Evaluating Variability in Tumor Measurements from Same-day Repeat CT Scans of Patients with Non–Small Cell Lung Cancer

Open access
Zhao, Binsheng, James, Leonard P., Moskowitz, Chaya S., Guo, Pingzhen, Ginsberg, Michelle S., Lefkowitz, Robert A., Qin, Yilin, Riely, Gregory J., Kris, Mark G., Schwartz, Lawrence H.·2009·Radiology

Abstract

PURPOSE: To evaluate the variability of tumor unidimensional, bidimensional, and volumetric measurements on same-day repeat computed tomographic (CT) scans in patients with non-small cell lung cancer. MATERIALS AND METHODS: This HIPAA-compliant study was approved by the institutional review board, with informed patient consent. Thirty-two patients with non-small cell lung cancer, each of whom underwent two CT scans of the chest within 15 minutes by using the same imaging protocol, were included in this study. Three radiologists independently measured the two greatest diameters of each lesion on both scans and, during another session, measured the same tumors on the first scan. In a separate analysis, computer software was applied to assist in the calculation of the two greatest diameters and the volume of each lesion on both scans. Concordance correlation coefficients (CCCs) and Bland-Altman plots were used to assess the agreements between the measurements of the two repeat scans (reproducibility) and between the two repeat readings of the same scan (repeatability). RESULTS: The reproducibility and repeatability of the three radiologists' measurements were high (all CCCs, >or=0.96). The reproducibility of the computer-aided measurements was even higher (all CCCs, 1.00). The 95% limits of agreements for the computer-aided unidimensional, bidimensional, and volumetric measurements on two repeat scans were (-7.3%, 6.2%), (-17.6%, 19.8%), and (-12.1%, 13.4%), respectively. CONCLUSION: Chest CT scans are well reproducible. Changes in unidimensional lesion size of 8% or greater exceed the measurement variability of the computer method and can be considered significant when estimating the outcome of therapy in a patient.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1148/radiol.2522081593

Cites

18 of 18 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 18 references

Cited by

See all 339

Metadata

Authors
Zhao, B., James, L. P., Moskowitz, C. S., Guo, P., Ginsberg, M. S., Lefkowitz, R. A., Qin, Y., Riely, G. J., Kris, M. G., & Schwartz, L. H.
Venue
Radiology
Published
2009-07
Volume
252
Issue
1
Pages
263-272
ISSN
  • 0033-8419 (ISSN-L)
  • 1527-1315
OpenAlex
W2124932867
Cited by
339
References
18
FWCI
9.66
Type
journal article
Language
en
Open access
green

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Evaluating Variability in Tumor Measurements from Same-day Repeat CT Scans of Patients with Non–Small Cell Lung Cancer. https://app.cochat.ai/papers/10.1148/radiol.2522081593
The paper
Zhao, B., James, L. P., Moskowitz, C. S., Guo, P., Ginsberg, M. S., Lefkowitz, R. A., Qin, Y., Riely, G. J., Kris, M. G., & Schwartz, L. H. (2009). Evaluating Variability in Tumor Measurements from Same-day Repeat CT Scans of Patients with Non–Small Cell Lung Cancer. Radiology. https://doi.org/10.1148/radiol.2522081593

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.