Verification of a novel point-of-care HbA1c device in real world clinical practice by comparison to three high performance liquid chromatography instruments
Open accessAbstract
Introduction: A real world clinical study was designed and conducted to evaluate the performance of a novel point-of-care device for determination of glycated haemoglobin A 1c (HbA 1c ), A1C EZ 2.0, in daily clinical practice.Materials and methods: Five hundred and fourteen subjects were included in this study, and divided into three groups.HbA 1c was measured by A1C EZ 2.0 and three different high performance liquid chromatography (HPLC) devices: Bio-Rad Variant II Turbo, Tosoh HLC-723 G8 and Premier Hb9210 separately.Precision of A1C EZ 2.0 was also evaluated.Results: Results obtained from A1C EZ 2.0 and all HPLC devices are correlated.Passing-Bablok regression analysis shows the equation of A1C EZ 2.0 results against the mean of HPLC devices with corresponding 95% confidence intervals (95% CI) for the intercept and slope is y = 0.10 (-0.17 to 0.10) + 1.00 (1.00 to 1.04) x.Bland-Altman difference plot shows that the mean relative difference between A1C EZ 2.0 and Variant II Turbo, G8, Hb9210 and all HPLC results is 2.5%, 0.6%, 0.4% and 1.1%, respectively.In addition, 121 pairs of results determined by using both venous and capillary blood prove that the difference of two kinds of blood sample causes no notable variation when measured by A1C EZ 2.0.Precision study gives 2.3% and 1.9% of total coefficient of variation for normal and abnormal HbA 1c sample in A1C EZ 2.0.Conclusions: HbA 1c values measured by A1C EZ 2.0 were in good accordance with the results obtained with the reference HPLC devices.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The systematic review and meta-analysis included 11 studies involving 1,593 participants and 3,285 paired HbA1c measurements comparing POC devices with reference HPLC-based laboratory methods.
“A total of 11 studies involving 1,593 participants and 3,285 paired measurements were included.”
From Comparison of point‑of‑care HbA1c devices with laboratory HPLC methods: A systematic review and meta‑analysis (2026) · cites this paper as [27]
The Materials and methods (Study design) describe a single analytical verification study, not a systematic review or meta-analysis. Statistical analysis reports 514 subjects and 635 pairs of data, contrary to the attributed totals of 11 studies, 1,593 participants and 3,285 paired measurements. Page numbers are unavailable in our copy. (LLM, conf=0.90)
Checked by a reasoning model — may be wrong.
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Effects of hemoglobin variants and chemically modified derivatives on assays for glycohemoglobin
Bry, L., Chen, P. C., & Sacks, D. B. (2001) · Clinical Chemistry · doi:10.1093/clinchem/47.2.153 · Cited by 523 · see citing works
Cagliero, E., Levina, E. V., & Nathan, D. M. (1999) · Diabetes Care · doi:10.2337/diacare.22.11.1785 · Cited by 246 · see citing works
Screening for Diabetes and Pre-Diabetes With Proposed A1C-Based Diagnostic Criteria
Olson, D. E., Rhee, M. K., Herrick, K., Ziemer, D. C., Twombly, J. G., & Phillips, L. S. (2010) · Diabetes Care · doi:10.2337/dc10-0433 · Cited by 234 · see citing works
Lenters-Westra, E., & Slingerland, R. J. (2010) · Clinical Chemistry · doi:10.1373/clinchem.2009.130641 · Cited by 228 · see citing works
Flagged first, then by citations
Research works printed but not linked: 19
- [1]Xu Y Wang L He J Bi Y Li M Wang T Prevalence and control of diabetes in Chinese adults. JAMA. 2013; 310: 948– 59. 10.1001/jama.2013.168118 24002281 · DOI
- [2]Weykamp C John WG Mosca A A review of the challenge in measuring hemoglobin A1c. J Diabetes Sci Technol. 2009; 3: 439– 45. 10.1177/193229680900300306 20144280 PMC2769874 · DOI
- [3]Nathan DM Singer DE Hurxthal K Goodson JD The clinical information value of the glycosylated hemoglobin assay. N Engl J Med. 1984; 310: 341– 6. 10.1056/NEJM198402093100602 6690962 · DOI
- [4]Dreval’ AV Cheporeva ON Red’kin IuA Misnikova IV Correlation of HBA(1c) and postprandial glycemia during a standard breakfast test in patients with type 2 diabetes mellitus receiving glibenclamide. Ter Arkh. 2010; 82: 41– 4. [in Russian] 20873244
- [6]Ferenczi A Reddy K Lorber DL Effect of immediate hemoglobin A1c results on treatment decisions in office practice. Endocr Pract. 2001; 7: 85– 8. 10.4158/EP.7.2.85 11421550 · DOI
- [7]Miller CD Barnes CS Phillips LS Ziemer DC Gallina DL Cook CB Rapid A1c availability improves clinical decision-making in an urban primary care clinic. Diabetes Care. 2003; 26: 1158– 63. 10.2337/diacare.26.4.1158 12663590 · DOI
- [9]Lenters-Westra E Slingerland RJ Three of 7 hemoglobin A1c point-of-care instruments do not meet generally accepted analytical performance criteria. Clin Chem. 2014; 60: 1062– 72. 10.1373/clinchem.2014.224311 24865164 · DOI
- [10]Zhou R Wang W Song ZX Tong Q Wang QT Evaluation of a new hemoglobin A1c analyzer for point-of-care testing. J Clin Lab Anal. 2018; 32: e22172. 10.1002/jcla.22172 28220976 PMC6817028 · DOI
- [11]Bilić-Zulle L Comparison of methods: Passing and Bablok regression. Biochem Med (Zagreb). 2011; 21: 49– 52. 10.11613/BM.2011.010 22141206 · DOI · Paper page
- [12]Altman DG Bland JM Measurement in medicine: the analysis of method comparison studies. Statistician. 1983; 32: 307– 17. 10.2307/2987937 · DOI · Paper page
- [14]Lin CN Emery TJ Little RR Hanson SE Rohlfing CL Jaisson S Effects of hemoglobin C, D, E, and S traits on measurements of HbA1c by six methods. Clin Chim Acta. 2012; 413: 819– 21. 10.1016/j.cca.2011.12.019 22244931 PMC5068911 · DOI
- [15]Kupke IR Kather B Zeugner S On the composition of capillary and venous blood serum. Clin Chim Acta. 1981; 112: 177– 85. 10.1016/0009-8981(81)90376-4 7237825 · DOI
- [16]Font MT Brichs MC Alvarez MC Olivella JM Turo JS Fernandez MP Capillary HbA1c determination on type 2 diabetes patients in a primary health centre. Aten Primaria. 2011; 43: 536– 43. 21411189 PMC7024957
- [17]Valdivia Pérez A Rico Blazquez M Luis Martin F Mingo Plaza D Lozano Garcia I Pizarro Blanco M Assessment of HbA1c in capillary blood compared to venous blood. Aten Primaria. 2010; 42: 126– 7, author reply 128. 20141867 PMC7022120
- [18]Heylen O Van Neyghem S Exterbille S Wehlou C Gorus F Weets I Evaluation of the Sebia CAPILLARYS 2 flex piercing for the measurement of HbA(1c) on venous and capillary blood samples. Am J Clin Pathol. 2014; 141: 867– 77. 10.1309/AJCPRU5QC2JBANSV 24838332 · DOI
- [19]Keramati T Razi F Tootee A Larijani B Comparability of hemoglobin A1c level measured in capillary versus venous blood sample applying two point-of-care instruments. J Diabetes Metab Disord. 2014; 13: 94. 10.1186/s40200-014-0094-1 25408924 PMC4234901 · DOI
- [20]Eldeirawi K Lipton RB Predictors of hemoglobin A1c in a national sample of nondiabetic children: the Third National Health and Nutrition Examination Survey, 1988-1994. Am J Epidemiol. 2003; 157: 624– 32. 10.1093/aje/kwg023 12672682 · DOI
- [22]Bao Y Ma X Li H Zhou M Hu C Wu H Glycated haemoglobin A1c for diagnosing diabetes in Chinese population: cross sectional epidemiological survey. BMJ. 2010; 340: c2249. 10.1136/bmj.c2249 20478961 PMC2871988 · DOI
- [23]Xin Z Yuan MX Li HX Hua L Feng JP Shi J Evaluation for fasting and 2-hour glucose and HbA1c for diagnosing diabetes based on prevalence of retinopathy in a Chinese population. PLoS One. 2012; 7: e40610. 10.1371/journal.pone.0040610 22808204 PMC3395674 · DOI
Cited by
Sorted by citations
Wei, Y., Gu, M., Ren, Y., & Hu, Z. (2026) · Biomedical Reports · doi:10.3892/br.2026.2120 · Cited by 3 · see citing works
- Authors
- Wang, Y., Peng, W., Tang, J., Dong, L., Gu, C., Zhang, X., Zhou, J., & Jia, W.
- Affiliations
- Shanghai Jiao Tong University (CN)
- Wuxi Municipal Bureau on Science and Technology (CN)
- Corresponding author
- Zhou, Jian
- Venue
- Biochemia Medica
- Published
- 2018-04-15
- Volume
- 28
- Issue
- 2
- Pages
- 020705-020705
- ISSN
- 1330-0962 (ISSN-L)
- 1846-7482
- PMC
- PMC5898960
- OpenAlex
- W2797485890
- Cited by
- 11
- References
- 4
- Citations by year
- 2026: 2 · 2024: 2 · 2023: 3 · 2022: 1 · 2021: 1 · 2019: 2
- FWCI
- 0.52
- Type
- journal article
- Language
- en
- Open access
- diamond
- doi.org
- Licence
- cc-by
- Locations
- Biochemia Medica (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, other-oa)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Clinical Biochemistry › Advanced Glycation End Products research
- Keywords
- HbA1c, point-of-care testing, HPLC, method comparison, diabetes monitoring, capillary blood, venous blood, measurement precision, real-world clinical practice
- MeSH terms
- Adolescent, Adult, Aged, Aged, 80 and over, Analysis of Variance, Chromatography, High Pressure Liquid, Diabetes Mellitus, Female, Glycated Hemoglobin, Humans, Male, Middle Aged, Sensitivity and Specificity, Reproducibility of Results, Case-Control Studies, Point-of-Care Systems, Automation, Laboratory
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review