Multi–Detector Row CT Attenuation Measurements: Assessment of Intra- and Interscanner Variability with an Anthropomorphic Body CT Phantom

Birnbaum, Bernard A., Hindman, Nicole, Lee, Julie, Babb, James S. · 2007 · Radiology

Abstract

PURPOSE: To determine the dependence of absolute computed tomographic (CT) attenuation values on multi-detector row CT scanner type, convolution kernel, and tube current by using an anthropomorphic phantom. MATERIALS AND METHODS: A customized phantom was designed with tissue-equivalent materials to simulate contrast material-enhanced liver, spleen, pancreas, aorta, kidney, 0- and 50-HU cylindric renal cysts, muscle, and fat. The phantom was scanned with five multi-detector row CT scanners (LightSpeed QXi, GE Healthcare, Milwaukee, Wis; MX8000, Philips Medical Systems, Best, the Netherlands; and Volume Zoom, Sensation 16 and Sensation 64, Siemens Medical Solutions, Forchheim, Germany) on five separate occasions with 120 kVp, low and high tube current settings, 3.00-3.75-mm section thickness, 50% overlap, and standard and high-spatial-resolution kernels. Standardized regions of interest (ROIs) were used to obtain 3510 attenuation measurements. Attenuation dependence on scanner, kernel, and tube current was evaluated by using F tests derived with mixed-model regression. Within the mixed-model framework, the Tukey honestly significant difference procedure and a Bonferroni multiple comparison correction were used to assess differences among imaging regimens and tube current settings, respectively, in terms of tissue attenuation and ROI standard deviation. RESULTS: Tube current had no significant effect (P>.4) on observed tissue attenuation. Significant (P<.0001) differences were observed between imaging regimens with respect to mean attenuation for each tissue type. Convolution kernel modification had an inconsistent effect on tissue attenuation, depending on the scanner. All multi-detector row CT scanners displayed intrascanner variability in tissue attenuation (minimum range: 8.4 HU for fat tissue with the Sensation 16; maximum range: 63.4 HU for liver tissue with the Sensation 64). The scanners behaved differently at the lower range of the CT number scale, where 0-HU cyst attenuation ranged from -15.7 to 23.9 HU and one vendor's equipment showed significantly lower mean attenuation values. CONCLUSION: CT attenuation values vary significantly between different manufacturers' multi-detector row CT scanners, among different generations of multi-detector row CT scanning equipment, and with individual combinations of scanner and convolution kernel. SUPPLEMENTAL MATERIAL: http://radiology.rsnajnls.org/cgi/content/full/242/1/109/DC1

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1148/radiol.2421052066

Cites

11 of 11 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 11 references

Cited by

See all 274

Metadata

Authors
Birnbaum, B. A., Hindman, N., Lee, J., & Babb, J. S.
Venue
Radiology
Published
2007-01
Volume
242
Issue
1
Pages
109-119
ISSN
  • 0033-8419 (ISSN-L)
  • 1527-1315
OpenAlex
W2148518552
Cited by
274
References
11
FWCI
7.77
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Multi–Detector Row CT Attenuation Measurements: Assessment of Intra- and Interscanner Variability with an Anthropomorphic Body CT Phantom. https://app.cochat.ai/papers/10.1148/radiol.2421052066
The paper
Birnbaum, B. A., Hindman, N., Lee, J., & Babb, J. S. (2007). Multi–Detector Row CT Attenuation Measurements: Assessment of Intra- and Interscanner Variability with an Anthropomorphic Body CT Phantom. Radiology. https://doi.org/10.1148/radiol.2421052066

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.