Tungsten anode spectral model using interpolating cubic splines: Unfiltered x‐ray spectra from 20 kV to 640 kV
Open accessThe DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
Purpose: Monte Carlo methods were used to generate lightly filtered high resolution x‐ray spectra spanning from 20 kV to 640 kV. Methods: X‐ray spectra were simulated for a conventional tungsten anode. The Monte Carlo N‐Particle eXtended radiation transport code (MCNPX 2.6.0) was used to produce 35 spectra over the tube potential range from 20 kV to 640 kV, and cubic spline interpolation procedures were used to create piecewise polynomials characterizing the photon fluence per energy bin as a function of x‐ray tube potential. Using these basis spectra and the cubic spline interpolation, 621 spectra were generated at 1 kV intervals from 20 to 640 kV. The tungsten anode spectral model using interpolating cubic splines (TASMICS) produces minimally filtered (0.8 mm Be) x‐ray spectra with 1 keV energy resolution. The TASMICS spectra were compared mathematically with other, previously reported spectra. Results: Using paired t‐test analyses, no statistically significant difference (i.e., p > 0.05) was observed between compared spectra over energy bins above 1% of peak bremsstrahlung fluence. For all energy bins, the correlation of determination (R2) demonstrated good correlation for all spectral comparisons. The mean overall difference (MOD) and mean absolute difference (MAD) were computed over energy bins (above 1% of peak bremsstrahlung fluence) and over all the kV permutations compared. MOD and MAD comparisons with previously reported spectra were 2.7% and 9.7%, respectively (TASMIP), 0.1% and 12.0%, respectively [R. Birch and M. Marshall, “Computation of bremsstrahlung x‐ray spectra and comparison with spectra measured with a Ge(Li) detector,” Phys. Med. Biol. 24, – (1979)], 0.4% and 8.1%, respectively (Poludniowski), and 0.4% and 8.1%, respectively (AAPM TG 195). The effective energy of TASMICS spectra with 2.5 mm of added Al filtration ranged from 17 keV (at 20 kV) to 138 keV (at 640 kV); with 0.2 mm of added Cu filtration the effective energy was 9 keV at 20 kV and 169 keV at 640 kV. Conclusions: Ranging from 20 kV to 640 kV, 621 x‐ray spectra were produced and are available at 1 kV tube potential intervals. The spectra are tabulated at 1 keV intervals. TASMICS spectra were shown to be largely equivalent to published spectral models and are available in spreadsheet format for interested users by emailing the corresponding author (JMB).
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Hubbell, J. H., & Seltzer, S. M. (1995) · doi:10.6028/nist.ir.5632 · Cited by 2,033
Cited by
Virtual clinical trials in medical imaging: a review
Abadi, E., Segars, W. P., Tsui, B. M. W., Kinahan, P. E., Bottenus, N., Frangi, A. F., Maidment, A., Lo, J., & Samei, E. (2020) · Journal of Medical Imaging · doi:10.1117/1.jmi.7.4.042805 · Cited by 224
Technical Note: SpekPy v2.0—a software toolkit for modeling x‐ray tube spectra
Poludniowski, G., Omar, A., Bujila, R., & Andreo, P. (2021) · Medical Physics · doi:10.1002/mp.14945 · Cited by 203
Tudda, A., Donzelli, E., Nicolini, G., Semperboni, S., Bossi, M., Cavaletti, G., Castriconi, R., Mangili, P., Vecchio, A. D., Sarno, A., et al. (2022) · Medical Physics · doi:10.1002/mp.15348 · Cited by 34
Metadata
- Authors
- Hernandez, A. M., & Boone, J. M.
- Venue
- Medical Physics
- Publisher
- Wiley
- Published
- 2014-03-07
- Volume
- 41
- Issue
- 4
- Pages
- 042101-042101
- ISSN
- 0094-2405 (ISSN-L)
- 2473-4209
- OpenAlex
- W2095036764
- Cited by
- 182
- References
- 1
- FWCI
- 9.69
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Tungsten anode spectral model using interpolating cubic splines: Unfiltered x‐ray spectra from 20 kV to 640 kV. https://app.cochat.ai/papers/10.1118/1.4866216
- The paper
- Hernandez, A. M., & Boone, J. M. (2014). Tungsten anode spectral model using interpolating cubic splines: Unfiltered x‐ray spectra from 20 kV to 640 kV. Medical Physics. https://doi.org/10.1118/1.4866216
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.