Potential errors in pulse oximetry: I. Pulse oximeter evaluation*
Open accessAbstract
There is no absolute reference for oxygen saturation, although multiwavelength in vitro oximeters are accepted as the 'gold standard'. Regardless of whether fractional or functional saturation is used by manufacturers to calibrate their oximeters, evaluation against fractional saturation is recommended since this is the clinically relevant variable. The use of standard notation and comparisons based on bias and precision is recommended. The accuracy of pulse oximetry is intrinsically limited by the use of only two wavelengths, and is dependent on the initial calibration population. The empirical algorithms used to convert the signal to its 'readout value' and the quality control of hardware may both be important sources of variability between oximeters. Change in blood temperature may introduce errors in pulse oximeter and in vitro oximeter saturation readings, but these will be clinically insignificant. Changes in blood pH should not decrease pulse oximetry accuracy.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
A review of 13 models of pulse oximeters suggested that a high percentage of carboxyhemoglobin or methaemoglobin are related to failures in saturation readings and can hide hypoxemia.
“A review of 13 models of pulse oximeters suggested that a high percentage of carboxyhemoglobin or methaemoglobin are related to failures in saturation readings and can hide hypoxemia [86].”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [86]
In 1991, commercial pulse oximeters had measurement error caused by ambient light.
“In 1991, it was found that commercial pulse oximeters had measurement error caused by ambient light [199].”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [199]
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965 · see citing works
Beckett, A. H., Rowland, M., & TURNER, P. (1965) · The Lancet · doi:10.1016/s0140-6736(65)91037-8 · Cited by 16,917 · see citing works
Tremper, K. K., & Barker, S. J. (1989) · Anesthesiology · doi:10.1097/00000542-198901000-00019 · Cited by 438 · see citing works
Effects of Methemoglobinemia on Pulse Oximetry and Mixed Venous Oximetry
Barker, S. J., Tremper, K. K., & Hyatt, J. (1989) · Anesthesiology · doi:10.1097/00000542-198901000-00021 · Cited by 384 · see citing works
Pulse oximetry: Analysis of theory, technology, and practice
Wukitsch, M. W., Petterson, M. T., Tobler, D. R., & Pologe, J. A. (1988) · Journal of Clinical Monitoring and Computing · doi:10.1007/bf01617328 · Cited by 305 · see citing works
Pulse Oximetry: Technical Aspects of Machine Design
Pologe, J. A. (1987) · International Anesthesiology Clinics · doi:10.1097/00004311-198702530-00009 · Cited by 110 · see citing works
A comparison of the performance of 20 pulse oximeters under conditions of poor perfusion
Clayton, D. G., Webb, R. K., Ralston, A., et al. (1991) · Anaesthesia · doi:10.1111/j.1365-2044.1991.tb09303.x · Cited by 109 · see citing works
Self-citationPulse Oximetry: Uses and Limitations
DeMeulenaere, S. (2007) · The Journal for Nurse Practitioners · doi:10.1016/j.nurpra.2007.02.021 · Cited by 109 · see citing works
Siggaard‐Andersen, O., Nørgaard‐Pedersen, B., & Rem, J. (1972) · Clinica Chimica Acta · doi:10.1016/0009-8981(72)90380-4 · Cited by 90 · see citing works
Definition and notation of hemoglobin oxygen saturation.
Zijlstra, W. G., & Oeseburg, B. (1989) · PubMed · Cited by 22 · see citing works
Reusing the Nellcor Pulse Oximeter Probe
Racys, V. A., & Nahrwold, M. L. (1987) · Anesthesiology · doi:10.1097/00000542-198705000-00039 · Cited by 15 · see citing works
A TEST OBJECT FOR ASSESSING PULSE OXIMETERS
Munley, A. J., Sik, M. J., & Shaw, A. G. (1989) · The Lancet · doi:10.1016/s0140-6736(89)92447-1 · Cited by 8 · see citing works
Reusing the Nellcor Pulse Probe
KAO, Y. J., & Badgwell, J. M. (1987) · Anesthesiology · doi:10.1097/00000542-198711000-00059 · Cited by 6 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L. (2021) · Biosensors · doi:10.3390/bios11040126 · Cited by 348 · see citing works
Meta-analysis of arterial oxygen saturation monitoring by pulse oximetry in adults
Jensen, L. A., Onyskiw, J. E., & Prasad, N. G. N. (1998) · Heart & Lung · doi:10.1016/s0147-9563(98)90086-3 · Cited by 243 · see citing works
A Review of Wearable Multi-Wavelength Photoplethysmography
Ray, D., Collins, T., Woolley, S., & Ponnapalli, P. (2023) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2021.3121476 · Cited by 155 · see citing works
Cabanas, A. M., Fuentes-Guajardo, M., Latorre, K., León, D., & Martín-Escudero, P. (2022) · Sensors · doi:10.3390/s22093402 · Cited by 124 · see citing works
- Authors
- RALSTON, A. C., WEBB, R. K., & RUNCIMAN, W. B.
- Affiliations
- Royal Adelaide Hospital (AU)
- Corresponding author
- RUNCIMAN, W. B.
- ORCID
- RUNCIMAN, W. B.
- Venue
- Anaesthesia
- Publisher
- Wiley
- Published
- 1991-03
- Volume
- 46
- Issue
- 3
- Pages
- 202-206
- ISSN
- 0003-2409 (ISSN-L)
- 1365-2044
- OpenAlex
- W2766731651
- Cited by
- 166
- References
- 13
- Citations by year
- 2026: 2 · 2025: 3 · 2024: 7 · 2023: 8 · 2022: 5 · 2021: 11 · 2020: 2 · 2019: 3 · 2018: 2 · 2017: 6 · 2016: 6 · 2015: 4 · 2014: 1 · 2013: 2 · 2012: 2
- FWCI
- 10.25
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Anaesthesia (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
- Keywords
- pulse oximetry, measurement error, oxygen saturation, bias and precision, empirical algorithms, blood temperature, blood pH
- MeSH terms
- Humans, Hydrogen-Ion Concentration, Oximetry, Oxygen, Oxyhemoglobins, Reference Standards, Signal Processing, Computer-Assisted, Statistics as Topic, Temperature
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review