Isotopic (δ13C, δ18O) Analysis of Small Amounts of Carbonate in Silicate Rocks by the Continuous Flow Isotope Ratio Mass Spectrometry Method

Open access
Dubinina, E. O., Chizhova, Y. N., Kossova, S. A. · 2023 · Geochemistry International

The DOI matches its Crossref record; 38 of the 38 references it cites were found in a registry or an index; 1 of the 37 sources its claims cite were checked for support.

Abstract

An experimental study of the main factors affecting the accuracy of oxygen and carbon isotopic analysis in carbonates dispersed in silicate matrix is carried out. Artificial 1, 2, 5, and 10% mixtures of quartz with carbonates with different isotopic parameters (KH-2, Ko, MCA-8) were analyzed by continuous flow isotope ratio mass spectrometry (CF IRMS). It is established that, in addition to the influence of the instrumental nonlinearity, the results are affected by two factors: trace amounts of CO 2 , constantly present in the system (the blank effect) and the presence of chemically neutral silicate particles (the matrix effect). The blank effect depends on the isotopic parameters of the sample and has very little influence on the estimated carbonate content in the rock. The matrix effect, on the contrary, strongly affects the estimated carbonate content, and produces the isotopic shift towards underestimated contents of heavy 13 C and 18 O isotopes. It is shown that this effect is related to the processes occurring near the CO 2 –acid–quartz interface, which are accompanied by kinetic fractionation of carbon and oxygen isotopes. Both effects are dependent on the amount of silicate matrix in the system and most clearly manifested during analysis of carbonate-poor rocks. When the carbonate content in the rock is about 1–2%, deviations from the true δ 13 C and δ 18 O values can reach the first ppm, while carbonate content obtained by chromatographic peak calibration can be underestimated by 20–40%.

Overview

CoChat Summary

## TL;DR

This experimental paper tests how well continuous-flow IRMS measures δ13C and δ18O in carbonate-poor silicate rocks by analyzing artificial quartz–carbonate mixtures with 1, 2, 5, and 10 wt % carbonate. It finds that measurement errors come from three sources—instrumental nonlinearity, a blank effect, and a silicate-matrix effect—with the matrix effect causing the largest bias in low-carbonate samples.

## Key findings

  • Artificial quartz–carbonate mixtures were prepared from three carbonate standards (KH-2, Ko, MSA-8) at 1, 2, 5, and 10 wt % carbonate and measured by CF IRMS; pure carbonate standards were used for calibration.
  • The blank effect was estimated at about 1–3% of the total analyzed gas, corresponding to approximately γ(b) ≈ 0.005 μmol CO2, and was judged to have little influence on carbonate-content estimates but to distort isotope values at low gas amounts.
  • Instrumental nonlinearity caused only small shifts: about 0.2–0.4‰ for δ13C and up to 0.6–0.7‰ for δ18O in the region of extremely low peak areas.
  • The silicate-matrix effect strongly affected carbonate-content estimates: carbonate contents less than 10% and 5% were expected to be underestimated by 20% and 30–40%, respectively; at about 1–2% carbonate, δ13C and δ18O deviations could reach the first ppm, and carbonate content from chromatographic peak area could be underestimated by up to 40%.

## Methods

  • The study used artificial silicate-carbonate mixtures made from Polaris quartz and three carbonate reference materials (KH-2 limestone, Ko calcite, MSA-8 calcite) at 1, 2, 5, and 10 wt % carbonate; pure carbonates were analyzed in the same measurement series.
  • Samples were ground, dried, baked in borosilicate vials, reacted with orthophosphoric acid in a GasBench II–Delta V+ CF IRMS system, and the fourth chromatographic peak area was used to estimate CO2 amount.
  • The authors separately assessed instrumental nonlinearity with repeated portions of the same CO2 gas, checked blank contributions with empty vials, acid-only vials, and pure quartz, and compared mixtures based on different carbonates to distinguish blank from matrix effects.

## Limitations

  • The paper states that matrix effects can only be accounted for if the exact mechanism of interaction between the matrix surface and CO2 is understood.
  • It also states that further experiments, including studies of different matrix materials, are required; this is an explicit limitation of the current study.
  • Evident from the text: the work uses artificial quartz-based mixtures rather than natural rocks, so the reported corrections and thresholds are demonstrated for these experimental systems rather than directly for all silicate rocks.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1134/s0016702923110034

Cites

38 of 38 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Figures

Fig. 1 Photos of starting materials: (a) calcite MSA-8 calcite, (b) KH-2 limestone, (c) Ko calcite, and (d) Polaris v.
Fig. 1 Photos of starting materials: (a) calcite MSA-8 calcite, (b) KH-2 limestone, (c) Ko calcite, and (d) Polaris v.From the PDF
Fig. 2 Area of the fourth chromatographic sample peak (S, Vs) and expected amount of analyzed gas CO 2 (0), μmol, for a mixture based on (a) KH-2, (b) Ko, and (c) MSA-8. Trend lines plotted for the fourth peak of pure carbonate samples were used to calculate in silicate-carbonate mixtures. 2 СО *
Fig. 2 Area of the fourth chromatographic sample peak (S, Vs) and expected amount of analyzed gas CO 2 (0), μmol, for a mixture based on (a) KH-2, (b) Ko, and (c) MSA-8. Trend lines plotted for the fourth peak of pure carbonate samples were used to calculate in silicate-carbonate mixtures. 2 СО *From the PDF
Fig. 3 Losses of analyzed gaseous CO 2 as a function of carbonate content in silicate-carbonate mixtures: (a) absolute and (b) relative. Relative losses were calculated using Eq. (2).
Fig. 3 Losses of analyzed gaseous CO 2 as a function of carbonate content in silicate-carbonate mixtures: (a) absolute and (b) relative. Relative losses were calculated using Eq. (2).From the PDF
Fig. 4 Deviation of (a) δ 13 C and (b) δ 18 O measured in silicate-carbonate mixtures from the true values depending on the amount of analyzed gas. Dotted line is confidence interval (2 σ ).
Fig. 4 Deviation of (a) δ 13 C and (b) δ 18 O measured in silicate-carbonate mixtures from the true values depending on the amount of analyzed gas. Dotted line is confidence interval (2 σ ).From the PDF
Fig. 5 Deviation of (a) δ 13 C and (b) δ 18 O measured in silicate-carbonate mixtures from the true values depending on the absolute content of CaCO 3 in the silicate-carbonate mixture. Data from the present work and experimental results from (Zha et al., 2010) are presented. Vertical dashed line is amount of CaCO 3 (70 μg) recommended by Zha et al. (2010) as a threshold amount for isotopic analysis of silicate-carbonate mixtures.
Fig. 5 Deviation of (a) δ 13 C and (b) δ 18 O measured in silicate-carbonate mixtures from the true values depending on the absolute content of CaCO 3 in the silicate-carbonate mixture. Data from the present work and experimental results from (Zha et al., 2010) are presented. Vertical dashed line is amount of CaCO 3 (70 μg) recommended by Zha et al. (2010) as a threshold amount for isotopic analysis of silicate-carbonate mixtures.From the PDF
Fig. 6 Deviations of 13 C values from the true value as a function of peak area: ( 1 ) instrumental nonlinearity determined by the prolonged experiment; ( 2 , 3 ) data for silicate-carbonate mixtures prepared on the basis of KH-2 and MCA-8, respectively.
Fig. 6 Deviations of 13 C values from the true value as a function of peak area: ( 1 ) instrumental nonlinearity determined by the prolonged experiment; ( 2 , 3 ) data for silicate-carbonate mixtures prepared on the basis of KH-2 and MCA-8, respectively.From the PDF
Fig. 7 Calculated deviation of δ 13 C and δ 18 O in the gas lost due to the matrix effect from the true values.
Fig. 7 Calculated deviation of δ 13 C and δ 18 O in the gas lost due to the matrix effect from the true values.From the PDF
Fig. 8 Experimental results on the effect of the volume of orthophosphoric acid on the amount of gas released from pure carbonates: ( 1 ) peak areas of samples with 0.05 mL of acid added; ( 2 ) peak areas of samples with 0.2 mL of acid added.
Fig. 8 Experimental results on the effect of the volume of orthophosphoric acid on the amount of gas released from pure carbonates: ( 1 ) peak areas of samples with 0.05 mL of acid added; ( 2 ) peak areas of samples with 0.2 mL of acid added.From the PDF
Fig. 9 Calculated line of isotopic parameter alteration of KH-2 carbonate during its decomposition with CO 2 removal at 600 ° . The fractionation factors of carbon and oxygen isotopes in the CO 2 -CaCO 3 system are taken from Bottinga (1968) and (Chacko and Deines, 2008). Numbers near the line are the fraction of remaining carbonate after partial decomposition (% of the initial amount). Dots are the measured carbon- and oxygen-isotopic compositions of samples of silicate-carbonate mixtures prepared on the basis of KH-2.
Fig. 9 Calculated line of isotopic parameter alteration of KH-2 carbonate during its decomposition with CO 2 removal at 600 ° . The fractionation factors of carbon and oxygen isotopes in the CO 2 -CaCO 3 system are taken from Bottinga (1968) and (Chacko and Deines, 2008). Numbers near the line are the fraction of remaining carbonate after partial decomposition (% of the initial amount). Dots are the measured carbon- and oxygen-isotopic compositions of samples of silicate-carbonate mixtures prepared on the basis of KH-2.From the PDF

Verification

Claims the paper makes

0 Valid · 2 Not valid · 21 Not checked · 80 Own results and interpretations

47 more claims

Create a free account to see all 103 claims and to verify the 2 references still open.

See all claims — free account
  • Borderline: one of the tests for a substantial claim is unclear. 1
  • It may state two separate claims as one. 41
  • The quote only partly states this claim. 2
  • Probably not a substantial claim. 15

Create a free account to see these claims.

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

Each contradiction between two passages listed here was confirmed by an independent check.

1 of 1

Citations

Cites

Show all 38 references

Research works printed but not linked: 1

  1. D. Yong and H. Song, “Refined protocol for the δ13C analysis of calcite and dolomite in carbonate mixture samples,” Rapid Commun Mass Spectrom. 34 (10), e8743 (2020). https://doi.org/10.1002/rcm.8743 · DOI

Cited by

See all 1

Metadata

Authors
Dubinina, E. O., Chizhova, Y. N., & Kossova, S. A.
Venue
Geochemistry International
Published
2023-11
Volume
61
Issue
11
Pages
1164-1176
ISSN
  • 0016-7029 (ISSN-L)
  • 1556-1968
OpenAlex
W4388706101
Cited by
1
References
38
Citations by year
2024: 1
FWCI
0.17
Type
journal article
Language
en
Open access
Licence
Locations
Geochemistry International (publishedVersion, cc-by)
Funders
Russian Academy of Sciences
Topics
  • Physical Sciences › Earth and Planetary Sciences › Geochemistry and Petrology › Geological and Geochemical Analysis
  • Physical Sciences › Earth and Planetary Sciences › Geochemistry and Petrology › Geochemistry and Elemental Analysis
  • Physical Sciences › Chemistry › Analytical Chemistry › Analytical chemistry methods development
Keywords
carbonate isotopes, carbon-13, oxygen-18, silicate rocks, matrix effects, kinetic isotope fractionation, quartz, continuous-flow isotope ratio mass spectrometry, analytical accuracy

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

The claim verdicts this report counts are experimental: their accuracy is not yet measured.

A language model drew the claims from the paper on 2026-09-30, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).

A second model labelled each claim on 2026-09-30 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).

Cited sources are checked with a language model’s help (gpt-6-sol).

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Isotopic (δ13C, δ18O) Analysis of Small Amounts of Carbonate in Silicate Rocks by the Continuous Flow Isotope Ratio Mass Spectrometry Method. https://app.cochat.ai/papers/10.1134/s0016702923110034
The paper
Dubinina, E. O., Chizhova, Y. N., & Kossova, S. A. (2023). Isotopic (δ13C, δ18O) Analysis of Small Amounts of Carbonate in Silicate Rocks by the Continuous Flow Isotope Ratio Mass Spectrometry Method. Geochemistry International. https://doi.org/10.1134/s0016702923110034

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.