European interlaboratory comparison of breath13CO2 analysis

Open access
Stellaard, F, Geypens, B · 1998 · Gut
Cited by 80

The DOI matches its Crossref record.

Abstract

The BIOMED I programme Stable Isotopes in Gastroenterology and Nutrition (SIGN) has focused upon evaluation and standardisation of stable isotope breath tests using 13 C labelled substrates. The programme dealt with comparison of 13 C substrates, test meals, test conditions, analysis techniques, and calculation procedures. Analytical techniques applied for 13 CO 2 analysis were evaluated by taking an inventory of instrumentation, calibration protocols, and analysis procedures. Two ring tests were initiated measuring 13 C abundances of carbonate materials. Evaluating the data it was found that seven different models of isotope ratio mass spectrometers (IRMS) were used by the participants applying both the dual inlet system and the continuous flow configuration. Eight different brands of certified 13 C reference materials were used with a 13 C abundance varying from δ 13 C PDB −37.2 to +2.0 ‰. CO 2 was liberated from certified material by three techniques and different working standards were used varying from −47.4 to +0.4 ‰ in their δ 13 C PDB value. The standard deviations (SDs) found for all measurements by all participants were 0.25 ‰ and 0.50 ‰ for two carbonates used in the ring tests. The individual variation for the single participants varied from 0.02 ‰ (dual inlet system) to 0.14 ‰ (continuous flow system). The measurement of the difference between two carbonates showed a SD of 0.33 ‰ calculated for all participants. Internal precision of IRMS as indicated by the specifications of the different instrument suppliers is <0.05 ‰ for dual inlet systems and <0.3‰ for continuous flow systems. In this respect it can be concluded that all participants are working well within the instrument specifications even including sample preparation. Increased overall interlaboratory variation is therefore likely to be due to non-instrumental conditions. It is possible that consistent differences in sample handling leading to isotope fractionation are the causes for interlaboratory variation. Breath analysis does not require sample preparation. As such, interlaboratory variation will be less than observed for the carbonate samples and within the range indicated as internal precision for continuous flow instruments. From this it is concluded that pure analytical interlaboratory variation is acceptable despite the many differences in instrumentation and analytical protocols. Coordinated metabolic studies appear possible, in which different European laboratories perform 13 CO 2 analysis. Evaluation of compatibility of the analytical systems remains advisable, however.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1136/gut.43.2008.s2

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

Metadata

Authors
Stellaard, F., & Geypens, B.
Venue
Gut
Publisher
BMJ
Published
1998-11-01
Volume
43
Issue
suppl 3
Pages
S2-S6
ISSN
  • 0017-5749 (ISSN-L)
  • 1468-3288
OpenAlex
W1996597513
Cited by
80
References
0
FWCI
7.92
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.