Improved kinetic analysis of dynamic PET data with optimized HYPR‐LR

Open access
Floberg, John M., Mistretta, Charles A., Weichert, Jamey P., Hall, Lance T., Holden, James E., Christian, Bradley T. · 2012 · Medical Physics
Cited by 45

The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.

Abstract

Purpose: Highly constrained backprojection‐local reconstruction (HYPR‐LR) has made a dramatic impact on magnetic resonance angiography (MRA) and shows promise for positron emission tomography (PET) because of the improvements in the signal‐to‐noise ratio (SNR) it provides dynamic images. For PET in particular, HYPR‐LR could improve kinetic analysis methods that are sensitive to noise. In this work, the authors closely examine the performance of HYPR‐LR in the context of kinetic analysis, they develop an implementation of the algorithm that can be tailored to specific PET imaging tasks to minimize bias and maximize improvement in variance, and they provide a framework for validating the use of HYPR‐LR processing for a particular imaging task. Methods: HYPR‐LR can introduce errors into non sparse PET studies that might bias kinetic parameter estimates. An implementation of HYPR‐LR is proposed that uses multiple temporally summed composite images that are formed based on the kinetics of the tracer being studied (HYPR‐LR‐MC). The effects of HYPR‐LR‐MC and of HYPR‐LR using a full composite formed with all the frames in the study (HYPR‐LR‐FC) on the kinetic analysis of Pittsburgh compound‐B ([11C]‐PIB) are studied. HYPR‐LR processing is compared to spatial smoothing. HYPR‐LR processing was evaluated using both simulated and human studies. Nondisplaceable binding potential (BP ND) parametric images were generated from fifty noise realizations of the same numerical phantom and eight [11 C]‐PIB positive human scans before and after HYPR‐LR processing or smoothing using the reference region Logan graphical method and receptor parametric mapping (RPM2). The bias and coefficient of variation in the frontal and parietal cortex in the simulated parametric images were calculated to evaluate the absolute performance of HYPR‐LR processing. Bias in the human data was evaluated by comparing parametric image BP ND values averaged over large regions of interest (ROIs) to Logan estimates of the BP ND from TACs averaged over the same ROIs. Variance was assessed qualitatively in the parametric images and semiquantitatively by studying the correlation between voxel BP ND estimates from Logan analysis and RPM2. Results: Both the simulated and human data show that HYPR‐LR‐FC overestimates BP ND values in regions of high [11 C]‐PIB uptake. HYPR‐LR‐MC virtually eliminates this bias. Both implementations of HYPR‐LR reduce variance in the parametric images generated with both Logan analysis and RPM2, and HYPR‐LR‐FC provides a greater reduction in variance. This reduction in variance nearly eliminates the noise‐dependent Logan bias. The variance reduction is greater for the Logan method, particularly for HYPR‐LR‐MC, and the variance in the resulting Logan images is comparable to that in the RPM2 images. HYPR‐LR processing compares favorably with spatial smoothing, particularly when the data are analyzed with the Logan method, as it provides a reduction in variance with no loss of spatial resolution. Conclusions: : HYPR‐LR processing shows significant potential for reducing variance in parametric images, and can eliminate the noise‐dependent Logan bias. HYPR‐LR‐FC processing provides the greatest reduction in variance but introduces a positive bias into the BP ND of high‐uptake border regions. The proposed method for forming HYPR composite images, HYPR‐LR‐MC, eliminates this bias at the cost of less variance reduction.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1118/1.4718669

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 1 references

Cited by

See all 45

Metadata

Authors
Floberg, J. M., Mistretta, C. A., Weichert, J. P., Hall, L. T., Holden, J. E., & Christian, B. T.
Venue
Medical Physics
Publisher
Wiley
Published
2012-05-22
Volume
39
Issue
6Part1
Pages
3319-3331
ISSN
  • 0094-2405 (ISSN-L)
  • 2473-4209
OpenAlex
W2042277982
Cited by
45
References
1
FWCI
2.00
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Improved kinetic analysis of dynamic PET data with optimized HYPR‐LR. https://app.cochat.ai/papers/10.1118/1.4718669
The paper
Floberg, J. M., Mistretta, C. A., Weichert, J. P., Hall, L. T., Holden, J. E., & Christian, B. T. (2012). Improved kinetic analysis of dynamic PET data with optimized HYPR‐LR. Medical Physics. https://doi.org/10.1118/1.4718669

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.