The DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Is Dual-Chamber Programming Inferior to Single-Chamber Programming in an Implantable Cardioverter-Defibrillator?: Results of the INTRINSIC RV (Inhibition of Unnecessary RV Pacing With AVSH in ICDs) Study
Open accessAbstract
Background— The INTRINSIC RV (Inhibition of Unnecessary RV Pacing with AVSH in ICDs) study tested the hypothesis that dual-chamber rate-responsive (DDDR) with atrioventricular search hysteresis (AVSH) 60-130 programming is not inferior to single-chamber (VVI)–40 programming in an implantable cardioverter defibrillator with respect to all-cause mortality and heart failure hospitalizations using an equivalence margin of 5%. Methods and Results— At 108 centers, 1530 patients with an implantable cardioverter defibrillator indication received a VITALITY AVT (Guidant Corporation, St. Paul, Minn) implantable cardioverter defibrillator programmed consistently to DDDR AVSH 60-130 for the first week. Of those, 988 patients with <20% right ventricular pacing at 1 week were randomized to DDDR AVSH 60-130 or to VVI-40 programming. Among those randomized, 502 were assigned to DDDR AVSH and 486 to VVI. Groups were similar with regard to coronary disease (68%), gender (21% female), and New York Heart Association functional class >I (79%). A total of 32 patients (6.4%) in the DDDR AVSH arm and 46 patients (9.5%) in the VVI arm died or were hospitalized for heart failure during a mean follow-up of 10.4 months (relative risk=0.67, P =0.072 in favor of DDDR AVSH). DDDR AVSH was not inferior to VVI programming ( P <0.001). All-cause mortality was not significantly different between the DDDR AVSH arm (3.6%) and the VVI arm (5.1%; P =0.23). The mean percent right ventricular pacing in the DDDR AVSH arm was 10% (median 4%) versus 3% (median 0%) in the VVI arm. Conclusions— In the INTRINSIC RV trial, among those randomized, DDDR AVSH was associated with similar outcomes as with VVI backup pacing.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
The DAVID Trial Investigators* (2002) · JAMA · doi:10.1001/jama.288.24.3115 · Cited by 2,056
Sweeney, M. O., Hellkamp, A. S., Ellenbogen, K. A., Greenspon, A. J., Freedman, R. A., Lee, K. L., & Lamas, G. A. (2003) · Circulation · doi:10.1161/01.cir.0000072769.17295.b1 · Cited by 1,773
Andersen, H. R., Nielsen, J. C., Thomsen, P. E. B., Thuesen, L., Mortensen, P. T., Vesterlund, T., & Pedersen, A. K. (1997) · The Lancet · doi:10.1016/s0140-6736(97)03425-9 · Cited by 841
Connolly, S. J., Kerr, C. R., Gent, M., Roberts, R. S., Yusuf, S., Gillis, A. M., Sami, M. H., Talajic, M., Tang, A. S. L., Klein, G. J., et al. (2000) · New England Journal of Medicine · doi:10.1056/nejm200005113421902 · Cited by 775
Cited by
Zeppenfeld, K., Tfelt-Hansen, J., de Riva, M., Winkel, B. G., Behr, E. R., Blom, N. A., Charron, P., Corrado, D., Dagres, N., de Chillou, C., et al. (2022) · European Heart Journal · doi:10.1093/eurheartj/ehac262 · Cited by 3,140
ACC/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities
Epstein, A. E., DiMarco, J. P., Ellenbogen, K. A., Estes, N. A. M., Freedman, R. A., Gettes, L. S., Gillinov, A. M., Gregoratos, G., Hammill, S. C., Hayes, D. L., et al. (2008) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2008.02.032 · Cited by 2,546
Writing Committee Members, Epstein, A. E., DiMarco, J. P., Ellenbogen, K. A., Estes, N. A. M., Freedman, R. A., Gettes, L. S., Gillinov, A. M., Gregoratos, G., Hammill, S. C., et al. (2008) · Circulation · doi:10.1161/circualtionaha.108.189742 · Cited by 1,774
Epstein, A. E., DiMarco, J. P., Ellenbogen, K. A., Estes, N. A. M., Freedman, R. A., Gettes, L. S., Gillinov, A. M., Gregoratos, G., Hammill, S. C., Hayes, D. L., et al. (2013) · Circulation · doi:10.1161/cir.0b013e318276ce9b · Cited by 1,489
Tracy, C. M., Epstein, A. E., Darbar, D., DiMarco, J. P., Dunbar, S. B., Estes, N. A. M., Ferguson, T. B., Hammill, S. C., Karasik, P. E., Link, M. S., et al. (2013) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2012.11.007 · Cited by 732
Metadata
- Authors
- Olshansky, B., Day, J. D., Moore, S., Gering, L., Rosenbaum, M., McGuire, M., Brown, S., & Lerew, D. R.
- Venue
- Circulation
- Published
- 2007-01-02
- Volume
- 115
- Issue
- 1
- Pages
- 9-16
- ISSN
- 0009-7322 (ISSN-L)
- 1524-4539
- OpenAlex
- W2105152904
- Cited by
- 169
- References
- 4
- FWCI
- 8.97
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Is Dual-Chamber Programming Inferior to Single-Chamber Programming in an Implantable Cardioverter-Defibrillator?: Results of the INTRINSIC RV (Inhibition of Unnecessary RV Pacing With AVSH in ICDs) Study. https://app.cochat.ai/papers/10.1161/circulationaha.106.629428
- The paper
- Olshansky, B., Day, J. D., Moore, S., Gering, L., Rosenbaum, M., McGuire, M., Brown, S., & Lerew, D. R. (2007). Is Dual-Chamber Programming Inferior to Single-Chamber Programming in an Implantable Cardioverter-Defibrillator?: Results of the INTRINSIC RV (Inhibition of Unnecessary RV Pacing With AVSH in ICDs) Study. Circulation. https://doi.org/10.1161/circulationaha.106.629428
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.