The Pulse Oximeter: Applications and Limitations—An Analysis of 2000 Incident Reports

Open access
Runciman, W. B., Webb, R. K., Barker, L., Currie, M. · 1993 · Anaesthesia and Intensive Care
Cited by 122

The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.

Abstract

The first 2000 incidents reported to the Australian Incident Monitoring Study were analysed with respect to the role of the pulse oximeter. Of these 184 (9%) were first detected by a pulse oximeter and there were a further 177 (9%) in which desaturation was recorded. Of the 1256 incidents which occurred in association with general anaesthesia 48% were “human detected” and 52% “monitor detected”. The pulse oximeter was ranked first and detected 27% of these monitor detected incidents; this figure would have been over 40% if an oximeter had always been used and its more informative modulated pulse tone relied upon instead of that of the “bleep” of the ECG. The pulse oximeter is the “front-line” monitor for endobronchial intubation, the fourth most common incident in association with general anaesthesia (it detected 87% of the 76 cases in which it was in use). It also played an invaluable role as a “back-up” monitor in 40 life-threatening situations in which “front-line” monitors (e.g. oxygen analyser, low pressure alarm, capnograph) were either not in use, were being used incorrectly or failed. Other situations detected, in order of frequency of detection, were: circuit disconnection, circuit leak, desaturation (severe shunt), oesophageal intubation, aspiration and/or regurgitation, pulmonary oedema, endotracheal tube obstruction, severe hypotension, failure of oxygen delivery, hypoxic gas mixture, hypoventilation, anaphylaxis, air embolism, bronchospasm, malignant hyperthermia, and tension pneumothorax. There were 15 reports of “failure”; four because the model in use had no modulated tone or alarm, four in which performance was in fact adequate, three were probe problems, two involved “over-reading”, one “under-reading” and in one new device the alarm failed. In the theoretical analysis of the 1256 general anaesthesia incidents it was concluded that pulse oximetry, used on its own, would have detected 82% of these incidents, had they been allowed to evolve (nearly 60% before any potential for organ damage). It is highly recommended that a suitable pulse oximeter be used on all patients from the time of arrival in the induction room until return of protective reflexes and demonstration of adequate saturation when breathing room air.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1177/0310057x9302100509

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 1 references

Cited by

See all 122

Metadata

Authors
Runciman, W. B., Webb, R. K., Barker, L., & Currie, M.
Venue
Anaesthesia and Intensive Care
Published
1993-10
Volume
21
Issue
5
Pages
543-550
ISSN
  • 0310-057X (ISSN-L)
  • 1448-0271
OpenAlex
W2407488597
Cited by
122
References
1
FWCI
8.57
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: The Pulse Oximeter: Applications and Limitations—An Analysis of 2000 Incident Reports. https://app.cochat.ai/papers/10.1177/0310057x9302100509
The paper
Runciman, W. B., Webb, R. K., Barker, L., & Currie, M. (1993). The Pulse Oximeter: Applications and Limitations—An Analysis of 2000 Incident Reports. Anaesthesia and Intensive Care. https://doi.org/10.1177/0310057x9302100509

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.