Accuracy of two ear oximeters at rest and during exercise in pulmonary patients.

Andrew L. Ries, John T. Farrow, Jack L. Clausen · 1985 · PubMed
Cited by 106

The DOI matches its OpenAlex record.

Abstract

The accuracy of 2 ear oximeters (Hewlett-Packard 47201A and Biox IIA) was evaluated at rest and during exercise in 116 pulmonary patients. Ear oximetry measurements were compared with arterial oxygen saturation (SaO2) measured from simultaneous arterial blood samples. When used according to the manufacturer's instructions for stabilization during exercise testing, both oximeters demonstrated comparable accuracy in estimating arterial blood SaO2 both at rest and during exercise in these patients (95% confidence limits, 2 SEE = +/- 4 to 5%). The oximeters were more accurate in measuring a change in SaO2 from a previous reading (95% confidence limits, 2 SEE = +/- 2.5 to 3.5%). COHgb levels greater than 4% and dark skin pigmentation appeared to decrease the accuracy of ear oximetry measurements. Within the limitations of accuracy, both ear oximeters may be useful for clinical exercise testing in pulmonary patients.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

OpenAlex · 10.1164/arrd.1985.132.3.685

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 106

Metadata

Authors
Ries, A. L., Farrow, J. T., & Clausen, J. L.
Venue
PubMed
Publisher
National Institutes of Health
Published
1985-09-01
Volume
132
Issue
3
Pages
685-9
OpenAlex
W1676735874
Cited by
106
References
0
FWCI
9.03
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Accuracy of two ear oximeters at rest and during exercise in pulmonary patients. https://app.cochat.ai/papers/10.1164/arrd.1985.132.3.685
The paper
Ries, A. L., Farrow, J. T., & Clausen, J. L. (1985). Accuracy of two ear oximeters at rest and during exercise in pulmonary patients. PubMed. https://doi.org/10.1164/arrd.1985.132.3.685

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.