Assessment of the feasibility of an ultra-low power, wireless digital patch for the continuous ambulatory monitoring of vital signs
Open accessThe DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Abstract
Background and objectives: Vital signs are usually recorded at 4–8 h intervals in hospital patients, and deterioration between measurements can have serious consequences. The primary study objective was to assess agreement between a new ultra-low power, wireless and wearable surveillance system for continuous ambulatory monitoring of vital signs and a widely used clinical vital signs monitor. The secondary objective was to examine the system's ability to automatically identify and reject invalid physiological data. Setting: Single hospital centre. Participants: Heart and respiratory rate were recorded over 2 h in 20 patients undergoing elective surgery and a second group of 41 patients with comorbid conditions, in the general ward. Outcome measures: Primary outcome measures were limits of agreement and bias. The secondary outcome measure was proportion of data rejected. Results: The digital patch provided reliable heart rate values in the majority of patients (about 80%) with normal sinus rhythm, and in the presence of abnormal ECG recordings (excluding aperiodic arrhythmias such as atrial fibrillation). The mean difference between systems was less than ±1 bpm in all patient groups studied. Although respiratory data were more frequently rejected as invalid because of the high sensitivity of impedance pneumography to motion artefacts, valid rates were reported for 50% of recordings with a mean difference of less than ±1 brpm compared with the bedside monitor. Correlation between systems was statistically significant (p<0.0001) for heart and respiratory rate, apart from respiratory rate in patients with atrial fibrillation (p=0.02). Conclusions: Overall agreement between digital patch and clinical monitor was satisfactory, as was the efficacy of the system for automatic rejection of invalid data. Wireless monitoring technologies, such as the one tested, may offer clinical value when implemented as part of wider hospital systems that integrate and support existing clinical protocols and workflows.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
A Real-Time QRS Detection Algorithm
Pan, J., & Tompkins, W. J. (1985) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1985.325532 · Cited by 7,855
Abraham, W. T., Adamson, P. B., Bourge, R. C., Aaron, M. F., Costanzo, M. R., Stevenson, L. W., Strickland, W., Neelagaru, S., Raval, N., Krueger, S., et al. (2011) · The Lancet · doi:10.1016/s0140-6736(11)60101-3 · Cited by 1,744
The CONNECT (Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision) Trial
Crossley, G. H., Boyle, A., Vitense, H., Chang, Y., & Mead, R. H. (2011) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2010.12.012 · Cited by 526 · cited 1×
“Some have used implantable devices: for example, the study of Abraham et al(10) monitored pulmonary artery pressure intermittently in heart failure; others have used equipment implanted for other reasons.(11) Wireless technology platforms are being applied to increasing numbers of clinical situations: interesting exam…”
Method
Cited by
Wearable devices in clinical trials: hype and hypothesis
Izmailova, E. S., Wagner, J. A., & Perakslis, E. D. (2018) · Clinical Pharmacology & Therapeutics · doi:10.1002/cpt.966 · Cited by 265
Leenen, J. P. L., Leerentveld, C., van Dijk, J. D., van Westreenen, H. L., Schoonhoven, L., & Patijn, G. A. (2020) · Journal of Medical Internet Research · doi:10.2196/18636 · Cited by 224
“…(–9.4/7.0)–1.0 (–8.0/6.0)Cardiovascular disorders (atrial fibrillation) ————–0.4 (–11.4/10.6)0.9 (–3.5/5.3)Metabolic disorders ————0.1 (–7.7/7.9)–0.02 (–6.98/7.02)Diabetes ————–0.43 (–6.10/5.20)–0.23 (–0.61/0.15)SensiumVitals, Hernandez-Silveira et al, 2015 [61] ———0.82 (–1.13/2.78) 2.93…”
InfluentialContinuous Monitoring of Vital Signs Using Wearable Devices on the General Ward: Pilot Study
Weenk, M., van Goor, H., Frietman, B., Engelen, L. J., van Laarhoven, C. J., Smit, J., Bredie, S. J., & van de Belt, T. H. (2017) · JMIR mHealth and uHealth · doi:10.2196/mhealth.7208 · Cited by 223
“Recently, the SensiumVitals digital patch was tested in hospitalized patients [54].”
Improving detection of patient deterioration in the general hospital ward environment
Vincent, J.-L., Einav, S., Pearse, R., Jaber, S., Kranke, P., Overdyk, F. J., Whitaker, D. K., Gordo, F., Dahan, A., & Hoeft, A. (2018) · European Journal of Anaesthesiology · doi:10.1097/eja.0000000000000798 · Cited by 174
Pipek, L. Z., Nascimento, R. F. V., Acencio, M. M. P., & Teixeira, L. R. (2021) · Scientific Reports · doi:10.1038/s41598-021-98453-3 · Cited by 87
Metadata
- Authors
- Hernandez-Silveira, M., Ahmed, K., Ang, S.-S., Zandari, F., Mehta, T., Weir, R., Burdett, A., Toumazou, C., & Brett, S. J.
- Venue
- BMJ Open
- Publisher
- BMJ
- Published
- 2015-05
- Volume
- 5
- Issue
- 5
- Pages
- e006606
- ISSN
- 2044-6055 (ISSN-L)
- OpenAlex
- W1860274335
- Cited by
- 83
- References
- 4
- FWCI
- 2.36
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.