Individuals Achieve More Accurate Results with Meters That Are Codeless and Employ Dynamic Electrochemistry
Open accessAbstract
Background: Studies have shown that controlling blood glucose can reduce the onset and progression of the long-term microvascular and neuropathic complications associated with the chronic course of diabetes mellitus. Improved glycemic control can be achieved by frequent testing combined with changes in medication, exercise, and diet. Technological advancements have enabled improvements in analytical accuracy of meters, and this paper explores two such parameters to which that accuracy can be attributed. Methods: Four blood glucose monitoring systems (with or without dynamic electrochemistry algorithms, codeless or requiring coding prior to testing) were evaluated and compared with respect to their accuracy. Results: Altogether, 108 blood glucose values were obtained for each system from 54 study participants and compared with the reference values. The analysis depicted in the International Organization for Standardization table format indicates that the devices with dynamic electrochemistry and the codeless feature had the highest proportion of acceptable results overall (System A, 101/103). Results were significant when compared at the 10% bias level with meters that were codeless and utilized static electrochemistry ( p = .017) or systems that had static electrochemistry but needed coding ( p = .008). Conclusions: Analytical performance of these blood glucose meters differed significantly depending on their technologic features. Meters that utilized dynamic electrochemistry and did not require coding were more accurate than meters that used static electrochemistry or required coding.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Clinical value of blood glucose monitoring systems is associated with measurement accuracy.
“Nowadays, a broad variety of BGMS is available on the market and it is understood that clinical value is associated with measurement accuracy.”
From System accuracy evaluation of 18 CE-marked current-generation blood glucose monitoring systems based on EN ISO 15197:2015 (2020) · cites this paper as [9]
Cites
Economic costs of diabetes in the U.S. in 2007
American Diabetes Association (2008) · Diabetes Care · doi:10.2337/dc08-9017 · Cited by 1,323 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Freckmann, G., Schmid, C., Baumstark, A., Pleus, S., Link, M., & Haug, C. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600510 · Cited by 237 · see citing works
Pleus, S., Baumstark, A., Jendrike, N., Mende, J., Link, M., Zschornack, E., Haug, C., & Freckmann, G. (2020) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2019-001067 · Cited by 63 · see citing works
- Authors
- Rao, A., Wiley, M., Iyengar, S., Nadeau, D., & Carnevale, J.
- Affiliations
- AgaMatrix Inc, Salem, New Hampshire
- HealthReach Diabetes, Endocrine and Nutrition Center, Hampton, New Hampshire
- Corresponding author
- Wiley, Meg
- ORCID
- Rao, Anoop
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2010-01
- Volume
- 4
- Issue
- 1
- Pages
- 145-150
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC2825635
- OpenAlex
- W2099919606
- Cited by
- 12
- References
- 1
- Citations by year
- 2020: 1 · 2019: 1 · 2018: 1 · 2016: 1 · 2015: 1 · 2013: 3 · 2012: 1
- FWCI
- 1.69
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrochemical sensors and biosensors
- Keywords
- blood glucose meters, analytical accuracy, glycemic control, diabetes mellitus, point-of-care glucose testing
- MeSH terms
- Electronic Data Processing, Blood Glucose, Calibration, Diabetes Mellitus, Electrochemistry, Female, Humans, Hypoglycemic Agents, Insulin, Kinetics, Male, Reference Values, Blood Glucose Self-Monitoring, Reproducibility of Results, Biosensing Techniques, Monitoring, Ambulatory
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review