Evaluation of a wearable wireless device with artificial intelligence, iThermonitor WT705, for continuous temperature monitoring for patients in surgical wards: a prospective comparative study
Open accessAbstract
OBJECTIVES: To evaluate a new-generation, non-invasive, wireless axillary thermometer with artificial intelligence, iThermonitor (WT705, Raiing Medical, Beijing, China), and to ascertain its feasibility for perioperative continuous body temperature monitoring in surgical patients. SETTING: Departments of Biliary Surgery and Operating Room and the post-anaesthesia care unit of a university teaching hospital in Chengdu, China. PARTICIPANTS: A total of 526 adult surgical patients were consecutively enrolled. DESIGN: This was a prospective observational study. Axillary temperatures were continuously recorded with iThermonitor throughout the whole perioperative period. The temperatures of the contralateral armpit were measured with mercury thermometers at 8:00, 12:00, 16:00 and 20:00 every day and were used as references. OUTCOME MEASURES: The outcomes were the accuracy and precision of the temperatures measured with iThermonitor, the validity to detect fever and the feasibility of continuous wear. Pairs of temperatures were evaluated with Student's t-test, Pearson's correlation and repeated-measures Bland-Altman plot. RESULTS: 0.755, p<0.001). The 95% limits of agreement (LoA) ranged from -0.63°C to 0.73°C, with 5.11% of the differences outside the 95% LoA. The intraclass correlation coefficient was 0.753. Continuous temperature monitoring captured more fevers than intermittent observation (117/526 vs 91/526, p<0.001), detected fever up to 4.35 hours earlier, and captured a higher peak temperature (0.29°C±0.27°C, 95% CI: 0.26-0.31). All subjects felt that wearing iThermonitor was more or less comfortable and did not affect their daily activities. CONCLUSIONS: iThermonitor is promising for continuous remote temperature monitoring in surgical patients. However, further developments are still needed to improve the precision of this device, especially for temperature detection in underweight patients and those with lower body temperature. TRIAL REGISTRATION NUMBER: ChiCTR1900024549; Results (registered on 5 July 2019).
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The 18 included studies were rated as having a moderate risk of bias.
“Overall, the 18 studies were rated as having a moderate risk of bias (29–48).”
From Artificial intelligence in nursing: an integrative review of clinical and operational impacts (2025) · cites this paper as [34]
Liu et al. (34) evaluated the iThermonitor WT705, an AI-assisted wearable thermometer, in surgical wards; the device identified fevers in 22% of patients compared to 17% with standard mercury thermometers and detected fevers up to 4.35 h earlier.
“Liu et al. (34) evaluated the iThermonitor WT705, an AI-assisted wearable thermometer, in surgical wards. The device identified fevers in 22% of patients compared to 17% with standard mercury thermometers and detected fevers up to 4.35 h earlier, allowing nurses to manage potential infections or complications proactively.”
From Artificial intelligence in nursing: an integrative review of clinical and operational impacts (2025) · cites this paper as [34]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Agreement Between Methods of Measurement with Multiple Observations Per Individual
Bland, J. M., & Altman, D. G. (2007) · Journal of Biopharmaceutical Statistics · doi:10.1080/10543400701329422 · Cited by 1,864 · see citing works
Zou, G. (2011) · Statistical Methods in Medical Research · doi:10.1177/0962280211402548 · Cited by 252 · see citing works
Niven, D. J., Gaudet, J. E., Laupland, K. B., Mrklas, K. J., Roberts, D. J., & Stelfox, H. T. (2015) · Annals of Internal Medicine · doi:10.7326/m15-1150 · Cited by 240 · see citing works
Malignant Hyperthermia Deaths Related to Inadequate Temperature Monitoring, 2007–2012
Larach, M. G., Brandom, B. W., Allen, G. C., Gronert, G. A., & Lehman, E. B. (2014) · Anesthesia & Analgesia · doi:10.1213/ane.0000000000000421 · Cited by 213 · see citing works
Downey, C. L., Chapman, S. J., Randell, R., Brown, J. M., & Jayne, D. G. (2018) · International Journal of Nursing Studies · doi:10.1016/j.ijnurstu.2018.04.013 · Cited by 184 · see citing works
The Future of Wearable Technologies and Remote Monitoring in Health Care
Liao, Y., Thompson, C. A., Peterson, S. K., Mandrola, J. M., & Beg, M. S. (2019) · American Society of Clinical Oncology Educational Book · doi:10.1200/edbk_238919 · Cited by 180 · see citing works
Fever and Fever of Unknown Origin: Review, Recent Advances, and Lingering Dogma
Wright, W. F., & Auwaerter, P. G. (2020) · Open Forum Infectious Diseases · doi:10.1093/ofid/ofaa132 · Cited by 151 · see citing works
Marcy, S. M., Kohl, K. S., Dagan, R., Nalin, D. R., Blum, M., Jones, M. C., Hansen, J. R., Labadie, J., Lee, L., Martin, B. L., et al. (2003) · Vaccine · doi:10.1016/j.vaccine.2003.09.007 · Cited by 144 · see citing works
Accuracy of Peripheral Thermometers for Estimating Temperature
Niven, D. J., Laupland, K. B., & Stelfox, H. T. (2016) · Annals of Internal Medicine · doi:10.7326/l16-0069 · Cited by 135 · see citing works
Breteler, M. J. M., Huizinga, E., van Loon, K., Leenen, L. P. H., Dohmen, D., Kalkman, C. J., & Blokhuis, T. J. (2018) · BMJ Open · doi:10.1136/bmjopen-2017-020162 · Cited by 124 · see citing works
Mužný, M., Henriksen, A., Giordanengo, A., Mužík, J., Grøttland, A., Blixgård, H., Hartvigsen, G., & Årsand, E. (2019) · International Journal of Medical Informatics · doi:10.1016/j.ijmedinf.2019.104017 · Cited by 76 · see citing works
Downey, C. L., Ng, S., Jayne, D. G., & Wong, D. (2019) · BMJ Open · doi:10.1136/bmjopen-2019-031150 · Cited by 71 · see citing works
Temperature measurements with a temporal scanner: systematic review and meta-analysis
Geijer, H., Udumyan, R., Lohse, G., & Nilsagård, Y. (2016) · BMJ Open · doi:10.1136/bmjopen-2015-009509 · Cited by 66 · see citing works
Lu, S., & Dai, Y. (2009) · International Journal of Nursing Studies · doi:10.1016/j.ijnurstu.2008.11.006 · Cited by 62 · see citing works
Rubia-Rubia, J., Arias, Á., Sierra, A., & Aguirre‐Jaime, A. (2010) · International Journal of Nursing Studies · doi:10.1016/j.ijnurstu.2010.11.003 · Cited by 59 · see citing works
Performance of a non-contact infrared thermometer in healthy newborns
Sollai, S., Dani, C., Berti, E., Fancelli, C., Galli, L., de Martino, M., & Chiappini, E. (2016) · BMJ Open · doi:10.1136/bmjopen-2015-008695 · Cited by 48 · see citing works
Apa, H., Gözmen, S., Bayram, N., Çatkoğlu, A., Devrim, F., Karaarslan, U., Günay, İ., Ünal, N., & Devrim, İ. (2013) · Pediatric Emergency Care · doi:10.1097/pec.0b013e3182a2d419 · Cited by 43 · see citing works
Evaluation of Wearable Digital Devices in a Phase I Clinical Trial
Izmailova, E. S., McLean, L., Bhatia, G., Hather, G., Cantor, M., Merberg, D., Perakslis, E. D., Benko, C., & Wagner, J. A. (2019) · Clinical and Translational Science · doi:10.1111/cts.12602 · Cited by 41 · see citing works
Pei, L., Huang, Y., Mao, G., & Sessler, D. I. (2017) · Anesthesia & Analgesia · doi:10.1213/ane.0000000000002706 · Cited by 39 · see citing works
Calonder, E. M., Sendelbach, S. E., Hodges, J. S., Gustafson, C., Machemer, C. H., Johnson, D., & Reiland, L. (2010) · Journal of PeriAnesthesia Nursing · doi:10.1016/j.jopan.2010.01.006 · Cited by 32 · see citing works
Middleton, S., McElduff, P., Drury, P., et al. (2018) · International Journal of Nursing Studies · doi:10.1016/j.ijnurstu.2018.09.014 · Cited by 24 · see citing works
Wearable Digital Thermometer Improves Fever Detection
Abbasi, J. (2017) · JAMA · doi:10.1001/jama.2017.10248 · Cited by 16 · see citing works
Lu, S., Dai, Y., & Yen, C. (2009) · International Journal of Nursing Studies · doi:10.1016/j.ijnurstu.2009.05.002 · Cited by 14 · see citing works
Balla, H. Z., Theodorsson, E., & Ström, J. O. (2019) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365513.2018.1519722 · Cited by 14 · see citing works
Dakappa, P. H. (2016) · JOURNAL OF CLINICAL AND DIAGNOSTIC RESEARCH · doi:10.7860/jcdr/2016/21617.8586 · Cited by 12 · see citing works
Origin, Evolution and Clinical Application of the Thermometer
Wright, W. F., & Mackowiak, P. A. (2016) · The American Journal of the Medical Sciences · doi:10.1016/j.amjms.2015.11.019 · Cited by 11 · see citing works
Kakarmath, S. S., de Redon, E., Centi, A. J., Palacholla, R. S., Kvedar, J. C., Jethwani, K., & Agboola, S. O. (2018) · JMIR Pediatrics and Parenting · doi:10.2196/10804 · Cited by 11 · see citing works
Awareness assessment of harmful effects of mercury in a health care set-up in India
Halder, N., Peshin, S. S., Pandey, R. M., et al. (2013) · Toxicology and Industrial Health · doi:10.1177/0748233713488237 · Cited by 10 · see citing works
Opersteny, E., Anderson, H., Bates, J., Davenport, K., Husby, J., Myking, K., & Oron, A. P. (2017) · Journal of Pediatric Nursing · doi:10.1016/j.pedn.2017.02.003 · Cited by 10 · see citing works
Li, P., Yang, Y. L., & Xiong, W. (2015) · Biological Trace Element Research · doi:10.1007/s12011-015-0391-7 · Cited by 9 · see citing works
Opioids and postoperative shivering
Sessler, D. I. (2016) · Journal of Clinical Anesthesia · doi:10.1016/j.jclinane.2015.11.006 · Cited by 8 · see citing works
Mercurial hospital waste: management, becoming and recommendations.
Chadli‐Debbiche, A., Allani, R., Jouini, R., et al. (2018) · PubMed · Cited by 3 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Artificial intelligence in nursing: an integrative review of clinical and operational impacts
Hassanein, S., El Arab, R. A., Abdrbo, A., Abu-Mahfouz, M. S., Gaballah, M. K. F., Seweid, M. M., Almari, M., & Alzghoul, H. (2025) · Frontiers in Digital Health · doi:10.3389/fdgth.2025.1552372 · Cited by 134 · see citing works
Koomson, V. J., & Lang’at, C. K. (2022) · Sustainable Development Goals Series · doi:10.1007/978-3-030-83864-5_31 · Cited by 4 · see citing works
- Authors
- Liu, Y., Liu, C., Gao, M., Wang, Y., Bai, Y., Xu, R., & Gong, R.
- Affiliations
- Sichuan University (CN)
- West China Hospital of Sichuan University (CN)
- Corresponding author
- Gong, Renrong
- Venue
- BMJ Open
- Publisher
- BMJ
- Published
- 2020-11
- Volume
- 10
- Issue
- 11
- Pages
- e039474
- ISSN
- 2044-6055 (ISSN-L)
- PMC
- PMC7677341
- OpenAlex
- W3101682050
- Cited by
- 35
- References
- 32
- Citations by year
- 2026: 6 · 2025: 10 · 2024: 8 · 2023: 4 · 2022: 4 · 2021: 3
- FWCI
- 1.75
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- BMJ Open (publishedVersion, cc-by-nc)
- PubMed Central (publishedVersion, cc-by-nc)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- HighWire Press Open Archive (submittedVersion)
- Funders
- Sichuan University West China Nursing Discipline Development Special Fund Project (NO.HXHL19031)
- Topics
- Health Sciences › Medicine › Critical Care and Intensive Care Medicine › Thermal Regulation in Medicine
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Infrared Thermography in Medicine
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Keywords
- perioperative monitoring, surgical patients, fever detection, non-invasive temperature measurement, artificial intelligence, remote patient monitoring
- MeSH terms
- Beijing, Wearable Electronic Devices, Adult, Artificial Intelligence, Body Temperature, China, Humans, Prospective Studies, Temperature, Thermometers, Reproducibility of Results
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review