Recommendations for Human Blood Pressure Determinations by Sphygmomanometers
Open accessAbstract
IT SHOULD be clearly recognized that arte- rial pressures cannot be measured with precision by means of sphygmomanometers.Direct registration of pressures by cali- brated intra-arterial manometers has shown (1) that even during quiet breathing and slight sinus arrhythmia, systolic and diastolic pres- sures vary from beat to beat by several mm.Hg, and that these differences are greatly in- tensified during states of arrhythmia and deep breathing; (2) that auscultatory systolic read- ings from the brachial artery average 3 or 4 mm.Hg too low and show average scatter of ±8 mm.Hg; and (3) that auscultatory diastolic pressures taken at the point of dulling of the sounds average about 8 mm.Hg too high.The errors of clinical measurement of blood pres- sure can be summarized by saying that in normal persons a mean error of ±8 mm.Hg may be expected in individual readings of sys- tolic and diastolic pressures.Despite this, clin- ical blood pressure determinations have proved very serviceable for practising physicians, clin- icians, insurance carriers, and others concerned with physical examinations, diagnosis, prog- nosis, or therapy.It is important, however, that any basic deficiency of sphygmomanom- etry be not increased by additional errors due to apparatus or technic.The following recommendations have been drawn up with the idea of aiding examiners to avoid pitfalls and, as far as possible, to establish The authors constituted the Committee to Revise "Standardization of High Blood Pressure Readings."This Committee, whose Chairman was Dr. Carl J. Wiggers, functioned under the auspices of the Council for High Blood Pressure Research of the Scientific Council of the American Heart Association.These recommendations are being presented simul_ taneously in Circulation and in the Journal of the American Medical Association.less.A length of bag sufficient to half-encircle
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“the mean error of the cuff method averaged 8 mmHg from intra-arterial brachial BP for both systolic and diastolic BP [40].”
From Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure (2022) · cites this paper as [40]
In the citing paper
“As early as in 1951, a committee of the Council for High BP Research of the American Heart Association reported that the sphygmomanometer auscultation method underestimates intra-arterial brachial systolic BP by an average of 3–4 mmHg, but overestimates diastolic BP by an average 8 mmHg [40].”
From Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure (2022) · cites this paper as [40]
Citations
Cites
Still being checked.
Cited by
The Relation of Adiposity to Blood Pressure and Development of Hypertension: The Framingham Study
KANNEL, W. B., BRAND, N., SKINNER, J. J., DAWBER, T. R., & MCNAMARA, P. M. (1967) · Annals of Internal Medicine · doi:10.7326/0003-4819-67-1-48 · Cited by 817
Characterization of the oscillometric method for measuring indirect blood pressure
Geddes, L. A., Voelz, M., Combs, C., Reiner, D., & Babbs, C. F. (1982) · Annals of Biomedical Engineering · doi:10.1007/bf02367308 · Cited by 340
Perceived Exertion and Pulse Rate during Graded Exercise in Various Age Groups
Borg, G., & Linderholm, H. (1967) · Acta Medica Scandinavica · doi:10.1111/j.0954-6820.1967.tb12626.x · Cited by 304
Measuring Blood Pressure Accurately
Jones, D. W., Appel, L. J., Sheps, S. G., Roccella, E. J., & Lenfant, C. J. M. (2003) · JAMA · doi:10.1001/jama.289.8.1027 · Cited by 263
EVIDENCE FOR RELATIONSHIP BETWEEN SODIUM (CHLORIDE) INTAKE AND HUMAN ESSENTIAL HYPERTENSION
Dahl, L. K. (1954) · Archives of Internal Medicine · doi:10.1001/archinte.1954.00250040017003 · Cited by 211
Metadata
- Authors
- BORDLEY, J., CONNOR, C. A. R., HAMILTON, W. F., KERR, W. J., & WIGGERS, C. J.
- Venue
- Circulation
- Published
- 1951-10
- Volume
- 4
- Issue
- 4
- Pages
- 503-509
- ISSN
- 0009-7322 (ISSN-L)
- 1524-4539
- OpenAlex
- W2988207071
- Cited by
- 262
- References
- 0
- Citations by year
- 2025: 1 · 2022: 3 · 2021: 1 · 2020: 3 · 2019: 5 · 2018: 3 · 2014: 1 · 2013: 5 · 2012: 3
- FWCI
- 5.08
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Circulation (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Keywords
- blood pressure measurement, measurement error, auscultatory blood pressure, brachial artery
- MeSH terms
- Blood Pressure, Blood Pressure Determination, Humans, Sphygmomanometers
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Recommendations for Human Blood Pressure Determinations by Sphygmomanometers. https://app.cochat.ai/papers/10.1161/01.cir.4.4.503
- The paper
- BORDLEY, J., CONNOR, C. A. R., HAMILTON, W. F., KERR, W. J., & WIGGERS, C. J. (1951). Recommendations for Human Blood Pressure Determinations by Sphygmomanometers. Circulation. https://doi.org/10.1161/01.cir.4.4.503
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.