Temporal Scanner Thermometry: A New Method of Core Temperature Estimation in ICU Patients
The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
Background and Aims: Temperature measurement is a routine task of patient care, with considerable clinical impact, especially in the ICU. This study was conducted to evaluate the accuracy and variability of the Temporal Artery Thermometer (TAT) in ICU-patients. Therefore, a convenience sample of 57 adult patients, with indwelling pulmonary artery catheters (PAC) in a 40-bed intensive care unit in a university teaching hospital was used. Methods: The study design was a prospective, descriptive comparative analysis. Body temperature was thereby measured simultaneously with the TAT and the Axillary Thermometer (AT), and was compared with the temperature recording of the PAC. The use of vasoactive medication was recorded. Results and conclusions: Mean temperature of all measurements was: PAC: 37.1°C (SD: 0.87), TAT: 37.0°C (SD: 0.68) and axillary thermometer: 36.6°C (SD: 0.94). The measurements of the TAT and the PAC were not significantly different (mean difference: 0.14°C; SD: 0.51; p= 0.33); whereas the measurements of the PAC and the AT differed significantly (mean difference: 0.46°C; SD: 0.39; p< 0.001). Mean difference in PAC versus TAT analyses, between patients with vasopressor therapy (0.12°C; SD: 0.55), and without vasopressor therapy (0.19°C; SD: 0.48) was not statistically significant (p= 0.47). Conclusion: We can conclude that the temporal scanner has a relatively good reliability with an acceptable accuracy and variability in patients with normothermia. The results are comparable to those of the AT, but they do not seem to be sufficient to prove any substantial benefit compared to rectal, oral or bladder thermometry.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Comparing methods of measurement: why plotting difference against standard method is misleading
Bland, J. M., & Altman, D. G. (1995) · The Lancet · doi:10.1016/s0140-6736(95)91748-9 · Cited by 2,343
Cited by
Niven, D. J., Gaudet, J. E., Laupland, K. B., Mrklas, K. J., Roberts, D. J., & Stelfox, H. T. (2015) · Annals of Internal Medicine · doi:10.7326/m15-1150 · Cited by 240
Temperature measurements with a temporal scanner: systematic review and meta-analysis
Geijer, H., Udumyan, R., Lohse, G., & Nilsagård, Y. (2016) · BMJ Open · doi:10.1136/bmjopen-2015-009509 · Cited by 66
Non-contact infrared assessment of human body temperature: The journal Temperature toolbox
Foster, J., Lloyd, A. B., & Havenith, G. (2021) · Temperature · doi:10.1080/23328940.2021.1899546 · Cited by 46
“Agreement between NCIT of T forehead and a reference T core (tympanic) has been assessed in the static hospital environment [5,59–62].”
Metadata
- Authors
- Myny, D., Waele, J. D., Defloor, T., Blot, S., & Colardyn, F.
- Venue
- Scottish Medical Journal
- Publisher
- SAGE Publications
- Published
- 2005-02
- Volume
- 50
- Issue
- 1
- Pages
- 15-18
- ISSN
- 0036-9330 (ISSN-L)
- 2045-6441
- OpenAlex
- W2145996086
- Cited by
- 51
- References
- 1
- FWCI
- 0.88
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.