Open access

Abstract

Automated 'oscillometric' blood pressure (BP) measuring devices (BPMDs) were developed in the 1970s to replace manual auscultatory BP measurement by mercury sphygmomanometer. Automated BPMDs that have passed accuracy testing versus a reference auscultatory sphygmomanometer using a scientifically accepted validation protocol are recommended for clinical use globally. Currently, there are many thousands of unique automated BPMDs manufactured by hundreds of companies, with each device using proprietary algorithms to estimate BP and using a method of operation that is largely unchanged since inception. Validated automated BPMDs provide similar BP values to those recorded using manual auscultation albeit with potential sources of error mostly associated with using empirical algorithms to derive BP from waveform pulsations. Much of the work to derive contemporary BP thresholds and treatment targets used to manage cardiovascular disease risk was obtained using automated BPMDs. While there is room for future refinement to improve accuracy for better individual risk stratification, validated BPMDs remain the recommended standard for office and out-of-office BP measurement to be used in hypertension diagnosis and management worldwide.

Overview

CoChat Summary

## TL;DR

This paper is a review of how manual auscultatory sphygmomanometers and automated “oscillometric” blood pressure devices work, and what each method actually measures. It argues that validated automated devices remain the recommended standard for clinical and out-of-office blood pressure measurement, while also noting device-specific algorithmic errors and room for improvement.

## Key findings

  • Automated BPMDs estimate blood pressure from cuff pressure waveform signals using proprietary algorithms; the paper states the basic method has remained largely unchanged for decades and that there are more than 3,500 models from hundreds of companies.
  • Validated automated BPMDs are said to provide blood pressure values similar to manual auscultation, and they are recommended globally for hypertension diagnosis and management.
  • The paper reports that compared with intra-arterial brachial BP, auscultatory sphygmomanometry underestimates systolic BP by 3–4 mmHg on average and overestimates diastolic BP by about 8 mmHg; automated BPMDs underestimate systolic BP by 8.0 mmHg on average and overestimate diastolic BP by 4.5 mmHg.
  • The paper notes that brachial systolic BP averages 8.0 mmHg higher than proximal aortic systolic BP, while diastolic BP is only slightly lower at the brachial artery (−1.0 mmHg), highlighting that peripheral cuff BP is not the same as central aortic BP.

## Methods

  • Narrative review / technical overview comparing manual auscultatory measurement with automated oscillometric BP devices.
  • Uses prior literature, validation standards, and reported quantitative comparisons; no new sample or primary dataset is described in the text provided.
  • Main analysis is descriptive: it explains device operation, waveform processing, algorithmic estimation of BP, and summarizes known sources of measurement error and published comparison values.

## Limitations

  • The paper states that each unique automated BPMD has different processes, algorithms, and fitting functions, and that these are “none of which are publicly disclosed,” so accuracy must be determined device by device.
  • It explicitly notes limitations of the oscillometric method, including errors that can be systematic, random, or related to specific patient populations, and that estimated BP can vary with device design and waveform-envelope construction.
  • Evident from the text: this is a review article rather than an original validation study, so its conclusions depend on previously published studies and do not present new empirical data.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1038/s41371-022-00693-x

Cites

69 of 69 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Figures

Fig. 1 Summary of components for the operation of an automated blood pressure measurement device (BPMD) using the oscillometric method. Cuff inflation and deflation is controlled by a microcomputer with a miniature air pump and valve system to change cuff pressure and facilitate BP measurements. Pressure waveforms generated by cardiac contraction and relaxation are sensed by the inflated cuff and transmitted via the tube to the internal pressure transducer, which provides the input signal for estimation of systolic and diastolic BP via the processing unit.
Fig. 1 Summary of components for the operation of an automated blood pressure measurement device (BPMD) using the oscillometric method. Cuff inflation and deflation is controlled by a microcomputer with a miniature air pump and valve system to change cuff pressure and facilitate BP measurements. Pressure waveforms generated by cardiac contraction and relaxation are sensed by the inflated cuff and transmitted via the tube to the internal pressure transducer, which provides the input signal for estimation of systolic and diastolic BP via the processing unit.CC BY · PubMed Central
Fig. 2 Overview of the oscillometric method to estimate blood pressure (BP) with superimposed example Korotkoff sounds phase I and V. The oscillometric waveform extracted from the cuff pressure curve is processed to construct the oscillometric waveform envelope, whereby the mean arterial pressure (MAP) is identified at the maximal amplitude of the envelope. Systolic BP (SBP) and diastolic BP (DBP) are then estimated using empirical fixed-ratio coefficients or other algorithms.
Fig. 2 Overview of the oscillometric method to estimate blood pressure (BP) with superimposed example Korotkoff sounds phase I and V. The oscillometric waveform extracted from the cuff pressure curve is processed to construct the oscillometric waveform envelope, whereby the mean arterial pressure (MAP) is identified at the maximal amplitude of the envelope. Systolic BP (SBP) and diastolic BP (DBP) are then estimated using empirical fixed-ratio coefficients or other algorithms.CC BY · PubMed Central
Fig. 3 Summary of the level of difference from intra-arterial brachial systolic and diastolic BP for cuff BP measured by sphygmomanometer or automated BP measuring device (BPMD). Data represent mean difference (95% confidence intervals) and are derived from individual participant level meta-analysis in ≥309 participants for BP measured by sphygmomanometer and ≥354 participants by automated BPMDs [38]. Example pulse pressures are denoted by the solid arrows. Green shaded areas represent indicative zones for aortic systolic and diastolic BP. The crosses indicate zones where non-invasive BP is markedly different from either brachial or aortic intra-arterial BPs, and the hatched arrow represents the potential underestimation of pulse pressure using an automated BPMD. Figure modified from [66].
Fig. 3 Summary of the level of difference from intra-arterial brachial systolic and diastolic BP for cuff BP measured by sphygmomanometer or automated BP measuring device (BPMD). Data represent mean difference (95% confidence intervals) and are derived from individual participant level meta-analysis in ≥309 participants for BP measured by sphygmomanometer and ≥354 participants by automated BPMDs [38]. Example pulse pressures are denoted by the solid arrows. Green shaded areas represent indicative zones for aortic systolic and diastolic BP. The crosses indicate zones where non-invasive BP is markedly different from either brachial or aortic intra-arterial BPs, and the hatched arrow represents the potential underestimation of pulse pressure using an automated BPMD. Figure modified from [66].CC BY · PubMed Central

Verification

Claims the paper makes

10 Valid · 2 Not valid · 14 Not checked · 5 Own results and interpretations

9 more claims

Create a free account to see all 31 claims and to verify the 4 references still open.

See all claims — free account
  • Probably not a substantial claim. 4
  • It may state two separate claims as one. 16

Create a free account to see these claims.

Citations

Cites

Show all 69 references

Research works printed but not linked: 2

  1. Birkenhager WH, Reid JL, editors. Handbook of hypertension. The Netherlands: Elsevier Science Publishers B.V.; 1991.
  2. Jilek J, Stork M. Arterial pulsations in the blood pressure cuff: are they hemodynamic pulses or oscillations? Int J Biol Biomed Eng. 2012;6:35–41.

Cited by

See all 84

Metadata

Authors
Sharman, J. E., Tan, I., Stergiou, G. S., Lombardi, C., Saladini, F., Butlin, M., Padwal, R., Asayama, K., Avolio, A., Brady, T. M., Murray, A., & Parati, G.
Affiliations
  • University of Tasmania (AU)
  • The George Institute for Global Health (AU)
  • National and Kapodistrian University of Athens (GR)
  • IRCCS Istituto Auxologico Italiano (IT)
  • Macquarie University (AU)
  • University of Alberta (CA)
  • Teikyo University (JP)
  • Johns Hopkins University (US)
  • Newcastle University (GB)
Corresponding author
Sharman, James E.
Venue
Journal of Human Hypertension
Published
2022-05-30
Volume
37
Issue
2
Pages
93-100
ISSN
  • 0950-9240 (ISSN-L)
  • 1476-5527
OpenAlex
W4281710849
Cited by
84
References
70
Citations by year
2026: 32 · 2025: 23 · 2024: 16 · 2023: 12 · 2022: 1
FWCI
2.64
Type
journal article
Language
en
Open access
Licence
Locations
  • Journal of Human Hypertension (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • BOA (University of Milano-Bicocca) (submittedVersion, cc-by)
  • Lirias (submittedVersion, other-oa)
  • Figshare (submittedVersion, cc-by)
  • Newcastle University ePrints (Newcastle Univesity) (cc-by)
Funders
Engineering and Physical Sciences Research Council (EP/N025342/1)
Topics
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
  • Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
Keywords
automated blood pressure measurement, oscillometric blood pressure devices, device validation, measurement accuracy, hypertension diagnosis, antihypertensive therapy, cardiovascular risk stratification, blood pressure thresholds, out-of-office blood pressure monitoring
MeSH terms
Blood Pressure, Blood Pressure Determination, Humans, Hypertension, Oscillometry, Sphygmomanometers

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

The claim verdicts this report counts are experimental: their accuracy is not yet measured.

A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).

A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).

Cited sources are checked with a language model’s help (gpt-6-sol and qwen3.8-omni-flash).

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure. https://app.cochat.ai/papers/10.1038/s41371-022-00693-x
The paper
Sharman, J. E., Tan, I., Stergiou, G. S., Lombardi, C., Saladini, F., Butlin, M., Padwal, R., Asayama, K., Avolio, A., Brady, T. M., Murray, A., & Parati, G. (2022). Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure. Journal of Human Hypertension. https://doi.org/10.1038/s41371-022-00693-x

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.