Image-Based Predictive Modeling of Heart Mechanics
Abstract
Personalized biophysical modeling of the heart is a useful approach for noninvasively analyzing and predicting in vivo cardiac mechanics. Three main developments support this style of analysis: state-of-the-art cardiac imaging technologies, modern computational infrastructure, and advanced mathematical modeling techniques. In vivo measurements of cardiac structure and function can be integrated using sophisticated computational methods to investigate mechanisms of myocardial function and dysfunction, and can aid in clinical diagnosis and developing personalized treatment. In this article, we review the state-of-the-art in cardiac imaging modalities, model-based interpretation of 3D images of cardiac structure and function, and recent advances in modeling that allow personalized predictions of heart mechanics. We discuss how using such image-based modeling frameworks can increase the understanding of the fundamental biophysics behind cardiac mechanics, and assist with diagnosis, surgical guidance, and treatment planning. Addressing the challenges in this field will require a coordinated effort from both the clinical-imaging and modeling communities. We also discuss future directions that can be taken to bridge the gap between basic science and clinical translation.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Given knowledge of the boundary loading, and imaged geometry and displacements, finite element analysis can estimate material properties compatible with the imaged deformation.
“Given knowledge of the boundary loading, and imaged geometry and displacements, finite element analysis can estimate material properties compatible with the imaged deformation [147].”
From AI in Medical Imaging Informatics: Current Challenges and Future Directions (2020) · cites this paper as [147]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Verification of cardiac tissue electrophysiology simulators using an <i>N</i> -version benchmark
Niederer, S. A., Kerfoot, E., Benson, A. P., Bernabeu, M. O., Bernus, O., Bradley, C., Cherry, E. M., Clayton, R., Fenton, F. H., Garny, A., et al. (2011) · Philosophical Transactions of the Royal Society A Mathematical Physical and Engineering Sciences · doi:10.1098/rsta.2011.0139 · Cited by 275 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
AI in Medical Imaging Informatics: Current Challenges and Future Directions
Panayides, A. S., Amini, A., Filipovic, N. D., Sharma, A., Tsaftaris, S. A., Young, A., Foran, D., Do, N., Golemati, S., Kurc, T., et al. (2020) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2020.2991043 · Cited by 713 · see citing works
“Feature inversion, where the difference between an input image and its reconstruction from a representation at a given layer, is another approach that can capture the relevant patches of the image at the considered layer [136].”
Viral Vectors for Gene Therapy: Translational and Clinical Outlook
Kotterman, M. A., Chalberg, T. W., & Schaffer, D. V. (2015) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-071813-104938 · Cited by 444 · see citing works
Passive myocardial mechanical properties: meaning, measurement, models
Emig, R., Zgierski-Johnston, C., Timmermann, V., Taberner, A., Nash, M., Kohl, P., & Peyronnet, R. (2021) · Biophysical Reviews · doi:10.1007/s12551-021-00838-1 · Cited by 115 · see citing works
“Comprehensive reviews of the use of finite elements for analysing in vivo cardiac mechanics can be found elsewhere (Wang et al. 2015; Chabiniok et al. 2016).”
BackgroundSelf-citationGao, H., Aderhold, A., Mangion, K., Luo, X.-Y., Husmeier, D., & Berry, C. (2017) · Journal of the Royal Society Interface · doi:10.1098/rsif.2017.0203 · Cited by 77 · see citing works
Pashakhanloo, F., Herzka, D., Mori, S., Zviman, M., Halperin, H., Gai, N., Bluemke, D., Trayanova, N., & McVeigh, E. (2017) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-016-0317-3 · Cited by 54
“DTMRI data derived from normal hearts have been utilized by many groups to accurately represent the fiber orientation in cardiac models [36].”
BackgroundMethodAdvances in computational modelling for personalised medicine after myocardial infarction
Mangion, K., Gao, H., Husmeier, D., Luo, X.-Y., & Berry, C. (2017) · Heart · doi:10.1136/heartjnl-2017-311449 · Cited by 53
“7 Most of myocardial active models are based on ‘the sliding theory ’ at cellular level and upscaled to tissue level (table 1).”
BackgroundMethodInfluentialFinsberg, H., Xi, C., Tan, J. L., Zhong, L., Genet, M., Sundnes, J., Lee, L., & Wall, S. (2018) · International Journal for Numerical Methods in Biomedical Engineering · doi:10.1002/cnm.2982 · Cited by 51 · see citing works
“66 Our results support this finding.”
Modelling mitral valvular dynamics–current trend and future directions
Gao, H., Qi, N., Feng, L.-Y., Ma, X.-S., Danton, M., Berry, C., & Luo, X.-Y. (2017) · International Journal for Numerical Methods in Biomedical Engineering · doi:10.1002/cnm.2858 · Cited by 38 · see citing works
Niestrawska, J., Augustin, C. M., & Plank, G. (2020) · Acta Biomaterialia · doi:10.1016/j.actbio.2020.02.010 · Cited by 35 · see citing works
“Additionally, the microstructure of the eart is not as easily characterized as for other soft tissues such as lood vessels or skin and there is still a profound lack of available ata [111] .”
BackgroundDavies, V., Noè, U., Lazarus, A., Gao, H., Macdonald, B., Berry, C., Luo, X.-Y., & Husmeier, D. (2019) · Journal of the Royal Statistical Society, Series C: Applied Statistics · doi:10.1111/rssc.12374 · Cited by 30 · see citing works
“…with in vivo data from cardiac magnetic resonance imaging (MRI), computational modelling of cardiac biomechanics can provide unique insights into cardiac function in both healthy and diseased states (Wang et al., 2015; Chabiniok et al., 2016; Gao, Aderhold, Mangion, Luo, Husmeier and Berry, 2017).”
BackgroundGao, H., Mangion, K., Carrick, D., Husmeier, D., Luo, X.-Y., & Berry, C. (2017) · Scientific Reports · doi:10.1038/s41598-017-13635-2 · Cited by 26 · see citing works
Modelling Cardiac Tissue Growth and Remodelling
Wang, V. Y., Hussan, J., Yousefi, H., Bradley, C., Hunter, P., & Nash, M. (2017) · Journal of elasticity · doi:10.1007/s10659-017-9640-7 · Cited by 23 · see citing works
“Limiting factors include the phenomenological nature of the equations, and the lack of identifiability of the constitutive parameters [76].”
BackgroundSelf-citationRodero, C., Baptiste, T. M. G., Barrows, R. K., Lewalle, A., Niederer, S., & Strocchi, M. (2023) · Frontiers of Physics · doi:10.3389/fphy.2023.1306210 · Cited by 22 · see citing works
“In [2], Wang et al. discussed some of the main articles on cardiac biomechanical models, with an emphasis on the images from which the anatomical data were extracted (for an introductory overview of image-based anatomical and mechanical modelling, see [3]).”
BackgroundAtlas Based Computational Analysis of Heart Shape and Function in Congenital Heart Disease
Gilbert, K., Forsch, N., Hegde, S., Mauger, C., Omens, J., Perry, J., Pontré, B., Suinesiaputra, A., Young, A., & McCulloch, A. (2018) · Journal of Cardiovascular Translational Research · doi:10.1007/s12265-017-9778-5 · Cited by 22 · see citing works
“These methods enable estimation of regional wall stress, contractility, relaxation parameters, and ventricular stiffness [33, 60].”
Cardiac Image Modelling: Breadth and Depth in Heart Disease
Suinesiaputra, A., McCulloch, A., Nash, M., Pontré, B., & Young, A. (2016) · Medical Image Anal. · doi:10.1016/j.media.2016.06.027 · Cited by 22 · see citing works
“Finite element analysis of cardiac mechanics can incorporate realistic non-linear, anisotropic material properties, active force generation, resulting in realistic simulations of cardiac mechanics (Chabiniok et al., 2016; Wang et al., 2015).”
BackgroundSelf-citationNguyen, T. D., Kadri, O. E., & Voronov, R. S. (2020) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2020.529365 · Cited by 19 · see citing works
“…2015; Beheshti et al., 2016; Gray and Pathmanathan, 2018; Ni et al., 2018); tissue mechanics computations (Trayanova, 2011, 2012; Sun et al., 2014; Wang et al., 2015; Chabiniok et al., 2016; Niederer et al., 2019a,b), ventricular thrombosis (Mittal et al., 2016) and the use of models in…”
BackgroundInfluentialRomaszko, L., Borowska, A., Lazarus, A., Dalton, D., Berry, C., Luo, X.-Y., Husmeier, D., & Gao, H. (2021) · Artif. Intell. Medicine · doi:10.1016/j.artmed.2021.102140 · Cited by 18 · see citing works
“More recent methods for cardiac geometry reconstruction include manual iterative interventions for reconstruction [11], and warping an idealized ventricular geometry, e.”
MethodRumindo, G. K., Ohayon, J., Croisille, P., & Clarysse, P. (2020) · Medical Engineering and Physics · doi:10.1016/j.medengphy.2020.09.003 · Cited by 18
“The determination of subject-specific constitutive material parameters is a challenging inverse problem involving several key components: hearts shape, tissue structure and motion representation, myocardial material modeling, definition of appropriate boundary and loading conditions and optimization strategy [2].”
BackgroundBiomechanics of infarcted left ventricle: a review of modelling
Li, W.-G. (2020) · Biomedical Engineering Letters · doi:10.1007/s13534-020-00159-4 · Cited by 18 · see citing works
Gall, A. L., Vallée, F., Pushparajah, K., Hussain, T., Mebazaa, A., Chapelle, D., Gayat, E., & Chabiniok, R. (2020) · PLoS ONE · doi:10.1371/journal.pone.0232830 · Cited by 18 · see citing works
Deep Learning Over Reduced Intrinsic Domains for Efficient Mechanics of the Left Ventricle
Talou, G. M. M., Gamage, T. P. B., Sagar, M., & Nash, M. (2020) · Frontiers of Physics · doi:10.3389/fphy.2020.00030 · Cited by 17 · see citing works
Self-citationCombining computer modelling and cardiac imaging to understand right ventricular pump function
Walmsley, J., van Everdingen, W. V., Cramer, M., Prinzen, F., Delhaas, T., & Lumens, J. (2017) · Cardiovascular Research · doi:10.1093/cvr/cvx154 · Cited by 17 · see citing works
“However, most computer models use either generic or LV models to represent active and passive RV myocardial function.(12) Hill-type phenomenological models have been used in both geometrically detailed and reduceddimensional models to simulate contraction.”
Motiwale, S., & Sacks, M. S. (2025) · Mechanics of Soft Materials · doi:10.1007/s42558-024-00064-1 · Cited by 16
Dobutamine stress testing in patients with Fontan circulation augmented by biomechanical modeling
Ruijsink, B., Zugaj, K., Wong, J., Pushparajah, K., Hussain, T., Moireau, P., Razavi, R., Chapelle, D., & Chabiniok, R. (2020) · PLoS ONE · doi:10.1371/journal.pone.0229015 · Cited by 16
“As opposed to image or signal analysis, the physical and physiological principles of these models allow to estimate the biophysical properties of the heart and vascular system—quantities not directly measurable from the image and pressure data but accessible via constitutive relations and equilibrium equations (see re…”
BackgroundOptimization Framework for Patient-Specific Cardiac Modeling
Mineroff, J., McCulloch, A., Krummen, D., Ganapathysubramanian, B., & Krishnamurthy, A. (2019) · Cardiovascular Engineering and Technology · doi:10.1007/s13239-019-00428-z · Cited by 13 · see citing works
Review of Cardiac-Coronary Interaction and Insights from Mathematical Modeling
Fan, L., Wang, H.-F., Kassab, G. S., & Lee, L. (2024) · WIREs Mechanisms of Disease · doi:10.1002/wsbm.1642 · Cited by 12 · see citing works
Validating MRI-Derived Myocardial Stiffness Estimates Using In Vitro Synthetic Heart Models
Kolawole, F. O., Peirlinck, M., Cork, T., Levenston, M., Kuhl, E., & Ennis, D. B. (2023) · Annals of Biomedical Engineering · doi:10.1007/s10439-023-03164-7 · Cited by 11
Gusseva, M., Hussain, T., Friesen, C. H. H., Greil, G., Chapelle, D., & Chabiniok, R. (2021) · Annals of Biomedical Engineering · doi:10.1007/s10439-021-02895-9 · Cited by 9
Relative identifiability of anisotropic properties from magnetic resonance elastography
Miller, R. M., Kolipaka, A., Nash, M., & Young, A. (2017) · NMR in Biomedicine · doi:10.1002/nbm.3848 · Cited by 9 · see citing works
Self-citationZhang, Y., Adams, J., Wang, V. Y., Horwitz, L., Tartibi, M., Morgan, A. E., Kim, J. J., Wallace, A., Weinsaft, J., Ge, L., et al. (2021) · Comput. Biol. Medicine · doi:10.1016/j.compbiomed.2021.104840 · Cited by 8 · see citing works
Self-citationStructure Tensors for Dispersed Fibers in Soft Materials
Kalhöfer-Köchling, M., Bodenschatz, E., & Wang, Y. (2020) · Physical Review Applied · doi:10.1103/physrevapplied.13.064039 · Cited by 8 · see citing works
Multisite pacing and myocardial scars: a computational study
Albatat, M., Bergsland, J., Arevalo, H., Odland, H., Wall, S., Sundnes, J., & Balasingham, I. (2020) · Computer Methods in Biomechanics and Biomedical Engineering · doi:10.1080/10255842.2020.1711885 · Cited by 8
“Advances in computational modelling has improved the understanding of the interrelation between different underlying factors in HF (Wang et al. 2015).”
BackgroundDeveloping cardiac biomechanical models beyond the clinic: modeling stressors of daily life
Lewalle, A., Baptiste, T. M. G., Barrows, R. K., Cicci, L., Corrado, C., Lee, A. W., Rodero, C., Solís-Lemus, J., Strocchi, M., & Niederer, S. (2025) · Biomechanics and Modeling in Mechanobiology · doi:10.1007/s10237-025-01982-3 · Cited by 7 · see citing works
“Biomechanical models of the heart have become invaluable tools for building an integrated understanding of cardiac function (Trayanova and Rice 2011; Trayanova 2011; Niederer et al. 2019; Wang et al. 2015).”
Tea Disease Spot Recognition Based on Image Feature Points Extraction and Matching
Chen, J. (2020) · Global NEST Journal · doi:10.30955/gnj.003375 · Cited by 7
“In this study, the average spectral reflectance of ROI of four kinds of samples is selected from 80 leaves of anthracnose, 72 leaves of red disease spot, 80 leaves of tea white star and 60 healthy leaves respectively (Ismail et al. , 2019; Wang Nielsen and Nash, 2015), as shown in Figure 2.”
BackgroundKolawole, F. O., Wang, V. Y., Freytag, B., Loecher, M., Cork, T., Nash, M. P., Kuhl, E., & Ennis, D. B. (2025) · Scientific Reports · doi:10.1038/s41598-025-89243-2 · Cited by 6
Self-citationPhase Stabilization with Motion Compensated Diffusion Weighted Imaging
Hannum, A. J., Cork, T., Setsompop, K., & Ennis, D. B. (2024) · Magnetic Resonance in Medicine · doi:10.1002/mrm.30218 · Cited by 5 · see citing works
“The bulk contraction in the heart was a complex contracting and twisting motion, 59,60 thus compensating for velocity and acceleration in all three-directions proved conse-quential for phase consistency.”
Comparative Evaluation of Different Emulators for Cardiac Mechanics
Dalton, D., Lazarus, A., & Husmeier, D. (2020) · Proceedings of the 2nd International Conference on Statistics: Theory and Applications · doi:10.11159/icsta20.126 · Cited by 4
“Introduction Recent advances into differential-equation based models of cardiac mechanics have demonstrated the potential for their use in a clinical setting for the diagnosis of cardiac defects [1].”
BackgroundAwadi, R., Benameur, N., Kraiem, T., & Labidi, S. (2022) · CENTERIS/ProjMAN/HCist · doi:10.1016/j.procs.2023.01.399 · Cited by 4
Mu, S., Chan, W., & Yap, C. (2025) · arXiv.org · doi:10.48550/arxiv.2506.20696 · Cited by 3
“This has previously been performed for both healthy and pathological cardiac conditions [46, 19], providing details of myocardial deformation, stress distribution, and contractile function [24], which offered physiology insights into diseases and intervention scenarios [17].”
Correcting bias in cardiac geometries derived from multimodal images using spatiotemporal mapping
Zhao, D., Mauger, C., Gilbert, K., Wang, V. Y., Quill, G., Sutton, T., Lowe, B., Legget, M., Ruygrok, P., Doughty, R., et al. (2023) · Scientific Reports · doi:10.1038/s41598-023-33968-5 · Cited by 3 · see citing works
Self-citationWang, V. Y., Tartibi, M., Zhang, Y., Selvaganesan, K., Haraldsson, H., Auger, D., Faraji, F., Spaulding, K. A., Takaba, K., Collins, A., et al. (2021) · Magnetic Resonance in Medicine · doi:10.1002/mrm.28775 · Cited by 3
“The KMAF framework (Figure 1) has 4 principal components including: (1) cine- CMR facilitated DENSE segmentation; (2) raw 3D Eulerian displacement calculation; (3) displacement interpolation onto LV FE model nodes; and (4) 3D Lagrangian strain analysis.”
MethodSelf-citationImage-Based Polygonal Lattices for Mechanical Modeling of Biological Materials: 2D Demonstrations
Liu, D., Chen, C.-S., & Zhang, T. (2021) · ACS Biomaterials Science & Engineering · doi:10.1021/acsbiomaterials.0c01772 · Cited by 2
A Framework for Evaluating Myocardial Stiffness Using 3D-Printed Heart Phantoms
Kolawole, F. O., Peirlinck, M., Cork, T., Wang, V. Y., Dual, S., Levenston, M., Kuhl, E., & Ennis, D. (2021) · Functional Imaging and Modeling of the Heart · doi:10.1007/978-3-030-78710-3_30 · Cited by 2
Self-citationMu, S., Chan, W., & Yap, C. (2026) · Comput. Methods Programs Biomed. · doi:10.1016/j.cmpb.2026.109243 · Cited by 1
The Effect of Voxel Volume and Voxel Shape on Cardiac Diffusion Tensor Imaging Metrics
Hannum, A. J., Cork, T. E., Setsompop, K., & Ennis, D. B. (2025) · Magnetic Resonance in Medicine · doi:10.1002/mrm.70189 · Cited by 1 · see citing works
“While isotropic voxels may reduce physiological bias , the known in-slice (short axis) spatial gradients in cardiac structure and function [33, 34] suggest that anisotropic voxels may have the best performance [35].”
Acute effects of mechanical dyssynchrony on left ventricular function and coronary perfusion
Choy, J., Fan, L., Awakeem, Y., Cai, C., Raissi, F., Lee, L., & Kassab, G. S. (2025) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2025.1630854 · Cited by 1
“This index can only be used as a global measurement, however, and it is also a load-dependent parameter that might be affected by preload, afterload and geometry (Wang et al., 2015).”
Fan, L., Choy, J. S., Cai, C., Teague, S. D., Guccione, J. M., Lee, L. C., & Kassab, G. S. (2024) · Bioengineering · doi:10.3390/bioengineering11070735 · Cited by 3 · see citing works
“These quantities, however, are load-dependent and vary due to many factors, such as a change in preload, afterload, and geometries [5].”
BackgroundHasaballa, A., Gamage, T. P. B., Wang, V. Y., Zhao, D., Mauger, C., Gilbert, K., Wang, Z., Freytag, B., Cao, J., Young, A., et al. (2021) · Functional Imaging and Modeling of the Heart · doi:10.1007/978-3-030-78710-3_28 · Cited by 1
Self-citationCardiac Modeling for Multisystem Inflammatory Syndrome in Children (MIS-C, PIMS-TS)
Waugh, R., Hussein, M., Weller, J., Sharma, K., Greil, G., Kahn, J., Hussain, T., & Chabiniok, R. (2021) · Functional Imaging and Modeling of the Heart · doi:10.1007/978-3-030-78710-3_42 · Cited by 1
Translational Cardiovascular Modeling: Tetralogy of Fallot and Modeling of Diseases
Chabiniok, R., Škardová, K., Galabov, R., Eichler, P., Gusseva, M., Janoušek, J., Fučík, R., Tintěra, J., Oberhuber, T., & Hussain, T. (2021) · Nečas Center Series · doi:10.1007/978-3-030-88084-2_6 · Cited by 1 · see citing works
Automatic Assessment of Myocardial biomechanical Properties from Cardiac MRI via Deep Learning
Awadi, R., Benameur, N., Deriche, M., & Labidi, S. (2026) · 2025 2nd International Conference on Innovative and Intelligent Information Technologies (IC3IT) · doi:10.1109/ic3it69043.2026.11567605
“The precise quantification of these parameters is consequently imperative for comprehending the mechanisms underpinning diseases, informing therapeutic approaches, and facilitating individualized cardiac management [8].”
Maga, L., Peirlinck, M., & Noël, L. (2025) · Computational Mechanics · doi:10.1007/s00466-026-02860-0
Mu, S., Chan, W., & Yap, C. (2025) · arXiv.org · doi:10.48550/arxiv.2507.18677
“Myocardial biomechanics studies such as image-based FE modeling [35, 13] and myocardial stiffness estimation from medical images [24] are important tools that improved our understanding of cardiac physiology, and have much potential for future translation into clinical tools for disease evaluation…”
Ghafarinatanzi, M., & Périé, D. (2023) · Biomechanics and Modeling in Mechanobiology · doi:10.1007/s10237-022-01675-1
“While the transformation from prolate spherical coordinate to Cartesian follow (Wang et al. 2015)”
MethodA prediction model of microcirculation disorder in myocardium based on ultrasonic images
Tian, M.-J., Zheng, M.-J., Qiu, S., & Song, Y. (2022) · Journal of Ambient Intelligence and Humanized Computing · doi:10.1007/s12652-022-04440-5
“In terms of disease prediction: Wang et al. (2015) constructed a mechanical model to analyze heart movement.”
BackgroundImage-Based Modelling of Cardiac Mechanics
Al-saffar, M. O., Al-Dahhan, Z. T., & Al-taweel, R. B. (2021) · Al-Nahrain Journal for Engineering Sciences · doi:10.29194/njes.24020098
“As it is able to assist in stratifying different forms of heart failure , gaining an understanding of the causes of HF and informing mechanical therapy design, the clinical community is very involved in research into an in vivo assessment of the mechanical function of tissues [1].”
BackgroundA Framework for Evaluating Myocardial Stiffness Using 3D-Printed Hearts
Kolawole, F. O., Peirlinck, M., Cork, T., Wang, V. Y., Dual, S., Levenston, M., Kuhl, E., & Ennis, D. (2021) · bioRxiv · doi:10.1101/2021.02.21.432169
Self-citationMu, S., Chan, W., & Yap, C. (2026) · Engineering applications of artificial intelligence · doi:10.1016/j.engappai.2026.115398
Continuum Mechanics and Nonlinear Elasticity
Taber, L. (2020) · doi:10.1007/978-3-030-43209-6_3
Problems in Soft Tissue Biomechanics
Taber, L. (2020) · doi:10.1007/978-3-030-43209-6_4
Taber, L. (2020) · Continuum Modeling in Mechanobiology · doi:10.1007/978-3-030-43209-6_5
- Authors
- Wang, V. Y., Nielsen, P. M. F., & Nash, M. P.
- Affiliations
- Auckland Bioengineering Institute and
- University of Auckland (NZ)
- Venue
- Annual Review of Biomedical Engineering
- Publisher
- Annual Reviews
- Published
- 2015-12-07
- Volume
- 17
- Issue
- 1
- Pages
- 351-383
- ISSN
- 1523-9829 (ISSN-L)
- 1545-4274
- OpenAlex
- W2159293185
- Cited by
- 63
- References
- 0
- Citations by year
- 2026: 2 · 2025: 6 · 2024: 3 · 2023: 6 · 2022: 3 · 2021: 13 · 2020: 14 · 2019: 3 · 2018: 2 · 2017: 8 · 2016: 3
- FWCI
- 0.75
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Annual Review of Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- ResearchSpace (University of Auckland) (submittedVersion)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Function and Risk Factors
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Cardiac Imaging and Diagnostics
- Physical Sciences › Engineering › Biomedical Engineering › Elasticity and Material Modeling
- Keywords
- cardiac mechanics, image-based modeling, cardiac imaging modalities, myocardial function, 3D cardiac imaging, computational cardiac modeling, surgical guidance, treatment planning, clinical translation
- MeSH terms
- Animals, Biomechanical Phenomena, Biomedical Engineering, Heart, Hemodynamics, Humans, Magnetic Resonance Imaging, Models, Cardiovascular, Myocardial Contraction, Myocardium, Tomography, X-Ray Computed, Imaging, Three-Dimensional, Biophysical Phenomena, Diffusion Tensor Imaging
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review