Horvath, Lynn L., George, Benjamin J., Murray, Clinton K., Harrison, Linda S., Hospenthal, Duane R. · 2004 · Journal of Clinical Microbiology
Cited by 136

The DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.

Abstract

A direct comparison of two automated blood culture systems was conducted to compare their ability to detect Candida growth. The systems evaluated were the BACTEC 9240 (Bactec) and BacT/ALERT 3D (BacT). The aerobic, anaerobic, and mycology media for each system were evaluated: Bactec Plus Aerobic/F, Plus Anaerobic/F, and Myco/F Lytic bottles,, respectively, and BacT FA, SN, and MB bottles, respectively. Each blood culture bottle was inoculated with fresh blood from healthy donors. Fifty isolates of Candida spp. were used. The six different blood culture bottles were each inoculated with 1,000 yeasts per bottle and then incubated in the corresponding automated system. The BacT detected growth of 90% (135 of 150) of Candida pathogens, while Bactec detected 66% (100 of 150). Growth was detected in all BacT and Bactec mycology bottles, all BacT aerobic bottles, and by terminal subculture of all bottles. Sixty-five of 300 (22%) bottles had no growth detected; 50 from the Bactec (5 aerobic and 45 anaerobic) and 15 from the BacT (all anaerobic). Terminal subculture of “negative” bottles demonstrated viable yeast growth from all 65 bottles, representing 65 false-negatives. The mean time to growth detection in the BacT system was 25.62 h while the Bactec was 27.30 h ( P < 0.01). Both automated blood culture systems detected all episodes of simulated candidemia when specialized mycology media were used. However, when only standard aerobic and anaerobic media were used, the BacT performed better than the Bactec in overall growth detection, time to growth detection, and number of false-negatives.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1128/jcm.42.1.115-118.2004

Cites

4 of 4 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 4 references

Cited by

See all 136

Metadata

Authors
Horvath, L. L., George, B. J., Murray, C. K., Harrison, L. S., & Hospenthal, D. R.
Venue
Journal of Clinical Microbiology
Published
2004-01
Volume
42
Issue
1
Pages
115-118
ISSN
  • 0095-1137 (ISSN-L)
  • 1098-660X
OpenAlex
W2095629741
Cited by
136
References
4
FWCI
3.28
Type
journal article
Language
en
Open access
green

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Direct Comparison of the BACTEC 9240 and BacT/ALERT 3D Automated Blood Culture Systems for Candida Growth Detection. https://app.cochat.ai/papers/10.1128/jcm.42.1.115-118.2004
The paper
Horvath, L. L., George, B. J., Murray, C. K., Harrison, L. S., & Hospenthal, D. R. (2004). Direct Comparison of the BACTEC 9240 and BacT/ALERT 3D Automated Blood Culture Systems for Candida Growth Detection. Journal of Clinical Microbiology. https://doi.org/10.1128/jcm.42.1.115-118.2004

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.