The DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.
Prospective Randomized Comparison of Irrigated-Tip Versus Conventional-Tip Catheters for Ablation of Common Flutter
Open accessAbstract
Background —Radiofrequency (RF) ablation of common flutter requires the creation of a complete ablation line to produce bidirectional conduction block in the cavotricuspid isthmus. An irrigated-tip ablation catheter has been shown to be effective in patients in whom conventional ablation has failed. This randomized study compares the efficacy and safety of this catheter with those of a conventional catheter for de novo flutter ablation. Methods and Results —Cavotricuspid ablation was performed with a conventional (n=26) or an irrigated-tip catheter (n=24). RF was applied for 60 minutes with a temperature-controlled mode: 65°C to 70°C up to 70 W with a conventional catheter or 50°C up to 50 W (with a 17-mL/min saline flow rate) with the irrigated-tip catheter. The end point was the achievement of bidirectional isthmus block, and a crossover was performed after 21 unsuccessful applications. Procedural ablation parameters as well as number of applications, x-ray exposure, procedure duration, impedance rise, and clot formation were compared for each group. A coronary angiogram was performed before and after each ablation for the first 30 patients. Complete bidirectional isthmus block was achieved for all patients. Four patients crossed over from conventional to irrigated-tip catheters. The number of applications, procedure duration, and x-ray exposure were significantly higher with the conventional than with the irrigated-tip catheter: 13±10 versus 5±3 pulses, 53±41 versus 27±16 minutes, and 18±14 versus 9±6 minutes, respectively. No significant side effects occurred, and the coronary angiograms of the first 30 patients after ablation were unchanged. Conclusions —Irrigated-tip catheters were found to be more effective than and as safe as conventional catheters for flutter ablation, facilitating the rapid achievement of bidirectional isthmus block.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Jackman, W. M., Wang, X., Friday, K. J., Roman, C. A., Moulton, K. P., Beckman, K. J., McClelland, J. H., Twidale, N., Hazlitt, H. A., Prior, M. I., et al. (1991) · New England Journal of Medicine · doi:10.1056/nejm199106063242301 · Cited by 1,422
Irani, W. N., Grayburn, P. A., & Afridi, I. (1997) · Circulation · doi:10.1161/01.cir.95.4.962 · Cited by 151
Resolution of cardiomyopathy after ablation of atrial flutter
Luchsinger, J. A., & Steinberg, J. S. (1998) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(98)00183-1 · Cited by 121
Cited by
Warnes, C. A., Williams, R. G., Bashore, T. M., Child, J. S., Connolly, H. M., Dearani, J. A., del Nido, P., Fasules, J. W., Graham, T. P., Hijazi, Z. M., et al. (2008) · Circulation · doi:10.1161/circulationaha.108.190811 · Cited by 2,066
ACC/AHA 2008 Guidelines for the Management of Adults With Congenital Heart Disease
Warnes, C. A., Williams, R. G., Bashore, T. M., Child, J. S., Connolly, H. M., Dearani, J. A., del Nido, P., Fasules, J. W., Graham, T. P., Hijazi, Z. M., et al. (2008) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2008.10.001 · Cited by 1,533
Arrhythmias in Adult Patients With Congenital Heart Disease
Walsh, E. P., & Cecchin, F. (2007) · Circulation · doi:10.1161/circulationaha.105.592410 · Cited by 398
Coronary Arterial Spasm During Pulsed Field Ablation to Treat Atrial Fibrillation
Reddy, V. Y., Petru, J., Funasako, M., Kopriva, K., Hala, P., Chovanec, M., Janotka, M., Kralovec, S., & Neuzil, P. (2022) · Circulation · doi:10.1161/circulationaha.122.061497 · Cited by 291
Triedman, J. K., Alexander, M. E., Love, B. A., Collins, K. K., Berul, C. I., Bevilacqua, L. M., & Walsh, E. P. (2002) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(02)01858-2 · Cited by 260
Metadata
- Authors
- Jaïs, P., Shah, D. C., Haïssaguerre, M., Hocini, M., Garrigue, S., Le Metayer, P., & Clémenty, J.
- Venue
- Circulation
- Published
- 2000-02-22
- Volume
- 101
- Issue
- 7
- Pages
- 772-776
- ISSN
- 0009-7322 (ISSN-L)
- 1524-4539
- OpenAlex
- W1971257477
- Cited by
- 172
- References
- 3
- FWCI
- 7.99
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Prospective Randomized Comparison of Irrigated-Tip Versus Conventional-Tip Catheters for Ablation of Common Flutter. https://app.cochat.ai/papers/10.1161/01.cir.101.7.772
- The paper
- Jaïs, P., Shah, D. C., Haïssaguerre, M., Hocini, M., Garrigue, S., Le Metayer, P., & Clémenty, J. (2000). Prospective Randomized Comparison of Irrigated-Tip Versus Conventional-Tip Catheters for Ablation of Common Flutter. Circulation. https://doi.org/10.1161/01.cir.101.7.772
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.