Two-View and Single-View Tomosynthesis versus Full-Field Digital Mammography: High-Resolution X-Ray Imaging Observer Study
Abstract
PURPOSE: To compare the diagnostic accuracy of two-dimensional (2D) full-field digital mammography with that of two-view (mediolateral and craniocaudal) and single-view (mediolateral oblique) tomosynthesis in an observer study involving two institutions. MATERIALS AND METHODS: Ethical committee approval was obtained. All participating women gave informed consent. Two hundred twenty women (mean age, 56.3; range, 40-80 years) with breast density of 2-4 according to American College of Radiology criteria were recruited between November 2008 and September 2009 and underwent standard treatment plus tomosynthesis with a prototype photon-counting machine. After exclusion criteria were met, this resulted in a final test set of 130 women. Ten accredited readers classified the 130 cases (40 cancers, 24 benign lesions, and 66 normal images) using 2D mammography and two-view tomosynthesis. Another 10 readers reviewed the same cases using 2D mammography but single-view tomosynthesis. The multireader, multicase receiver operating characteristic (ROC) method was applied. The significance of the observed difference in accuracy between 2D mammography and tomosynthesis was calculated. RESULTS: For diagnostic accuracy, 2D mammography performed significantly worse than two-view tomosynthesis (average area under ROC curve [AUC] = 0.772 for 2D, AUC = 0.851 for tomosynthesis, P = .021). Significant differences were found for both masses and microcalcification (P = .037 and .049). The difference in AUC between the two modalities of -0.110 was significant (P = .03) only for the five readers with the least experience (<10 years of reading); with AUC of -0.047 for the five readers with 10 years or more experience (P = .25). No significant difference (P = .79) in reader performance was seen when 2D mammography (average AUC = 0.774) was compared with single-view tomosynthesis (average AUC = 0.775). CONCLUSION: Two-view tomosynthesis outperforms 2D mammography but only for readers with the least experience. The benefits were seen for both masses and microcalcification. No differences in classification accuracy was seen between and 2D mammography and single-view tomosynthesis.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Skaane, P., Bandos, A. I., Gullien, R., Eben, E. B., Ekseth, U., Haakenaasen, U., Izadi, M., Jebsen, I. N., Jahr, G., Krager, M., et al. (2013) · Radiology · doi:10.1148/radiol.12121373 · Cited by 1,012
Ciatto, S., Houssami, N., Bernardi, D., Caumo, F., Pellegrini, M., Brunelli, S., Tuttobene, P., Bricolo, P., Fantò, C., Valentini, M., et al. (2013) · The Lancet Oncology · doi:10.1016/s1470-2045(13)70134-7 · Cited by 838
Rodriguez-Ruiz, A., Lång, K., Gubern-Merida, A., Broeders, M., Gennaro, G., Clauser, P., Helbich, T. H., Chevalier, M., Tan, T., Mertelmeier, T., et al. (2019) · JNCI: Journal of the National Cancer Institute · doi:10.1093/jnci/djy222 · Cited by 692
Self-citationSkaane, P., Bandos, A. I., Eben, E. B., Jebsen, I. N., Krager, M., Haakenaasen, U., Ekseth, U., Izadi, M., Hofvind, S., & Gullien, R. (2014) · Radiology · doi:10.1148/radiol.13131391 · Cited by 332
Samala, R. K., Chan, H., Hadjiiski, L., Helvie, M. A., Wei, J., & Cha, K. (2016) · Medical Physics · doi:10.1118/1.4967345 · Cited by 289
Metadata
- Authors
- Wallis, M. G., Moa, E., Zanca, F., Leifland, K., & Danielsson, M.
- Venue
- Radiology
- Published
- 2012-03
- Volume
- 262
- Issue
- 3
- Pages
- 788-796
- ISSN
- 0033-8419 (ISSN-L)
- 1527-1315
- OpenAlex
- W2146448637
- Cited by
- 187
- References
- 0
- FWCI
- 20.16
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Two-View and Single-View Tomosynthesis versus Full-Field Digital Mammography: High-Resolution X-Ray Imaging Observer Study. https://app.cochat.ai/papers/10.1148/radiol.11103514
- The paper
- Wallis, M. G., Moa, E., Zanca, F., Leifland, K., & Danielsson, M. (2012). Two-View and Single-View Tomosynthesis versus Full-Field Digital Mammography: High-Resolution X-Ray Imaging Observer Study. Radiology. https://doi.org/10.1148/radiol.11103514
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.