An accurate method for computer‐generating tungsten anode x‐ray spectra from 30 to 140 kV
Abstract
A tungsten anode spectral model using interpolating polynomials (TASMIP) was used to compute x‐ray spectra at 1 keV intervals over the range from 30 kV to 140 kV. The TASMIP is not semi‐empirical and uses no physical assumptions regarding x‐ray production, but rather interpolates measured constant potential x‐ray spectra published by Fewell et al. [Handbook of Computed Tomography X‐ray Spectra (U.S. Government Printing Office, Washington, D.C., 1981)]. X‐ray output measurements (mR/mAs measured at 1 m) were made on a calibrated constant potential generator in our laboratory from 50 kV to 124 kV, and with 0–5 mm added aluminum filtration. The Fewell spectra were slightly modified (numerically hardened) and normalized based on the attenuation and output characteristics of a constant potential generator and metal‐insert x‐ray tube in our laboratory. Then, using the modified Fewell spectra of different kVs, the photon fluence Φ at each 1 keV energy bin over energies from 10 keV to 140 keV was characterized using polynomial functions of the form A total of 131 polynomial functions were used to calculate accurate x‐ray spectra, each function requiring between two and four terms. The resulting TASMIP algorithm produced x‐ray spectra that match both the quality and quantity characteristics of the x‐ray system in our laboratory. For photon fluences above 10% of the peak fluence in the spectrum, the average percent difference (and standard deviation) between the modified Fewell spectra and the TASMIP photon fluence was −1.43% (3.8%) for the 50 kV spectrum, −0.89% (1.37%) for the 70 kV spectrum, and for the 80, 90, 100, 110, 120, 130 and 140 kV spectra, the mean differences between spectra were all less than 0.20% and the standard deviations were less than ∼1.1%. The model was also extended to include the effects of generator‐induced kV ripple. Finally, the x‐ray photon fluence in the units of per mR was calculated as a function of HVL, kV, and ripple factor, for various (water‐equivalent) patient thicknesses (0, 10, 20, and 30 cm). These values may be useful for computing the detective quantum efficiency, of x‐ray detector systems. The TASMIP algorithm and ancillary data are made available on line at http://www.aip.org/epaps/epaps.html.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Sisniega, A., Zbijewski, W., Badal, A., Kyprianou, I. S., Stayman, J. W., Vaquero, J. J., & Siewerdsen, J. H. (2013) · Medical Physics · doi:10.1118/1.4801895 · Cited by 117
Determination of the weighted CT dose index in modern multi-detector CT scanners
Perisinakis, K., Damilakis, J., Tzedakis, A., Papadakis, A., Theocharopoulos, N., & Gourtsoyiannis, N. (2007) · Physics in Medicine and Biology · doi:10.1088/0031-9155/52/21/010 · Cited by 47
Metadata
- Authors
- Boone, J. M., & Seibert, J. A.
- Venue
- Medical Physics
- Publisher
- Wiley
- Published
- 1997-11
- Volume
- 24
- Issue
- 11
- Pages
- 1661-1670
- ISSN
- 0094-2405 (ISSN-L)
- 2473-4209
- OpenAlex
- W1979528439
- Cited by
- 706
- References
- 0
- FWCI
- 5.67
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: An accurate method for computer‐generating tungsten anode x‐ray spectra from 30 to 140 kV. https://app.cochat.ai/papers/10.1118/1.597953
- The paper
- Boone, J. M., & Seibert, J. A. (1997). An accurate method for computer‐generating tungsten anode x‐ray spectra from 30 to 140 kV. Medical Physics. https://doi.org/10.1118/1.597953
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.