Characterization of image quality and image‐guidance performance of a preclinical microirradiator

Open access
Clarkson, R., Lindsay, P. E., Ansell, S., Wilson, G., Jelveh, S., Hill, R. P., Jaffray, D. A. · 2011 · Medical Physics
Cited by 172

The DOI matches its Crossref record; 10 of the 10 references it cites were found in a registry or an index.

Abstract

Purpose: To assess image quality and image‐guidance capabilities of a cone‐beam CT based small‐animal image‐guided irradiation unit (micro‐IGRT). Methods: A micro‐IGRT system has been developed in collaboration with the authors' laboratory as a means to study the radiobiological effects of conformal radiation dose distributions in small animals. The system, the X‐Rad 225Cx, consists of a 225 kVp x‐ray tube and a flat‐panel amorphous silicon detector mounted on a rotational C‐arm gantry and is capable of both fluoroscopic x‐ray and cone‐beam CT imaging, as well as image‐guided placement of the radiation beams. Image quality (voxel noise, modulation transfer, CT number accuracy, and geometric accuracy characteristics) was assessed using water cylinder and micro‐CT test phantoms. Image guidance was tested by analyzing the dose delivered to radiochromic films fixed to BB's through the end‐to‐end process of imaging, targeting the center of the BB, and irradiation of the film/BB in order to compare the offset between the center of the field and the center of the BB. Image quality and geometric studies were repeated over a 5–7 month period to assess stability. Results: CT numbers reported were found to be linear and the noise for images of homogeneous water phantom was 30 HU at imaging doses of approximately 1 cGy (to water). The presampled MTF at 50% and 10% reached 0.64 and , respectively. Targeting accuracy by means of film irradiations was shown to have a mean displacement error of , with standard deviations of [0.02, 0.20, 0.17] mm. The system has proven to be stable over time, with both the image quality and image‐guidance performance being reproducible for the duration of the studies. Conclusions: The micro‐IGRT unit provides soft‐tissue imaging of small‐animal anatomy at acceptable imaging doses. The geometric accuracy and targeting systems permit dose placement with submillimeter accuracy and precision. The system has proven itself to be stable over 2 yr of routine laboratory use and provides a platform for the exploration of targeted radiation effects in small‐animal models.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1118/1.3533947

Cites

10 of 10 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 10 references

Cited by

See all 172

Metadata

Authors
Clarkson, R., Lindsay, P. E., Ansell, S., Wilson, G., Jelveh, S., Hill, R. P., & Jaffray, D. A.
Venue
Medical Physics
Publisher
Wiley
Published
2011-01-21
Volume
38
Issue
2
Pages
845-856
ISSN
  • 0094-2405 (ISSN-L)
  • 2473-4209
OpenAlex
W2002760403
Cited by
172
References
10
FWCI
13.46
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Characterization of image quality and image‐guidance performance of a preclinical microirradiator. https://app.cochat.ai/papers/10.1118/1.3533947
The paper
Clarkson, R., Lindsay, P. E., Ansell, S., Wilson, G., Jelveh, S., Hill, R. P., & Jaffray, D. A. (2011). Characterization of image quality and image‐guidance performance of a preclinical microirradiator. Medical Physics. https://doi.org/10.1118/1.3533947

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.