The DOI matches its Crossref record; 19 of the 19 references it cites were found in a registry or an index.
Continuous Wearable Monitoring Analytics Predict Heart Failure Hospitalization: The LINK-HF Multicenter Study
Abstract
BACKGROUND: Implantable cardiac sensors have shown promise in reducing rehospitalization for heart failure (HF), but the efficacy of noninvasive approaches has not been determined. The objective of this study was to determine the accuracy of noninvasive remote monitoring in predicting HF rehospitalization. METHODS: The LINK-HF study (Multisensor Non-invasive Remote Monitoring for Prediction of Heart Failure Exacerbation) examined the performance of a personalized analytical platform using continuous data streams to predict rehospitalization after HF admission. Study subjects were monitored for up to 3 months using a disposable multisensor patch placed on the chest that recorded physiological data. Data were uploaded continuously via smartphone to a cloud analytics platform. Machine learning was used to design a prognostic algorithm to detect HF exacerbation. Clinical events were formally adjudicated. RESULTS: One hundred subjects aged 68.4±10.2 years (98% male) were enrolled. After discharge, the analytical platform derived a personalized baseline model of expected physiological values. Differences between baseline model estimated vital signs and actual monitored values were used to trigger a clinical alert. There were 35 unplanned nontrauma hospitalization events, including 24 worsening HF events. The platform was able to detect precursors of hospitalization for HF exacerbation with 76% to 88% sensitivity and 85% specificity. Median time between initial alert and readmission was 6.5 (4.2-13.7) days. CONCLUSIONS: Multivariate physiological telemetry from a wearable sensor can provide accurate early detection of impending rehospitalization with a predictive accuracy comparable to implanted devices. The clinical efficacy and generalizability of this low-cost noninvasive approach to rehospitalization mitigation should be further tested. Registration: URL: https://www.clinicaltrials.gov. Unique Identifier: NCT03037710.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Multiparametric monitoring may be enabled with a patch worn for up to 3 months (Stehlik 2020).”
From 2021 ISHNE/HRS/EHRA/APHRS Expert Collaborative Statement on mHealth in Arrhythmia Management: Digital Medical Tools for Heart Rhythm Professionals: From the International Society for Holter and Noninvasive Electrocardiology/Heart Rhythm Society/European Heart Rhythm Association/Asia-Pacific Heart Rhythm Society (2021) · cites this paper as Stehlik 2020
In the citing paper
“The platform was able to successfully predict hospitalization for HF exacerbation with 76%-88% sensitivity and 85% specificity.”
From Unlocking Tomorrow’s Health Care: Expanding the Clinical Scope of Wearables by Applying Artificial Intelligence (2024) · cites this paper as [36]
Citations
Cites
Forecasting the Impact of Heart Failure in the United States
Heidenreich, P. A., Albert, N. M., Allen, L. A., Bluemke, D. A., Butler, J., Fonarow, G. C., Ikonomidis, J. S., Khavjou, O., Konstam, M. A., Maddox, T. M., et al. (2013) · Circulation Heart Failure · doi:10.1161/hhf.0b013e318291329a · Cited by 2,884
Self-citationAbraham, W. T., Adamson, P. B., Bourge, R. C., Aaron, M. F., Costanzo, M. R., Stevenson, L. W., Strickland, W., Neelagaru, S., Raval, N., Krueger, S., et al. (2011) · The Lancet · doi:10.1016/s0140-6736(11)60101-3 · Cited by 1,744
Telemonitoring in patients with heart failure
Chaudhry, S. I., Mattera, J. A., Curtis, J. P., Spertus, J. A., Herrin, J., Lin, Z., Phillips, C. O., Hodshon, B. V., Cooper, L. S., & Krumholz, H. M. (2010) · New England Journal of Medicine · doi:10.1056/nejmoa1010029 · Cited by 1,232
Artificial intelligence in cardiology
Johnson, K. W., Torres Soto, J., Glicksberg, B. S., Shameer, K., Miotto, R., Ali, M., Ashley, E., & Dudley, J. T. (2018) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2018.03.521 · Cited by 1,202
Intrathoracic Impedance Monitoring in Patients With Heart Failure
Yu, C., Wang, L., Chau, E. M. C., Chan, R. H.-W., Kong, S., TANG, M., Christensen, J., Stadler, R. W., & Lau, C. (2005) · Circulation · doi:10.1161/circulationaha.104.492207 · Cited by 708
Solomon, S. D., Dobson, J., Pocock, S., Skali, H., McMurray, J. J. V., Granger, C. B., Yusuf, S., Swedberg, K., Young, J. B., Michelson, E. L., et al. (2007) · Circulation · doi:10.1161/circulationaha.107.696906 · Cited by 681
Abraham, W. T., Stevenson, L. W., Bourge, R. C., Lindenfeld, J. A., Bauman, J. G., & Adamson, P. B. (2015) · The Lancet · doi:10.1016/s0140-6736(15)00723-0 · Cited by 621
van Veldhuisen, D. J., Braunschweig, F., Conraads, V. M., Ford, I., Cowie, M., Jondeau, G., Kautzner, J., Aguilera, R. M., Lunati, M., Yu, C., et al. (2011) · Circulation · doi:10.1161/circulationaha.111.043042 · Cited by 443
Gupta, A., Allen, L. A., Bhatt, D. L., Cox, M., DeVore, A. D., Heidenreich, P. A., Hernandez, A. F., Peterson, E. D., Matsouaka, R., Yancy, C. W., et al. (2017) · JAMA Cardiology · doi:10.1001/jamacardio.2017.4265 · Cited by 427
A Multisensor Algorithm Predicts Heart Failure Events in Patients With Implanted Devices
BOEHMER, J. P., Hariharan, R., Devecchi, F., Smith, A. L., Molon, G., Capucci, A., An, Q., Averina, V. A., Stolen, C., Thakur, P. H., et al. (2017) · JACC Heart Failure · doi:10.1016/j.jchf.2016.12.011 · Cited by 400
Research works printed but not linked: 1
- Mott J Pipke M. Similarity-based modeling of aircraft flight paths. Paper presented at 2004 IEEE Aerospace Conference Proceedings (IEEE Cat No04TH8720). March 6-13 2004 Big Sky MT USA. 2004;3:1580–1593.
Cited by
Sorted by citations
Gelder, I. C. V., Rienstra, M., Bunting, K. V., Casado-Arroyo, R., Caso, V., Crijns, H. J. G. M., Potter, T. J. R. D., Dwight, J., Guasti, L., Hanke, T., et al. (2024) · European Heart Journal · doi:10.1093/eurheartj/ehae176 · Cited by 3,594
Heart Failure After Myocardial Infarction: Incidence and Predictors
Jenča, D., Melenovský, V., Stehlik, J., Staněk, V., Kettner, J., Kautzner, J., Adámková, V., & Wohlfahrt, P. (2020) · ESC Heart Failure · doi:10.1002/ehf2.13144 · Cited by 644
Self-citationArmoundas, A. A., Narayan, S. M., Arnett, D. K., Spector-Bagdady, K., Bennett, D. A., Celi, L. A., Friedman, P. A., Gollob, M. H., Hall, J. L., Kwitek, A. E., et al. (2024) · Circulation · doi:10.1161/cir.0000000000001201 · Cited by 347
FDA Perspective on the Regulation of Artificial Intelligence in Health Care and Biomedicine
Warraich, H. J., Tazbaz, T., & Califf, R. M. (2025) · JAMA · doi:10.1001/jama.2024.21451 · Cited by 332
Wearable Devices in Cardiovascular Medicine
Hughes, A. N., Shandhi, M. M. H., Master, H., Dunn, J., & Brittain, E. L. (2023) · Circulation Research · doi:10.1161/circresaha.122.322389 · Cited by 341
Metadata
- Authors
- Stehlik, J., Schmalfuss, C., Bozkurt, B., Nativi-Nicolau, J., Wohlfahrt, P., Wegerich, S., Rose, K., Ray, R., Schofield, R., Deswal, A., Sekaric, J., Anand, S., Richards, D., Hanson, H., Pipke, M., & Pham, M.
- Affiliations
- Michael E. DeBakey VA Medical Center (US)
- Malcom Randall VA Medical Center (US)
- University of Utah (US)
- PhysIQ Inc, Chicago, IL (S.W., K.R., J.S., S.A., D.R., M.P.).
- VA Palo Alto Health Care System (US)
- George E. Wahlen Department of VA Medical Center (US)
- Corresponding author
- Stehlik, Josef
- ORCID
- Venue
- Circulation Heart Failure
- Published
- 2020-03
- Volume
- 13
- Issue
- 3
- Pages
- e006513-e006513
- ISSN
- 1941-3289 (ISSN-L)
- 1941-3297
- OpenAlex
- W3007492756
- Cited by
- 333
- References
- 19
- Citations by year
- 2026: 59 · 2025: 70 · 2024: 58 · 2023: 52 · 2022: 38 · 2021: 41 · 2020: 15
- FWCI
- 22.53
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Circulation Heart Failure (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Failure Treatment and Management
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Health Professions › Health Information Management › Artificial Intelligence in Healthcare
- Keywords
- heart failure, hospital readmission prediction, remote patient monitoring, wearable sensors, clinical alerts
- MeSH terms
- Cloud Computing, Smartphone, Machine Learning, Wearable Electronic Devices, Aged, Aged, 80 and over, Diagnosis, Computer-Assisted, Equipment Design, Female, Heart Failure, Humans, Male, Middle Aged, Patient Readmission, Predictive Value of Tests, Prognosis, Risk Factors, Telemetry, Time Factors, United States, Reproducibility of Results, Risk Assessment
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review