Evaluation of “point of care” devices in the measurement of low blood glucose in neonatal practice
Open accessThe DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.
Abstract
Background: Low blood glucose in newborns is difficult to detect clinically. Hence a reliable “point of care” device (glucometer) for early detection and treatment of low glucose is needed. Objective: To evaluate the performance of five readily available glucometers for the detection of low blood glucose in newborn infants. Method: Glucostix measurements were taken for newborns with risk factors using a Reflolux S (Boehringer) glucometer. If the initial reading was low (< 2.6 mmol/l), further measurements were taken with two other glucometers (phase I, Advantage and Glucotrend (Roche); phase II, Elite XL (Bayer) and Precision (Abbott)), and plasma glucose was measured in the laboratory (Aeroset; Abbott). Results: Over 10 months, 101 specimens were collected from 71 newborns (57 in phase I; 44 in phase II). The Advantage glucometer usually overestimated blood glucose with a mean difference of 1.07 mmol/l (p < 0.01) at all low glucose ranges. The Glucotrend, Precision, and Elite XL glucometers performed better; the mean differences were not significantly different from the laboratory measured value (0.17 mmol/l (p = 0.37); −0.12 mmol/l (p = 0.13), and 0.24 mmol/l (p = 0.13) respectively). For detection of glucose concentrations < 2.6 mmol/l, the Precision glucometer had the highest sensitivity (96.4%) and negative predictive value (90%). For lower glucose concentrations (< 2.0 mmol/l), the Glucotrend glucometer performed even better (sensitivity 92.3%, negative predictive value 96.3%). Conclusion: Point of care devices should have good precision in the low glucose concentration range, sensitivity, and accuracy for early detection of neonatal hypoglycaemia. None of the five glucometers was satisfactory as the sole measuring device. The Glucotrend and Precision glucometers have the greatest sensitivity and negative predictive value. However, confirmation with laboratory measurements of plasma glucose and clinical assessment are still of the utmost importance.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Statistical methods for assessing agreement between two methods of clinical measurement
Bland, J. M., & Altman, D. G. (2010) · International Journal of Nursing Studies · doi:10.1016/j.ijnurstu.2009.10.001 · Cited by 5,220
Self-Monitoring of Blood Glucose
American Diabetes Association (1994) · Diabetes Care · doi:10.2337/diacare.17.1.81 · Cited by 255
Cited by
Assessing the quality of glucose monitor studies: a critical evaluation of published reports
Mahoney, J., & Ellison, J. (2007) · Clinical Chemistry · doi:10.1373/clinchem.2006.083493 · Cited by 48
“…glucose[Text Word]) AND (performance[Text Word] OR evaluation[Text Word] OR accurate[Text Word] OR accuracy[Text Word] OR point-of-care[Text Word] OR meter[Text Word] OR meters[Text Word] OR glucometer[Text Word] OR glucometers[Text Word] OR monitor[Text Word] OR monitors[Text Word]) AND…”
BackgroundBalion, C., Grey, V., Ismaila, A., Blatz, S., & Seidlitz, W. (2006) · BMC Pediatrics · doi:10.1186/1471-2431-6-28 · Cited by 46
“Few studies reporting the performance of glucose meters in the NICU have been able to achieve any of the above criteria or other reasonable accuracy criteria [8-10].”
Wight, N., Marinelli, K. A., & The Academy Of Breastfeeding Medicine Protocol Committee (2006) · Breastfeeding Medicine · doi:10.1089/bfm.2006.1.178 · Cited by 43
“14,26,32 Bedside glucose tests may be used for screening, but laboratory levels must confirm results before a diagnosis of hypoglycemia can be made, especially in asymptomatic infants.”
Metadata
- Authors
- Ho, H. T., Yeung, W. K. Y., & Young, B. W. Y.
- Venue
- Archives of Disease in Childhood - Fetal and Neonatal Edition
- Publisher
- BMJ
- Published
- 2004-06-21
- Volume
- 89
- Issue
- 4
- Pages
- F356-F359
- ISSN
- 1359-2998 (ISSN-L)
- 1468-2052
- OpenAlex
- W2094524489
- Cited by
- 99
- References
- 3
- FWCI
- 1.90
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.