Strut Position, Blood Flow, and Drug Deposition: Implications for Single and Overlapping Drug-Eluting Stents
Open accessAbstract
Background— The intricacies of stent design, local pharmacology, tissue biology, and rheology preclude an intuitive understanding of drug distribution and deposition from drug-eluting stents (DES). Methods and Results— A coupled computational fluid dynamics and mass transfer model was applied to predict drug deposition for single and overlapping DES. Drug deposition appeared not only beneath regions of arterial contact with the strut but surprisingly also beneath standing drug pools created by strut disruption of flow. These regions correlated with areas of drug-induced fibrin deposition surrounding DES struts in porcine coronary arteries. Fibrin deposition immediately distal to individual isolated drug-eluting struts was twice as great as in the proximal area and for the stent as a whole was greater in distal segments than proximal segments. Adjacent and overlapping stent struts increased computed arterial drug deposition by far less than the sum of their combined drug load. In addition, drug eluted from the abluminal stent strut surface accounted for only 11% of total deposition, whereas, remarkably, drug eluted from the adluminal surface accounted for 43% of total deposition. Thus, local blood flow alterations and location of drug elution on the strut were far more important in determining arterial wall drug deposition and distribution than were drug load or arterial wall contact with coated strut surfaces. Conclusions— Simulations that coupled strut configurations with flow dynamics correlated with in vivo effects and revealed that drug deposition occurs less via contact between drug coating and the arterial wall than via flow-mediated deposition of blood-solubilized drug.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Chiu, J.-J., & Chien, S. (2011) · Physiological Reviews · doi:10.1152/physrev.00047.2009 · Cited by 2,280
Hemodynamic shear stress and the endothelium in cardiovascular pathophysiology
Davies, P. F. (2009) · Nature Clinical Practice Cardiovascular Medicine · doi:10.1038/ncpcardio1397 · Cited by 1,209
Vascular Responses to Drug Eluting Stents: Importance of Delayed Healing
Finn, A. V., Nakazawa, G., Joner, M., Kolodgie, F. D., Mont, E. K., Gold, H. K., & Virmani, R. (2007) · Arteriosclerosis, Thrombosis, and Vascular Biology · doi:10.1161/atvbaha.107.144220 · Cited by 900
In-Stent Restenosis in the Drug-Eluting Stent Era
Dangas, G. D., Claessen, B. E., Caixeta, A., Sanidas, E. A., Mintz, G. S., & Mehran, R. (2010) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2010.07.028 · Cited by 803
Serruys, P. W., Garcia-Garcia, H. M., & Onuma, Y. (2012) · European Heart Journal · doi:10.1093/eurheartj/ehr384 · Cited by 303
Metadata
- Authors
- Balakrishnan, B., Tzafriri, A. R., Seifert, P., Groothuis, A., Rogers, C., & Edelman, E. R.
- Venue
- Circulation
- Published
- 2005-06-07
- Volume
- 111
- Issue
- 22
- Pages
- 2958-2965
- ISSN
- 0009-7322 (ISSN-L)
- 1524-4539
- OpenAlex
- W2110036880
- Cited by
- 210
- References
- 0
- FWCI
- 9.71
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Strut Position, Blood Flow, and Drug Deposition: Implications for Single and Overlapping Drug-Eluting Stents. https://app.cochat.ai/papers/10.1161/circulationaha.104.512475
- The paper
- Balakrishnan, B., Tzafriri, A. R., Seifert, P., Groothuis, A., Rogers, C., & Edelman, E. R. (2005). Strut Position, Blood Flow, and Drug Deposition: Implications for Single and Overlapping Drug-Eluting Stents. Circulation. https://doi.org/10.1161/circulationaha.104.512475
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.