Characterization of image quality and image‐guidance performance of a preclinical microirradiator
Open accessThe DOI matches its Crossref record; 10 of the 10 references it cites were found in a registry or an index.
Abstract
Purpose: To assess image quality and image‐guidance capabilities of a cone‐beam CT based small‐animal image‐guided irradiation unit (micro‐IGRT). Methods: A micro‐IGRT system has been developed in collaboration with the authors' laboratory as a means to study the radiobiological effects of conformal radiation dose distributions in small animals. The system, the X‐Rad 225Cx, consists of a 225 kVp x‐ray tube and a flat‐panel amorphous silicon detector mounted on a rotational C‐arm gantry and is capable of both fluoroscopic x‐ray and cone‐beam CT imaging, as well as image‐guided placement of the radiation beams. Image quality (voxel noise, modulation transfer, CT number accuracy, and geometric accuracy characteristics) was assessed using water cylinder and micro‐CT test phantoms. Image guidance was tested by analyzing the dose delivered to radiochromic films fixed to BB's through the end‐to‐end process of imaging, targeting the center of the BB, and irradiation of the film/BB in order to compare the offset between the center of the field and the center of the BB. Image quality and geometric studies were repeated over a 5–7 month period to assess stability. Results: CT numbers reported were found to be linear and the noise for images of homogeneous water phantom was 30 HU at imaging doses of approximately 1 cGy (to water). The presampled MTF at 50% and 10% reached 0.64 and , respectively. Targeting accuracy by means of film irradiations was shown to have a mean displacement error of , with standard deviations of [0.02, 0.20, 0.17] mm. The system has proven to be stable over time, with both the image quality and image‐guidance performance being reproducible for the duration of the studies. Conclusions: The micro‐IGRT unit provides soft‐tissue imaging of small‐animal anatomy at acceptable imaging doses. The geometric accuracy and targeting systems permit dose placement with submillimeter accuracy and precision. The system has proven itself to be stable over 2 yr of routine laboratory use and provides a platform for the exploration of targeted radiation effects in small‐animal models.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Feldkamp, L. A., Davis, L. C., & Kress, J. W. (1984) · Journal of the Optical Society of America A · doi:10.1364/josaa.1.000612 · Cited by 6,319
Flat-panel cone-beam computed tomography for image-guided radiation therapy
Jaffray, D. A., Siewerdsen, J. H., Wong, J. W., & Martinez, A. A. (2002) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/s0360-3016(02)02884-5 · Cited by 1,330
Self-citationA system for stereotactic radiosurgery with a linear accelerator
Lutz, W., Winston, K. R., & Maleki, N. (1988) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/0360-3016(88)90446-4 · Cited by 757
Scaling down imaging: molecular mapping of cancer in mice
Weissleder, R. (2002) · Nature Reviews Cancer · doi:10.1038/nrc701 · Cited by 708
Wong, J., Armour, E., Kazanzides, P., Iordachita, I., Tryggestad, E., Deng, H., Matinfar, M., Kennedy, C., Liu, Z., Chan, T., et al. (2008) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2008.04.025 · Cited by 364
Partial volume rat lung irradiation: An evaluation of early DNA damage
Ma, K., Hill, R. P., & Dyk, J. V. (1998) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/s0360-3016(97)00736-0 · Cited by 182
Self-citationA quality assurance phantom for the performance evaluation of volumetric micro-CT systems
Du, L. Y., Umoh, J. U., Nikolov, H. N., Pollmann, S. I., Lee, T., & Holdsworth, D. W. (2007) · Physics in Medicine and Biology · doi:10.1088/0031-9155/52/23/021 · Cited by 165
Graves, E. E., Zhou, H., Chatterjee, R., Keall, P., Gambhir, S. S., Contag, C. H., & Boyer, A. L. (2007) · Medical Physics · doi:10.1118/1.2789498 · Cited by 99
Zhou, H., Rodriguez, M., van den Haak, F., Nelson, G., Jogani, R., Jiali, X., Zhu, X., Xian, Y., Tran, P. T., Felsher, D. W., et al. (2010) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2009.11.008 · Cited by 79
Langan, A., Ma, K., Yeung, I. W. T., Dyk, J. V., & Hill, R. P. (2006) · Radiotherapy and Oncology · doi:10.1016/j.radonc.2006.03.017 · Cited by 61
Self-citation
Cited by
Divergent clonal selection dominates medulloblastoma at recurrence
Morrissy, A. S., Garzia, L., Shih, D. J. H., Zuyderduyn, S., Huang, X., Skowron, P., Remke, M., Cavalli, F. M. G., Ramaswamy, V., Lindsay, P. E., et al. (2016) · Nature · doi:10.1038/nature16478 · Cited by 334
Self-citationRadiation-induced bystander and abscopal effects: important lessons from preclinical models
Daguenet, E., Louati, S., Wozny, A.-S., Vial, N., Gras, M., Guy, J.-B., Vallard, A., Rodriguez-Lafrasse, C., & Magné, N. (2020) · British Journal of Cancer · doi:10.1038/s41416-020-0942-3 · Cited by 178
Guidelines for preclinical and early phase clinical assessment of novel radiosensitisers
Harrington, K. J., Billingham, L. J., Brunner, T. B., Burnet, N. G., Chan, C. S., Hoskin, P., Mackay, R. I., Maughan, T. S., Macdougall, J., McKenna, W. G., et al. (2011) · British Journal of Cancer · doi:10.1038/bjc.2011.240 · Cited by 170
“It is now possible to model conformal clinical RT in animal models using state-of-the-art image-guided irradiation of tumours growing in orthotopic sites (Saha et al, 2010; Clarkson et al, 2011).”
BackgroundChattopadhyay, N., Cai, Z., Kwon, Y. L., Lechtman, E., Pignol, J.-P., & Reilly, R. M. (2013) · Breast Cancer Research and Treatment · doi:10.1007/s10549-012-2338-4 · Cited by 162
White and Gray Matter Abnormalities After Cranial Radiation in Children and Mice
Nieman, B. J., de Guzman, A. E., Gazdzinski, L. M., Lerch, J. P., Chakravarty, M. M., Pipitone, J., Strother, D., Fryer, C., Bouffet, E., Laughlin, S., et al. (2015) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2015.07.2293 · Cited by 62
Metadata
- Authors
- Clarkson, R., Lindsay, P. E., Ansell, S., Wilson, G., Jelveh, S., Hill, R. P., & Jaffray, D. A.
- Venue
- Medical Physics
- Publisher
- Wiley
- Published
- 2011-01-21
- Volume
- 38
- Issue
- 2
- Pages
- 845-856
- ISSN
- 0094-2405 (ISSN-L)
- 2473-4209
- OpenAlex
- W2002760403
- Cited by
- 172
- References
- 10
- FWCI
- 13.46
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Characterization of image quality and image‐guidance performance of a preclinical microirradiator. https://app.cochat.ai/papers/10.1118/1.3533947
- The paper
- Clarkson, R., Lindsay, P. E., Ansell, S., Wilson, G., Jelveh, S., Hill, R. P., & Jaffray, D. A. (2011). Characterization of image quality and image‐guidance performance of a preclinical microirradiator. Medical Physics. https://doi.org/10.1118/1.3533947
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.