Multi–Detector Row CT Attenuation Measurements: Assessment of Intra- and Interscanner Variability with an Anthropomorphic Body CT Phantom
Abstract
PURPOSE: To determine the dependence of absolute computed tomographic (CT) attenuation values on multi-detector row CT scanner type, convolution kernel, and tube current by using an anthropomorphic phantom. MATERIALS AND METHODS: A customized phantom was designed with tissue-equivalent materials to simulate contrast material-enhanced liver, spleen, pancreas, aorta, kidney, 0- and 50-HU cylindric renal cysts, muscle, and fat. The phantom was scanned with five multi-detector row CT scanners (LightSpeed QXi, GE Healthcare, Milwaukee, Wis; MX8000, Philips Medical Systems, Best, the Netherlands; and Volume Zoom, Sensation 16 and Sensation 64, Siemens Medical Solutions, Forchheim, Germany) on five separate occasions with 120 kVp, low and high tube current settings, 3.00-3.75-mm section thickness, 50% overlap, and standard and high-spatial-resolution kernels. Standardized regions of interest (ROIs) were used to obtain 3510 attenuation measurements. Attenuation dependence on scanner, kernel, and tube current was evaluated by using F tests derived with mixed-model regression. Within the mixed-model framework, the Tukey honestly significant difference procedure and a Bonferroni multiple comparison correction were used to assess differences among imaging regimens and tube current settings, respectively, in terms of tissue attenuation and ROI standard deviation. RESULTS: Tube current had no significant effect (P>.4) on observed tissue attenuation. Significant (P<.0001) differences were observed between imaging regimens with respect to mean attenuation for each tissue type. Convolution kernel modification had an inconsistent effect on tissue attenuation, depending on the scanner. All multi-detector row CT scanners displayed intrascanner variability in tissue attenuation (minimum range: 8.4 HU for fat tissue with the Sensation 16; maximum range: 63.4 HU for liver tissue with the Sensation 64). The scanners behaved differently at the lower range of the CT number scale, where 0-HU cyst attenuation ranged from -15.7 to 23.9 HU and one vendor's equipment showed significantly lower mean attenuation values. CONCLUSION: CT attenuation values vary significantly between different manufacturers' multi-detector row CT scanners, among different generations of multi-detector row CT scanning equipment, and with individual combinations of scanner and convolution kernel. SUPPLEMENTAL MATERIAL: http://radiology.rsnajnls.org/cgi/content/full/242/1/109/DC1
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
The current radiological approach to renal cysts.
Bosniak, M. A. (1986) · Radiology · doi:10.1148/radiology.158.1.3510019 · Cited by 1,046
Characterization of adrenal masses using unenhanced CT: an analysis of the CT literature
Boland, G. W., Lee, M. J., Gazelle, G. S., Halpern, E. F., McNicholas, M. M., & Mueller, P. R. (1998) · American Journal of Roentgenology · doi:10.2214/ajr.171.1.9648789 · Cited by 693
Bosniak, M. A. (1991) · Radiology · doi:10.1148/radiology.179.2.2014269 · Cited by 299
CT of the solitary pulmonary nodule
Siegelman, S., Zerhouni, E., Leo, F., Khouri, N., & Stitik, F. (1980) · American Journal of Roentgenology · doi:10.2214/ajr.135.1.1 · Cited by 277
The unreliability of CT numbers as absolute values
Levi, C., Gray, J., McCullough, E., & Hattery, R. (1982) · American Journal of Roentgenology · doi:10.2214/ajr.139.3.443 · Cited by 240
Vertebral bone mineral analysis: an integrated approach with CT.
Kalender, W. A., Klotz, E., & Suess, C. (1987) · Radiology · doi:10.1148/radiology.164.2.3602380 · Cited by 204
Faulkner, K. G., Glüer, C.-C., Majumdar, S., Lang, P. K., Engelke, K., & Genant, H. K. (1991) · American Journal of Roentgenology · doi:10.2214/ajr.157.6.1950872 · Cited by 113
Factors Influencing Quantitative CT Measurements of Solitary Pulmonary Nodules
Zerhouni, E. A., Spivey, J. F., Morgan, R. H., Leo, F. P., Stitik, F. P., & Siegelman, S. S. (1982) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-198212000-00005 · Cited by 97
Measuring Coronary Calcium on CT Images Adjusted for Attenuation Differences
Nelson, J. C., Kronmal, R. A., Carr, J. J., McNitt‐Gray, M. F., Wong, N. D., Loria, C. M., Goldin, J. G., Williams, O. D., & Detrano, R. (2005) · Radiology · doi:10.1148/radiol.2352040515 · Cited by 90
CT Characterization of Adrenal Masses: The Time Has Come
KOROBKIN, M. T. (2000) · Radiology · doi:10.1148/radiology.217.3.r00dc52629 · Cited by 87
Cited by
Quantitative assessment of liver fat with magnetic resonance imaging and spectroscopy
Reeder, S. B., Cruite, I., Hamilton, G., & Sirlin, C. B. (2011) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.22580 · Cited by 822
Schwenzer, N. F., Springer, F., Schraml, C., Stefan, N., Machann, J., & Schick, F. (2009) · Journal of Hepatology · doi:10.1016/j.jhep.2009.05.023 · Cited by 749
Abbara, S., Arbab-Zadeh, A., Callister, T. Q., Desai, M. Y., Mamuya, W., Thomson, L., & Weigold, W. G. (2009) · Journal of Cardiovascular Computed Tomography · doi:10.1016/j.jcct.2009.03.004 · Cited by 580
Management of the Incidental Renal Mass
Silverman, S. G., Israel, G. M., Herts, B. R., & Richie, J. P. (2008) · Radiology · doi:10.1148/radiol.2491070783 · Cited by 434
Radiologic evaluation of nonalcoholic fatty liver disease
Lee, S. S. (2014) · World Journal of Gastroenterology · doi:10.3748/wjg.v20.i23.7392 · Cited by 426
Metadata
- Authors
- Birnbaum, B. A., Hindman, N., Lee, J., & Babb, J. S.
- Venue
- Radiology
- Published
- 2007-01
- Volume
- 242
- Issue
- 1
- Pages
- 109-119
- ISSN
- 0033-8419 (ISSN-L)
- 1527-1315
- OpenAlex
- W2148518552
- Cited by
- 274
- References
- 11
- FWCI
- 7.77
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Multi–Detector Row CT Attenuation Measurements: Assessment of Intra- and Interscanner Variability with an Anthropomorphic Body CT Phantom. https://app.cochat.ai/papers/10.1148/radiol.2421052066
- The paper
- Birnbaum, B. A., Hindman, N., Lee, J., & Babb, J. S. (2007). Multi–Detector Row CT Attenuation Measurements: Assessment of Intra- and Interscanner Variability with an Anthropomorphic Body CT Phantom. Radiology. https://doi.org/10.1148/radiol.2421052066
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.