal resolution of an ECG‐triggered version of the DSCT high‐pitch spiral scan mode were evaluated with a moving coronary artery phantom, and radiation dose was assessed in comparison with other existing cardiac scan techniques. Results: No significant differences in quantitative measures of image quality were found between single‐source scans at and dual‐source scans at for spatial and low contrast resolution, CT number accuracy and linearity, SSPs, image uniformity, and noise. The pitch value had only a minor impact on radiation dose and image noise when the effective tube current time product (mA s/pitch) was kept constant. However, while not severe, artifacts were found to be more prevalent for the dual‐source scan mode when structures varied markedly along the axis, particularly for head scans. Images of the moving coronary artery phantom acquired with the ECG‐triggered high‐pitch scan mode were visually free from motion artifacts at heart rates of 60 and 70 bpm. However, image quality started to deteriorate for higher heart rates. At equivalent image quality, the ECG‐triggered high‐pitch scan mode demonstrated lower radiation dose than other cardiac scan techniques on the same DSCT equipment (25% and 60% dose reduction compared to ECG‐triggered sequential step‐and‐shoot and ECG‐gated spiral with x‐ray pulsing). Conclusions: A high‐pitch (up to), high‐temporal‐resolution (up to 75 ms) dual‐source CT scan mode produced equivalent image quality relative to single‐source scans using a more typical pitch value . The resultant reduction in the overall acquisition time may offer clinical advantage for cardiovascular, trauma, and pediatric CT applications. In addition, ECG‐triggered high‐pitch scanning may be useful as an alternative to ECG‐triggered sequential scanning for patients with low to moderate heart rates up to 70 bpm, with the potential to scan the heart within one heart beat at reduced radiation dose.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Managing Radiation Use in Medical Imaging: A Multifaceted Challenge
Hricak, H., Brenner, D. J., Adelstein, S. J., Frush, D. P., Hall, E. J., Howell, R. W., McCollough, C. H., Mettler, F. A., Pearce, M. S., Suleiman, O. H., et al. (2011) · Radiology · doi:10.1148/radiol.10101157 · Cited by 342
Self-citationComputed tomography and patient risk: Facts, perceptions and uncertainties
Power, S. P., Moloney, F., Twomey, M., James, K., O’Connor, O. J., & Maher, M. M. (2016) · World Journal of Radiology · doi:10.4329/wjr.v8.i12.902 · Cited by 335
Han, B. K., Rigsby, C. K., Hlavacek, A., Leipsic, J., Nicol, E. D., Siegel, M. J., Bardo, D., Abbara, S., Ghoshhajra, B., Lesser, J. R., et al. (2015) · Journal of Cardiovascular Computed Tomography · doi:10.1016/j.jcct.2015.07.004 · Cited by 211
Metadata
- Authors
- Flohr, T. G., Leng, S., Yu, L., Allmendinger, T., Bruder, H., Petersilka, M., Eusemann, C. D., Stierstorfer, K., Schmidt, B., & McCollough, C. H.
- Venue
- Medical Physics
- Publisher
- Wiley
- Published
- 2009-11-23
- Volume
- 36
- Issue
- 12
- Pages
- 5641-5653
- ISSN
- 0094-2405 (ISSN-L)
- 2473-4209
- OpenAlex
- W1966329603
- Cited by
- 167
- References
- 0
- FWCI
- 10.10
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.