Comparison of Dynamic Contour Tonometry with Goldmann Applanation Tonometry
Abstract
PURPOSE: The dynamic contour tonometer (DCT; Pascal tonometer) is a novel tonometer designed to measure intraocular pressure (IOP) independent of corneal properties. The purpose of this study was a comparison of the DCT with the Goldmann applanation tonometer (GAT) with respect to mean of IOP readings, the influence of ocular structural factors on IOP readings, and both intra- and interobserver variability, in a large group of healthy subjects. METHODS: In a prospective study of 228 eyes, IOP measurements by GAT and DCT were compared, and the effects of central corneal thickness (CCT), corneal curvature, axial length, and anterior chamber depth were analyzed. To evaluate intra- and interobserver variability, IOP was measured in eight eyes by four observers. RESULTS: There was a high concordance between the IOP readings obtained by DCT and GAT. However, IOP readings were consistently higher with DCT than with GAT (median difference: +1.7 mm Hg, interquartile range [25th-75th percentile] = 0.8-2.7 mm Hg). In contrast to GAT, multivariable regression analysis showed no significant effect of corneal thickness, corneal curvature, astigmatism, anterior chamber depth, and axial length on DCT readings. For repeated measurements the intraobserver variability was 0.65 mm Hg for the DCT and 1.1 mm Hg for the GAT (P = 0.008). Interobserver variability was 0.44 mm Hg for the DCT and 1.28 mm Hg for the GAT (P = 0.017). CONCLUSIONS: IOP measurements by DCT are highly concordant with IOP readings obtained from GAT but do not vary in CCT and have a lower intra- and interobserver variability. DCT seems to be an appropriate method of tonometry for routine clinical use.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
APPLANATION TONOMETRY AND CENTRAL CORNEAL THICKNESS
EHLERS, N., BRAMSEN, T., & SPERLING, S. (1975) · Acta Ophthalmologica · doi:10.1111/j.1755-3768.1975.tb01135.x · Cited by 931
Sources of error with use of Goldmann-type tonometers
Whitacre, M. M., & Stein, R. (1993) · Survey of Ophthalmology · doi:10.1016/0039-6257(93)90053-a · Cited by 747
Central corneal thickness in the ocular hypertension treatment study (OHTS)
Brandt, J. D., Beiser, J. A., Kass, M. A., & Gordon, M. O. (2001) · Ophthalmology · doi:10.1016/s0161-6420(01)00760-6 · Cited by 626
Feltgen, N. (2001) · British Journal of Ophthalmology · doi:10.1136/bjo.85.1.85 · Cited by 238
Intraocular Pressure Measurements Using Dynamic Contour Tonometry after Laser In Situ Keratomileusis
Kaufmann, C., Bachmann, L. M., & Thiel, M. A. (2003) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.02-0946 · Cited by 214
Self-citationDielemans, I., Vingerling, J. R., Hofman, A., Grobbee, D. E., & de Jong, P. T. V. M. (1994) · Graefe's Archive for Clinical and Experimental Ophthalmology · doi:10.1007/bf00176782 · Cited by 174
Accuracy of Goldmann tonometry in clinical practice
Sudesh, S., Moseley, M. J., & Thompson, J. R. (1993) · Acta Ophthalmologica · doi:10.1111/j.1755-3768.1993.tb04988.x · Cited by 62
Munger, R., Hodge, W., Mintsioulis, G., Agapitos, P. J., Jackson, W. B., & Damji, K. F. (1998) · PubMed · Cited by 50
Cited by
Tonnu, P.-A. (2005) · British Journal of Ophthalmology · doi:10.1136/bjo.2004.056622 · Cited by 357
Surveillance for ocular hypertension: an evidence synthesis and economic evaluation.
Burr, J., Botello-Pinzon, P., Takwoingi, Y., Hernández, R., Vazquez-Montes, M., Elders, A., Asaoka, R., Banister, K., van der Schoot, J., Fraser, C., et al. (2012) · Health Technology Assessment · doi:10.3310/hta16290 · Cited by 180
Ocular Pulse Amplitude in Healthy Subjects as Measured by Dynamic Contour Tonometry
Kaufmann, C. (2006) · Archives of Ophthalmology · doi:10.1001/archopht.124.8.1104 · Cited by 154
Self-citation
Metadata
- Authors
- Kaufmann, C., Bachmann, L. M., & Thiel, M. A.
- Venue
- Investigative Opthalmology & Visual Science
- Published
- 2004-09-01
- Volume
- 45
- Issue
- 9
- Pages
- 3118
- ISSN
- 0146-0404 (ISSN-L)
- 1552-5783
- OpenAlex
- W2065609053
- Cited by
- 281
- References
- 8
- FWCI
- 16.69
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.