Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure
Open accessAbstract
Automated 'oscillometric' blood pressure (BP) measuring devices (BPMDs) were developed in the 1970s to replace manual auscultatory BP measurement by mercury sphygmomanometer. Automated BPMDs that have passed accuracy testing versus a reference auscultatory sphygmomanometer using a scientifically accepted validation protocol are recommended for clinical use globally. Currently, there are many thousands of unique automated BPMDs manufactured by hundreds of companies, with each device using proprietary algorithms to estimate BP and using a method of operation that is largely unchanged since inception. Validated automated BPMDs provide similar BP values to those recorded using manual auscultation albeit with potential sources of error mostly associated with using empirical algorithms to derive BP from waveform pulsations. Much of the work to derive contemporary BP thresholds and treatment targets used to manage cardiovascular disease risk was obtained using automated BPMDs. While there is room for future refinement to improve accuracy for better individual risk stratification, validated BPMDs remain the recommended standard for office and out-of-office BP measurement to be used in hypertension diagnosis and management worldwide.
Overview
CoChat Summary
## TL;DR
This paper is a review of how manual auscultatory sphygmomanometers and automated “oscillometric” blood pressure devices work, and what each method actually measures. It argues that validated automated devices remain the recommended standard for clinical and out-of-office blood pressure measurement, while also noting device-specific algorithmic errors and room for improvement.
## Key findings
- Automated BPMDs estimate blood pressure from cuff pressure waveform signals using proprietary algorithms; the paper states the basic method has remained largely unchanged for decades and that there are more than 3,500 models from hundreds of companies.
- Validated automated BPMDs are said to provide blood pressure values similar to manual auscultation, and they are recommended globally for hypertension diagnosis and management.
- The paper reports that compared with intra-arterial brachial BP, auscultatory sphygmomanometry underestimates systolic BP by 3–4 mmHg on average and overestimates diastolic BP by about 8 mmHg; automated BPMDs underestimate systolic BP by 8.0 mmHg on average and overestimate diastolic BP by 4.5 mmHg.
- The paper notes that brachial systolic BP averages 8.0 mmHg higher than proximal aortic systolic BP, while diastolic BP is only slightly lower at the brachial artery (−1.0 mmHg), highlighting that peripheral cuff BP is not the same as central aortic BP.
## Methods
- Narrative review / technical overview comparing manual auscultatory measurement with automated oscillometric BP devices.
- Uses prior literature, validation standards, and reported quantitative comparisons; no new sample or primary dataset is described in the text provided.
- Main analysis is descriptive: it explains device operation, waveform processing, algorithmic estimation of BP, and summarizes known sources of measurement error and published comparison values.
## Limitations
- The paper states that each unique automated BPMD has different processes, algorithms, and fitting functions, and that these are “none of which are publicly disclosed,” so accuracy must be determined device by device.
- It explicitly notes limitations of the oscillometric method, including errors that can be systematic, random, or related to specific patient populations, and that estimated BP can vary with device design and waveform-envelope construction.
- Evident from the text: this is a review article rather than an original validation study, so its conclusions depend on previously published studies and do not present new empirical data.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures


![Fig. 3 Summary of the level of difference from intra-arterial brachial systolic and diastolic BP for cuff BP measured by sphygmomanometer or automated BP measuring device (BPMD). Data represent mean difference (95% confidence intervals) and are derived from individual participant level meta-analysis in ≥309 participants for BP measured by sphygmomanometer and ≥354 participants by automated BPMDs [38]. Example pulse pressures are denoted by the solid arrows. Green shaded areas represent indicative zones for aortic systolic and diastolic BP. The crosses indicate zones where non-invasive BP is markedly different from either brachial or aortic intra-arterial BPs, and the hatched arrow represents the potential underestimation of pulse pressure using an automated BPMD. Figure modified from [66].](https://figures.cochat.ai/paper-figures/624a90fe19e649cc9d5ab089a6e71cda/automated-oscillometric-blood-pressure-measuring-devices-how-they-work-and-what-figure-3.jpg)
Verification
Claims the paper makes
10 Valid · 2 Not valid · 14 Not checked · 5 Own results and interpretations
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Since each unique automated BPMD can have different processes, algorithms and fitting functions to derive BPs (none of which are publicly disclosed), the accuracy of each BPMD needs to be individually determined by comparison to a BP reference standard using a scientifically accepted validation protocol [14, 57, 58].”
§ Automated ‘oscillometric’ blood pressure measuring devices: how they work and what they measure > Automated BPMDs: how do they work, what do they measure?
What the sources say
- Not checked — no full text of the sourcecites [14] Ringrose J, Padwal R. Automated blood pressure measuring devices: how are they validated for accuracy? J Hum Hypertens. 2022 (in press). source
- Partly supportedcites [57] Stergiou GS, Alpert B, Mieke S, Asmar R, Atkins N, Eckert S, et al. A universal standard for the validation of blood pressure measuring devices: Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Collaboration Statement. Hypertension. 2018;71:368–74. sourceExperimental — accuracy not yet measured
The Objective calls for all BP measuring devices to be assessed for accuracy under an agreed validation protocol, and Validation Criteria and Reporting specifies comparison of test-device readings with reference readings. The paper does not say that each automated device uses different, undisclosed algorithms or that this is why individual validation is needed; searches for algorithms and device-model differences did not establish that rationale. Page numbers were not available in our copy. (LLM, conf=0.86)
- Partly supportedcites [58] Stergiou GS, Palatini P, Asmar R, Ioannidis JP, Kollias A, Lacy P, et al. Recommendations and Practical Guidance for performing and reporting validation studies according to the Universal Standard for the validation of blood pressure measuring devices by the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO). J Hypertens. 2019;37:459–66. sourceExperimental — accuracy not yet measured
The 'Validation data analysis—Pass/fail criteria' section (page not provided) describes assessing a test device's accuracy against reference BP readings under the AAMI/ESH/ISO standard. The copy does not say that each unique BPMD must be validated individually because its algorithms differ, or that those algorithms are undisclosed. (LLM, conf=0.86)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
The cited sources that could be checked couldn’t confirm or dispute this claim.
In the paper
“In particular, increased arterial stiffness leads to overestimation of systolic, diastolic and mean arterial BPs [60, 62–64].”
§ Automated ‘oscillometric’ blood pressure measuring devices: how they work and what they measure > Automated BPMDs: how do they work, what do they measure?
What the sources say
- Not checked — no full text of the sourcecites [60] Ursino M, Cristalli C. A mathematical study of some biomechanical factors affecting the oscillometric blood pressure measurement. IEEE Trans Biomed Eng. 1996;43:761–78. source
- Not checked — no full text of the sourcecites [62] Finnegan TP, Spence JD, Wong DG, Wells GA. Blood pressure measurement in the elderly: correlation of arterial stiffness with difference between intra-arterial and cuff pressures. J Hypertens. 1985;3:231–5. source
- Could not verifycites [63] van Popele NM, Grobbee DE, Bots ML, Asmar R, Topouchian J, Reneman RS, et al. Association between arterial stiffness and atherosclerosis: the Rotterdam Study. Stroke. 2001;32:454–60. sourceExperimental — accuracy not yet measured
The Arterial Stiffness methods section describes measuring blood pressure and calculating mean arterial pressure with a one-third pulse-pressure coefficient, but does not assess whether that calculation overestimates pressure. The Discussion says stiffness increases systolic and decreases diastolic blood pressure; it does not address overestimation of measured systolic, diastolic, or mean arterial pressure. (LLM, conf=0.76)
- Not checked — no full text of the sourcecites [64] Stergiou GS, Lourida P, Tzamouranis D, Baibas NM. Unreliable oscillometric blood pressure measurement: prevalence, repeatability and characteristics of the phenomenon. J Hum Hypertens. 2009;23:794–800. source
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“However, the range of optimal coefficients for accurate systolic and diastolic BP measurements varies greatly between devices [12, 45, 50, 54].”
§ Automated ‘oscillometric’ blood pressure measuring devices: how they work and what they measure > Automated BPMDs: how do they work, what do they measure?
What the sources say
- Not checked — no full text of the sourcecites [12] Geddes LA, Voelz M, Combs C, Reiner D, Babbs CF. Characterization of the oscillometric method for measuring indirect blood pressure. Ann Biomed Eng. 1982;10:271–80. source
- Partly supportedcites [45] Celler BG, Argha A, Le PN, Ambikairajah E. Novel methods of testing and calibration of oscillometric blood pressure monitors. PLoS One. 2018;13:e0201123. sourceExperimental — accuracy not yet measured
The Results section, “Derivation of characteristic ratios based on device readings” (page unavailable), reports significant differences between device groups in the systolic and diastolic ratios they used. But the Discussion and conclusion (page unavailable) identifies a common optimal systolic range, 0.5–0.6, and does not establish device-specific optimal ranges or an optimal diastolic ratio. The claim conflates differing device ratios with differing optimal coefficients. (LLM, conf=0.88)
- Not checked — no full text of the sourcecites [50] Forouzanfar M, Dajani HR, Groza VZ, Bolic M, Rajan S, Batkin I. Oscillometric blood pressure estimation: past, present, and future. IEEE Rev Biomed Eng. 2015;8:44–63. source
- Not checked — no full text of the sourcecites [54] Amoore JN, Vacher E, Murray IC, Mieke S, King ST, Smith FE, et al. Effect of the shapes of the oscillometric pulse amplitude envelopes and their characteristic ratios on the differences between auscultatory and oscillometric blood pressure measurements. Blood Press Monit. 2007;12:297–305. source
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“The systolic and diastolic BPs are estimated using proprietary algorithms, for example, based on empirical fixed-ratio coefficients [50, 53] that are designed to coincide with BPs measured by an auscultation sphygmomanometer.”
§ Automated ‘oscillometric’ blood pressure measuring devices: how they work and what they measure > Automated BPMDs: how do they work, what do they measure?
What the sources say
- Not checked — no full text of the sourcecites [50] Forouzanfar M, Dajani HR, Groza VZ, Bolic M, Rajan S, Batkin I. Oscillometric blood pressure estimation: past, present, and future. IEEE Rev Biomed Eng. 2015;8:44–63. source
- Partly supportedcites [53] Chandrasekhar A, Yavarimanesh M, Hahn JO, Sung SH, Chen CH, Cheng HM, et al. Formulas to explain popular oscillometric blood pressure estimation algorithms. Front Physiol. 2019;10:1415. sourceExperimental — accuracy not yet measured
The Introduction (page unavailable) describes population-based fixed ratios for estimating systolic and diastolic BP and says these algorithms are believed to be related to proprietary commercial-device algorithms. It does not say the ratios are designed to coincide with auscultatory sphygmomanometer readings; the Materials and Methods instead describe deriving the paper's fixed ratios from invasive BP measurements. (LLM, conf=0.86)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“hundreds of companies manufacturing more than 3500 different models of automated BPMDs, many of which are without evidence of having passed validation testing [25–29].”
§ Automated ‘oscillometric’ blood pressure measuring devices: how they work and what they measure > Introduction
What the sources say
- Supportedcites [25] Picone DS, Campbell NRC, Schutte AE, Olsen MH, Ordunez P, Whelton PK, et al. Validation status of blood pressure measuring devices sold globally. JAMA. 2022;327:680–1. source
The Results section reports 4287 listed devices, of which 3411 automated upper-arm/wrist devices from 457 unique manufacturers remained after exclusions, with no evidence of validation for 2602 (76.3%) and only 300 (8.8%) validated; the Discussion concludes 'A minority of automated upper arm cuff and wristbased devices globally have evidence of validation for accuracy.' This supports both 'hundreds of companies' (457 manufacturers) and 'many ... without evidence of having passed validation testing' (76.3%). The 'more than 3500 models' figure corresponds to the automated devices listed before the site/obsolescence exclusions (3606 of 4287 after removing 681 manually operated) rather than the final analytic sample of 3411, a minor numeric nuance that does not alter the attributed point. (LLM, conf=0.80)
- Not matchedcites [26] Blood Pressure Monitoring Devices Market Size, Share & Trends Analysis Report By Product (Sphygmomanometers/Aneroid BP Monitors, Ambulatory BP Monitors), By End-use, By Region, And Segment Forecasts, 2021. 20282021. Report No.: 978-1-68038-586-1.
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Not checked — no full text of the sourcecites [27] Alpert BS. Can ‘FDA-cleared’ blood pressure devices be trusted? A call to action. Blood Press Monit. 2017;22:179–81. source
- Partly supportedcites [28] Picone DS, Deshpande RA, Schultz MG, Fonseca R, Campbell NRC, Delles C, et al. Nonvalidated home blood pressure devices dominate the online marketplace in Australia. Hypertension. 2020;75:1593–9. sourceExperimental — accuracy not yet measured
The Results (p. 7) found that only 6.6% of 972 unique devices surveyed were validated or claimed equivalence, supporting the point that many lacked validation evidence. The Discussion (p. 11) also reports Medaval data for more than 3,000 devices, but the paper does not establish the claimed figure of more than 3,500 models. (LLM, conf=0.88)
- Partly supportedcites [29] Sharman JE, O’Brien E, Alpert B, Schutte AE, Delles C, Hecht Olsen M, et al. Lancet Commission on Hypertension group position statement on the global improvement of accuracy standards for devices that measure blood pressure. J Hypertension. 2020;38:21–29. sourceExperimental — accuracy not yet measured
The Introduction reports that fewer than 15% of about 3000 cuff-based BP devices had published evidence on accuracy performance, supporting the point that many lack validation evidence. It does not substantiate the cited sentence’s figure of more than 3500 automated models or its claim about hundreds of manufacturers. No page number is available in our copy. (LLM, conf=0.88)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
9 more claims
Create a free account to see all 31 claims and to verify the 4 references still open.
- Probably not a substantial claim. 4
- It may state two separate claims as one. 16
Citations
Cites
International Organization for Standardization
(2011) · SpringerReference · doi:10.1007/springerreference_16847 · Cited by 12,534
Whelton, P. K., Carey, R. M., Aronow, W. S., Casey, D. E., Collins, K. J., Dennison Himmelfarb, C., DePalma, S. M., Gidding, S., Jamerson, K. A., Jones, D. W., et al. (2018) · Hypertension · doi:10.1161/hyp.0000000000000065 · Cited by 7,391
Whelton, P. K., Carey, R. M., Aronow, W. S., Casey, D. E., Collins, K. J., Dennison Himmelfarb, C., DePalma, S. M., Gidding, S., Jamerson, K. A., Jones, D. W., et al. (2018) · Hypertension · doi:10.1161/hyp.0000000000000066 · Cited by 4,347
Ettehad, D., Emdin, C. A., Kiran, A., Anderson, S. G., Callender, T., Emberson, J., Chalmers, J., Rodgers, A., & Rahimi, K. (2016) · The Lancet · doi:10.1016/s0140-6736(15)01225-8 · Cited by 3,755
Williams, B., Mancia, G., Spiering, W., Agabiti Rosei, E., Azizi, M., Burnier, M., Clement, D. L., Coca, A., de Simone, G., Dominiczak, A., et al. (2018) · Journal of Hypertension · doi:10.1097/hjh.0000000000001940 · Cited by 2,885
O'Brien, E., Asmar, R., Beilin, L., Imai, Y., Mallion, J.-M., Mancia, G., Mengden, T., Myers, M., Padfield, P., Palatini, P., et al. (2003) · Journal of Hypertension · doi:10.1097/00004872-200305000-00001 · Cited by 1,655
Self-citationA general cardiovascular risk profile: The Framingham study
Kannel, W. B., McGee, D., & Gordon, T. (1976) · The American Journal of Cardiology · doi:10.1016/0002-9149(76)90061-8 · Cited by 1,189
Association Between Arterial Stiffness and Atherosclerosis: The Rotterdam Study
van Popele, N. M., Grobbee, D. E., Bots, M. L., Asmar, R., Topouchian, J., Reneman, R. S., Hoeks, A. P. G., van der Kuip, D. A. M., Hofman, A., & Witteman, J. C. M. (2001) · Stroke · doi:10.1161/01.str.32.2.454 · Cited by 1,142
Whelton, P. K., Carey, R. M., Aronow, W. S., Casey, D. E., Collins, K. J., Dennison Himmelfarb, C., DePalma, S. M., Gidding, S., Jamerson, K. A., Jones, D. W., et al. (2018) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2017.11.005 · Cited by 920
Research works printed but not linked: 2
- Birkenhager WH, Reid JL, editors. Handbook of hypertension. The Netherlands: Elsevier Science Publishers B.V.; 1991.
- Jilek J, Stork M. Arterial pulsations in the blood pressure cuff: are they hemodynamic pulses or oscillations? Int J Biol Biomed Eng. 2012;6:35–41.
Cited by
Advances in mechanical biomarkers
Eroles, M., & Rico, F. (2023) · Journal of Molecular Recognition · doi:10.1002/jmr.3022 · Cited by 37
Cuff-based blood pressure measurement: challenges and solutions
Pilz, N., Picone, D. S., Patzak, A. R., Opatz, O. S., Lindner, T., Fesseler, L., Heinz, V., & Bothe, T. L. (2024) · Blood Pressure · doi:10.1080/08037051.2024.2402368 · Cited by 42
Reshaping the healthcare world by AI-integrated wearable sensors following COVID-19
Khan, B., Khalid, R. T., Wara, K. U., Masrur, M. A., Khan, S., Khan, W. U., Amara, U., & Abdullah, S. (2025) · Chemical Engineering Journal · doi:10.1016/j.cej.2025.159478 · Cited by 30
Akter, S., Azhar, B. S., Kamruzzaman, M., Banna, M. H. A., Hasan, H., & Roshid, M. M. (2025) · BMC Nutrition · doi:10.1186/s40795-025-01121-2 · Cited by 35
Resonance sonomanometry for noninvasive, continuous monitoring of blood pressure
Jimenez, R., Yurk, D., Dell, S. J., Rutledge, A. C., Fu, M., Dempsey, W. P., Abu‐Mostafa, Y. S., Rajagopal, A., & Rajagopal, A. B. (2024) · PNAS Nexus · doi:10.1093/pnasnexus/pgae252 · Cited by 15
Metadata
- Authors
- Sharman, J. E., Tan, I., Stergiou, G. S., Lombardi, C., Saladini, F., Butlin, M., Padwal, R., Asayama, K., Avolio, A., Brady, T. M., Murray, A., & Parati, G.
- Affiliations
- University of Tasmania (AU)
- The George Institute for Global Health (AU)
- National and Kapodistrian University of Athens (GR)
- IRCCS Istituto Auxologico Italiano (IT)
- Macquarie University (AU)
- University of Alberta (CA)
- Teikyo University (JP)
- Johns Hopkins University (US)
- Newcastle University (GB)
- Corresponding author
- Sharman, James E.
- ORCID
- Venue
- Journal of Human Hypertension
- Published
- 2022-05-30
- Volume
- 37
- Issue
- 2
- Pages
- 93-100
- ISSN
- 0950-9240 (ISSN-L)
- 1476-5527
- PMC
- PMC9957730
- OpenAlex
- W4281710849
- Cited by
- 84
- References
- 70
- Citations by year
- 2026: 32 · 2025: 23 · 2024: 16 · 2023: 12 · 2022: 1
- FWCI
- 2.64
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by
- creativecommons.org/licenses/by/4.0 (tdm)
- creativecommons.org/licenses/by/4.0 (vor)
- Locations
- Journal of Human Hypertension (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- BOA (University of Milano-Bicocca) (submittedVersion, cc-by)
- Lirias (submittedVersion, other-oa)
- Figshare (submittedVersion, cc-by)
- Newcastle University ePrints (Newcastle Univesity) (cc-by)
- Funders
- Engineering and Physical Sciences Research Council (EP/N025342/1)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Keywords
- automated blood pressure measurement, oscillometric blood pressure devices, device validation, measurement accuracy, hypertension diagnosis, antihypertensive therapy, cardiovascular risk stratification, blood pressure thresholds, out-of-office blood pressure monitoring
- MeSH terms
- Blood Pressure, Blood Pressure Determination, Humans, Hypertension, Oscillometry, Sphygmomanometers
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol and qwen3.8-omni-flash).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure. https://app.cochat.ai/papers/10.1038/s41371-022-00693-x
- The paper
- Sharman, J. E., Tan, I., Stergiou, G. S., Lombardi, C., Saladini, F., Butlin, M., Padwal, R., Asayama, K., Avolio, A., Brady, T. M., Murray, A., & Parati, G. (2022). Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure. Journal of Human Hypertension. https://doi.org/10.1038/s41371-022-00693-x
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.