Examination of the Carbon Monoxide Diffusing Capacity (D l CO ) in Relation to Its K co and V a Components
Abstract
The single-breath carbon monoxide diffusing capacity (DL(CO)) is the product of two measurements during breath holding at full inflation: (1) the rate constant for carbon monoxide uptake from alveolar gas (kco [minute(-1)]) and (2) the "accessible" alveolar volume (Va). kco expressed per mm Hg alveolar dry gas pressure (Pb*) as kco/Pb*, and then multiplied by Va, equals Dl(CO); thus, Dl(CO) divided by Va (DL(CO)/Va, also called Kco) is only kco/Pb* in different units, remaining, essentially, a rate constant. The notion that DL(CO)/Va "corrects" DL(CO) for reduced Va is physiologically incorrect, because DL(CO)/Va is not constant as Va changes; thus, the term Kco reflects the physiology more appropriately. Crucially, the same DL(CO) may occur with various combinations of Kco and Va, each suggesting different pathologies. Decreased Kco occurs in alveolar-capillary damage, microvascular pathology, or anemia. Increased Kco occurs with (1) failure to expand normal lungs to predicted full inflation (extrapulmonary restriction); or (2) increased capillary volume and flow, either globally (left-to-right intracardiac shunting) or from flow and volume diversion from lost or damaged units to surviving normal units (e.g., pneumonectomy). Decreased Va occurs in (1) reduced alveolar expansion, (2) alveolar damage or loss, or (3) maldistribution of inspired gases with airflow obstruction. Kco will be greater than 120% predicted in case 1, 100-120% in case 2, and 40-120% in case 3, depending on pathology. Kco and Va values should be available to clinicians, as fundamental to understanding the clinical implications of DL(CO). The diffusing capacity for nitric oxide (DL(NO)), and the DL(NO)/DL(CO) ratio, provide additional insights.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Interpretative strategies for lung function tests
Pellegrino, R., Viegi, G., Brusasco, V., Crapo, R. O., Burgos, F., Casaburi, R., Coates, A., van der Grinten, C. P. M., Gustafsson, P., Hankinson, J., et al. (2005) · European Respiratory Journal · doi:10.1183/09031936.05.00035205 · Cited by 5,945
Standardisation of the single-breath determination of carbon monoxide uptake in the lung
MacIntyre, N., Crapo, R. O., Viegi, G., Johnson, D. C., van der Grinten, C. P. M., Brusasco, V., Burgos, F., Casaburi, R., Coates, A., Enright, P., et al. (2005) · European Respiratory Journal · doi:10.1183/09031936.05.00034905 · Cited by 2,255
Ogilvie, C. M., Forster, R. E., Blakemore, W. S., & Morton, J. W. (1957) · Journal of Clinical Investigation · doi:10.1172/jci103402 · Cited by 1,201
Roughton, F. J. W., & Forster, R. E. (1957) · Journal of Applied Physiology · doi:10.1152/jappl.1957.11.2.290 · Cited by 999
Hepatopulmonary Syndrome — A Liver-Induced Lung Vascular Disorder
Rodríguez-Roisin, R., & Krowka, M. J. (2008) · New England Journal of Medicine · doi:10.1056/nejmra0707185 · Cited by 687
Standardization of the measurement of transfer factor (diffusing capacity)
Cotes, J. E., Chinn, D. J., Quanjer, P. H., Roca, J., & Yernault, J.-C. (1993) · European Respiratory Journal · doi:10.1183/09041950.041s1693 · Cited by 680
Steen, V., & Medsger, T. A. (2003) · Arthritis & Rheumatism · doi:10.1002/art.10775 · Cited by 494
The diffusion of gases through the lungs of man
Krogh, M. (1915) · The Journal of Physiology · doi:10.1113/jphysiol.1915.sp001710 · Cited by 432
Measurements of capillary dimensions and blood volume in rapidly frozen lungs.
Glazier, J. B., Hughes, J. M., Maloney, J. E., & West, J. B. (1969) · Journal of Applied Physiology · doi:10.1152/jappl.1969.26.1.65 · Cited by 249
Self-citationPuri, S., Baker, B. L., Dutka, D. P., Oakley, C. M., Hughes, J. M. B., & Cleland, J. G. F. (1995) · Circulation · doi:10.1161/01.cir.91.11.2769 · Cited by 248
Self-citationBorland, C., & Higenbottam, T. (1989) · European Respiratory Journal · doi:10.1183/09031936.93.02010056 · Cited by 222
Measurements of capillary dimensions and blood volume in rapidly frozen lungs
(1969) · The American Journal of Cardiology · doi:10.1016/0002-9149(69)90087-3 · Cited by 221
Guenard, H., Varene, N., & Vaida, P. (1987) · Respiration Physiology · doi:10.1016/s0034-5687(87)80036-1 · Cited by 214
McGrath, M. W., & Thomson, M. L. (1959) · The Journal of Physiology · doi:10.1113/jphysiol.1959.sp006212 · Cited by 180
Carlsen, E., & Comroe, J. H. (1958) · The Journal of General Physiology · doi:10.1085/jgp.42.1.83 · Cited by 175
Lung function in white children aged 4 to 19 years: II--Single breath analysis and plethysmography.
Rosenthal, M., Cramer, D., Bain, S. H., Denison, D., Bush, A., & Warner, J. O. (1993) · Thorax · doi:10.1136/thx.48.8.803 · Cited by 171
European reference equations for CO and NO lung transfer
Aguilaniu, B., Maitre, J., Glénet, S. N., Gégout‐Petit, A., & Guénard, H. J.-P. (2008) · European Respiratory Journal · doi:10.1183/09031936.00063207 · Cited by 128
Johnson, D. C. (2000) · Respiratory Medicine · doi:10.1053/rmed.1999.0740 · Cited by 121
Clinical Significance of Elevated Diffusing Capacity
Saydain, G., Beck, K. C., Decker, P. A., Cowl, C. T., & Scanlon, P. D. (2004) · CHEST Journal · doi:10.1378/chest.125.2.446 · Cited by 120
Hughes, J. M. B., & Bates, D. V. (2003) · Respiratory Physiology & Neurobiology · doi:10.1016/j.resp.2003.08.004 · Cited by 117
Self-citationAlveolar-Capillary Membrane Dysfunction in Heart Failure
Guazzi, M. (2003) · CHEST Journal · doi:10.1378/chest.124.3.1090 · Cited by 109
In defence of the carbon monoxide transfer coefficientKCO(TL/VA)
Hughes, J. M. B., & Pride, N. B. (2001) · European Respiratory Journal · doi:10.1183/09031936.01.17201680 · Cited by 109
Self-citationReference Values of Pulmonary Diffusing Capacity During Exercise by a Rebreathing Technique
Hsia, C. C., McBrayer, D. G., & Ramanathan, M. (1995) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/ajrccm.152.2.7633723 · Cited by 95
Pulmonary function in sickle cell disease with or without acute chest syndrome
Santoli, F., Zerah, F., Vasile, N., Bachir, D., Galacteros, F., & Atlan, G. (1998) · European Respiratory Journal · doi:10.1183/09031936.98.12051124 · Cited by 92
Effect of lung volume and positional changes on pulmonary diffusing capacity and its components
Stam, H. J., Kreuzer, F., & Versprille, A. (1991) · Journal of Applied Physiology · doi:10.1152/jappl.1991.71.4.1477 · Cited by 86
Single Breath Diffusing Capacity for Carbon Monoxide in Stable Asthma
Collard, P., Njinou, B., Nejadnik, B., Keyeux, A., & Frans, A. (1994) · CHEST Journal · doi:10.1378/chest.105.5.1426 · Cited by 81
Significant blood resistance to nitric oxide transfer in the lung
Borland, C. D. R., Dunningham, H., Bottrill, F. E., Vuylsteke, A., Yılmaz, C., Dane, D. M., & Hsia, C. C. W. (2010) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00904.2009 · Cited by 80
Reference values of pulmonary diffusing capacity for nitric oxide in an adult population
Zavorsky, G. S., Cao, J., & Murias, J. M. (2007) · Nitric Oxide · doi:10.1016/j.niox.2007.10.002 · Cited by 74
Zavorsky, G. S., Quiron, K. B., Massarelli, P. S., & Lands, L. C. (2004) · CHEST Journal · doi:10.1378/chest.125.3.1019 · Cited by 74
Diffusing capacity for nitric oxide: Reference values and dependence on alveolar volume
van der Lee, I., Zanen, P., Stigter, N., et al. (2007) · Respiratory Medicine · doi:10.1016/j.rmed.2006.12.001 · Cited by 67
Serial Estimations of Carbon Monoxide Diffusing Capacity in Intrapulmonary Haemorrhage
Greening, A. P., & Hughes, J. M. B. (1981) · Clinical Science · doi:10.1042/cs0600507 · Cited by 63
Self-citationDeciphering the nitric oxide to carbon monoxide lung transfer ratio: physiological implications
Glénet, S. N., de Bisschop, C., Vargas, F., & Guénard, H. J.-P. (2007) · The Journal of Physiology · doi:10.1113/jphysiol.2007.133405 · Cited by 56
Chinn, D., Je, C., Flowers, R., Marks, A., & Reed, J. (1996) · European Respiratory Journal · doi:10.1183/09031936.96.09061269 · Cited by 56
Effect of alveolar volume on the interpretation of single breath DlCO
Frans, A., Nemery, B., Veriter, C., Lacquet, L. M., & Francis, C. (1997) · Respiratory Medicine · doi:10.1016/s0954-6111(97)90029-9 · Cited by 54
Nitric oxide diffusing capacity versus spirometry in the early diagnosis of emphysema in smokers
van der Lee, I., Gietema, H. A., Zanen, P., et al. (2009) · Respiratory Medicine · doi:10.1016/j.rmed.2009.06.005 · Cited by 40
Hart, N., Cramer, D., Ward, S., et al. (2002) · European Respiratory Journal · doi:10.1183/09031936.00.00286702 · Cited by 40
Evaluation of Diffusing Capacity in Patients With a Restrictive Lung Disease
Stam, H. J., Splinter, T. A. W., & Versprille, A. (2000) · CHEST Journal · doi:10.1378/chest.117.3.752 · Cited by 38
Effect of Circulatory Congestion on the Components of Pulmonary Diffusing Capacity in Morbid Obesity
Oppenheimer, B. W., Berger, K. I., Rennert, D. A., Pierson, R. N., Norman, R. G., Rapoport, D. M., Kral, J. G., & Goldring, R. M. (2006) · Obesity · doi:10.1038/oby.2006.134 · Cited by 37
Can a membrane oxygenator be a model for lung NO and CO transfer?
Borland, C. D. R., Dunningham, H., Bottrill, F. E., & Vuylsteke, A. (2006) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00949.2005 · Cited by 28
Transfer factor for carbon monoxide: a glance behind the scene
Fitting, J.-W. (2004) · Swiss Medical Weekly · doi:10.4414/smw.2004.10660 · Cited by 27
van der Lee, I., Zanen, P., van den Bosch, J. M. M., et al. (2005) · Respiratory Medicine · doi:10.1016/j.rmed.2005.04.014 · Cited by 26
Carbon monoxide transfer coefficient KCO (TL/VA): a flawed index L
Cotes, J. E., Hughes, J. M. B., & Pride, N. B. (2001) · European Respiratory Journal · doi:10.1183/09031936.01.00246701 · Cited by 15
Self-citation
Cited by
ERS/ATS technical standard on interpretive strategies for routine lung function tests
Stanojevic, S., Kaminsky, D. A., Miller, M. R., Thompson, B., Aliverti, A., Barjaktarevic, I., Cooper, B. G., Culver, B., Derom, E., Hall, G. L., et al. (2022) · European Respiratory Journal · doi:10.1183/13993003.01499-2021 · Cited by 1,656
2017 ERS/ATS standards for single-breath carbon monoxide uptake in the lung
Graham, B. L., Brusasco, V., Burgos, F., Cooper, B. G., Jensen, R., Kendrick, A., MacIntyre, N. R., Thompson, B. R., & Wanger, J. (2017) · European Respiratory Journal · doi:10.1183/13993003.00016-2016 · Cited by 972
Stanojevic, S., Graham, B. L., Cooper, B. G., Thompson, B. R., Carter, K. W., Francis, R. W., & Hall, G. L. (2017) · European Respiratory Journal · doi:10.1183/13993003.00010-2017 · Cited by 676
Metadata
- Authors
- Hughes, J. M. B., & Pride, N. B.
- Venue
- American Journal of Respiratory and Critical Care Medicine
- Publisher
- Oxford University Press (OUP)
- Published
- 2012-07-15
- Volume
- 186
- Issue
- 2
- Pages
- 132-139
- ISSN
- 1073-449X (ISSN-L)
- 1535-4970
- OpenAlex
- W2153769722
- Cited by
- 259
- References
- 42
- FWCI
- 11.15
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.