Comparison of performance of various sphygmomanometers with intra-arterial blood-pressure readings.
Open accessThe DOI matches its Crossref record.
Abstract
Seven types of sphygmomanometer were used in random order on each of nine hypertensive patients and the readings compared with simultaneous intra-arterial blood-pressure recordings. All the devices gave significantly different values for systolic pressure, and only two measured diastolic pressure without significant error. Systolic pressure was consistently underestimated (range 31-7 mm Hg), and all but one instrument overestimated diastolic pressure (range 10-2 mm Hg). The variability of readings was least with the standard mercury sphygmomanometer and the random-zero machine, while with some of the more automated devices single readings were in error up to -68/33 mm Hg. The strong correlations found between intra-arterial and cuff systolic pressures with all devices tested and significant correlations for diastolic pressure with all but one device indicate that, with one possible exception, the sphygmomanometers would give accurate results where a change in blood pressure was the main concern.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · Circulation · doi:10.1161/01.cir.0000154900.76284.f6 · Cited by 3,985
Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · Hypertension · doi:10.1161/01.hyp.0000150859.47929.8e · Cited by 2,051
Prognostic Value of Ambulatory Blood Pressure: Current Evidence and Clinical Implications
Verdecchia, P. (2000) · Hypertension · doi:10.1161/01.hyp.35.3.844 · Cited by 621
Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J. W., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · The Journal of Clinical Hypertension · doi:10.1111/j.1524-6175.2005.04377.x · Cited by 436
Avolio, A. P., Butlin, M., & Walsh, A. (2010) · Physiological Measurement · doi:10.1088/0967-3334/31/1/r01 · Cited by 310
ERROR IN BLOOD-PRESSURE MEASUREMENT DUE TO INCORRECT CUFF SIZE IN OBESE PATIENTS
Maxwell, M., Schroth, P., Waks, A., Karam, M., & Dornfeld, L. (1982) · The Lancet · doi:10.1016/s0140-6736(82)91163-1 · Cited by 307
Cardiovascular Outcome in White-Coat Versus Sustained Mild Hypertension: A 10-Year Follow-Up Study
Khattar, R. S., Senior, R., & Lahiri, A. (1998) · Circulation · doi:10.1161/01.cir.98.18.1892 · Cited by 299
Accuracy of Cuff-Measured Blood Pressure
Picone, D. S., Schultz, M. G., Otahal, P., Aakhus, S., Al-Jumaily, A. M., Black, J. A., Bos, W. J., Chambers, J. B., Chen, C.-H., Cheng, H.-M., et al. (2017) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2017.05.064 · Cited by 277
Sharman, J. E., Avolio, A. P., Baulmann, J., Benetos, A., Blacher, J., Blizzard, C. L., Boutouyrie, P., Chen, C.-H., Chowienczyk, P., Cockcroft, J. R., et al. (2017) · European Heart Journal · doi:10.1093/eurheartj/ehw632 · Cited by 231
“36 Therefore, calibrating the waveform to brachial cuff SBP and DBP is likely to result in underestimation of central SBP and central pulse pressure, but overestimation of central DBP relative to intra-arterial central DBP. 40 The magnitude of calibration-induced error may often exceed 10mmHg for…”
BackgroundA review of flexible force sensors for human health monitoring
Cheng, M., Zhu, G., Zhang, F., Tang, W.-L., Jianping, S., Yang, J.-Q., & Zhu, L.-Y. (2020) · Journal of Advanced Research · doi:10.1016/j.jare.2020.07.001 · Cited by 203
Continuous finger arterial pressure: Utility in the cardiovascular laboratory
Imholz, B. P. M., Wieling, W., Langewouters, G. J., & van Montfrans, G. A. (1991) · Clinical Autonomic Research · doi:10.1007/bf01826057 · Cited by 181
A guide to uncertainty quantification and sensitivity analysis for cardiovascular applications
Eck, V. G., Donders, W. P., Sturdy, J., Feinberg, J., Delhaas, T., Hellevik, L. R., & Huberts, W. (2016) · International Journal for Numerical Methods in Biomedical Engineering · doi:10.1002/cnm.2755 · Cited by 163
Metadata
- Authors
- Hunyor, S. N., Flynn, J. M., & Cochineas, C.
- Venue
- BMJ
- Publisher
- BMJ
- Published
- 1978-07-15
- Volume
- 2
- Issue
- 6131
- Pages
- 159-162
- ISSN
- 0007-1447 (ISSN-L)
- 1468-5833
- OpenAlex
- W1972409283
- Cited by
- 122
- References
- 0
- FWCI
- 6.07
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.