Evaluation of noninvasive blood pressure monitoring devices Spacelabs 90202 and 90207 versus resting and ambulatory 24-hour intra-arterial blood pressure.
Open accessThe DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.
Abstract
This study evaluated the accuracy of blood pressure values provided by the Spacelabs 90202 and 90207 devices in comparison with intra-arterial recording in 19 subjects at rest and in nine subjects in ambulatory conditions (Oxford method). At rest Spacelabs monitors reflected intra-arterial systolic blood pressure values very closely but overestimated to a considerable extent intra-arterial diastolic blood pressure (Spacelabs-intra-arterial differences, -0.8 +/- 9.2, NS, and 9.1 +/- 8.8 mm Hg, p less than 0.01, for systolic and diastolic blood pressures, respectively). In ambulatory conditions Spacelabs-intra-arterial average differences in 24-hour values were +0.4 +/- 5.1 mm Hg for systolic blood pressure (NS) and +14.0 +/- 2.9 mm Hg for diastolic blood pressure (p less than 0.01) when group data were considered. The performance of both Spacelabs devices was worse when assessed in individual subjects or for each hourly interval. In spite of these differences between noninvasive and intra-arterial absolute blood pressure values, however, Spacelabs 90202 and 90207 monitors were able to faithfully reflect directional hour-to-hour changes in intra-arterial blood pressure (chi 2 = 18.2 and chi 2 = 23.1 for systolic and diastolic blood pressures, respectively, p less than 0.01). No differences were found between the performance of the two Spacelabs devices. Thus, although the absolute accuracy of blood pressure values provided by these monitors in ambulatory subjects is still limited, they seem to be suitable for studies aimed at assessing 24-hour blood pressure profiles quantitatively as well as qualitatively.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Characterization of the oscillometric method for measuring indirect blood pressure
Geddes, L. A., Voelz, M., Combs, C., Reiner, D., & Babbs, C. F. (1982) · Annals of Biomedical Engineering · doi:10.1007/bf02367308 · Cited by 340
Continuous vs intermittent blood pressure measurements in estimating 24-hour average blood pressure.
di Rienzo, M., Grassi, G., Pedotti, A., & Mancia, G. (1983) · Hypertension · doi:10.1161/01.hyp.5.2.264 · Cited by 239
Self-citation
Cited by
Sega, R., Facchetti, R., Bombelli, M., Cesana, G., Corrao, G., Grassi, G., & Mancia, G. (2005) · Circulation · doi:10.1161/01.cir.0000160923.04524.5b · Cited by 1,009
Self-citationPrognostic accuracy of day versus night ambulatory blood pressure: a cohort study
Boggia, J., Li, Y., Thijs, L., Hansen, T. W., Kikuya, M., Björklund-Bodegård, K., Richart, T., Ohkubo, T., Kuznetsova, T., Torp-Pedersen, C., et al. (2007) · The Lancet · doi:10.1016/s0140-6736(07)61538-4 · Cited by 887
Mancia, G., Facchetti, R., Bombelli, M., Grassi, G., & Sega, R. (2006) · Hypertension · doi:10.1161/01.hyp.0000215363.69793.bb · Cited by 691
Self-citationLoneliness in everyday life: Cardiovascular activity, psychosocial context, and health behaviors.
Hawkley, L. C., Burleson, M. H., Berntson, G. G., & Cacioppo, J. T. (2003) · Journal of Personality and Social Psychology · doi:10.1037/0022-3514.85.1.105 · Cited by 608
Mancia, G., Zanchetti, A., Agebiti-Rosei, E., Benemio, G., De Cesaris, R., Fogari, R., Pessino, A., Porcellati, C., Salvetti, A., & Trimarco, B. (1997) · Circulation · doi:10.1161/01.cir.95.6.1464 · Cited by 596
Self-citation
Metadata
- Authors
- Groppelli, A., Omboni, S., Parati, G., & Mancia, G.
- Venue
- Hypertension
- Published
- 1992-08
- Volume
- 20
- Issue
- 2
- Pages
- 227-232
- ISSN
- 0194-911X (ISSN-L)
- 1524-4563
- OpenAlex
- W2129901270
- Cited by
- 157
- References
- 3
- FWCI
- 3.39
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Evaluation of noninvasive blood pressure monitoring devices Spacelabs 90202 and 90207 versus resting and ambulatory 24-hour intra-arterial blood pressure. https://app.cochat.ai/papers/10.1161/01.hyp.20.2.227
- The paper
- Groppelli, A., Omboni, S., Parati, G., & Mancia, G. (1992). Evaluation of noninvasive blood pressure monitoring devices Spacelabs 90202 and 90207 versus resting and ambulatory 24-hour intra-arterial blood pressure. Hypertension. https://doi.org/10.1161/01.hyp.20.2.227
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.