Dose Performance of a 64-Channel Dual-Source CT Scanner1

McCollough, Cynthia H., Primak, Andrew N., Saba, Osama, Bruder, Herbert, Stierstorfer, Karl, Raupach, Rainer, Suess, Christoph, Schmidt, Bernhard, Ohnesorge, Bernd M., Flohr, Thomas G. · 2007 · Radiology

Abstract

PURPOSE: To prospectively compare the dose performance of a 64-channel multi-detector row computed tomographic (CT) scanner and a 64-channel dual-source CT scanner from the same manufacturer. MATERIALS AND METHODS: To minimize dose in the cardiac (dual-source) mode, the evaluated dual-source CT system uses a cardiac beam-shaping filter, three-dimensional adaptive noise reduction, heart rate-dependent pitch, and electrocardiographically based modulation of the tube current. Weighted CT dose index per 100 mAs was measured for the head, body, and cardiac beam-shaping filters. Kerma-length product was measured in the spiral cardiac mode at four pitch values and three electrocardiographic modulation temporal windows. Noise was measured in an anthropomorphic phantom. Data were compared with data from a 64-channel multi-detector row CT scanner. RESULTS: For the multi-detector row and dual-source CT systems, respectively, weighted CT dose index per 100 mAs was 14.2 and 12.2 mGy (head CT), 6.8 and 6.4 mGy (body CT), and 6.8 and 5.3 mGy (cardiac CT). In the spiral cardiac mode (no electrocardiographically based tube current modulation, 0.2 pitch), equivalent noise occurred at volume CT dose index values of 23.7 and 35.0 mGy (coronary artery calcium CT) and 58.9 and 61.2 mGy (coronary CT angiography) for multi-detector row CT and dual-source CT, respectively. The use of heart rate-dependent pitch values reduced volume CT dose index to 46.2 mGy (0.265 pitch), 34.0 mGy (0.36 pitch), and 26.6 mGy (0.46 pitch) compared with 61.2 mGy for 0.2 pitch. The use of electrocardiographically based tube current-modulation and temporal windows of 110, 210, and 310 msec further reduced volume CT dose index to 9.1-25.1 mGy, dependent on the heart rate. CONCLUSION: For electrocardiographically gated coronary CT angiography, image noise equivalent to that of multi-detector row CT can be achieved with dual-source CT at doses comparable to or up to a factor of two lower than the doses at multi-detector row CT, depending on heart rate of the patient.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1148/radiol.2433061165

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 202

Metadata

Authors
McCollough, C. H., Primak, A. N., Saba, O., Bruder, H., Stierstorfer, K., Raupach, R., Suess, C., Schmidt, B., Ohnesorge, B. M., & Flohr, T. G.
Venue
Radiology
Published
2007-06
Volume
243
Issue
3
Pages
775-784
ISSN
  • 0033-8419 (ISSN-L)
  • 1527-1315
OpenAlex
W2140905817
Cited by
202
References
0
FWCI
27.36
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Dose Performance of a 64-Channel Dual-Source CT Scanner1. https://app.cochat.ai/papers/10.1148/radiol.2433061165
The paper
McCollough, C. H., Primak, A. N., Saba, O., Bruder, H., Stierstorfer, K., Raupach, R., Suess, C., Schmidt, B., Ohnesorge, B. M., & Flohr, T. G. (2007). Dose Performance of a 64-Channel Dual-Source CT Scanner1. Radiology. https://doi.org/10.1148/radiol.2433061165

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.