System Accuracy Evaluation of 43 Blood Glucose Monitoring Systems for Self-Monitoring of Blood Glucose according to DIN EN ISO 15197
Open accessThe DOI matches its Crossref record; 36 of the 36 references it cites were found in a registry or an index.
Abstract
Background: The accuracy of systems for self-monitoring of blood glucose is important, as reliable measurement results are a prerequisite for therapeutic decisions. Methods: This system accuracy evaluation study was performed according to DIN EN ISO 15197:2003 for 43 Conformité Européenne (CE)-labeled blood glucose (BG) monitoring systems. Measurement results of each system were compared with results of the designated comparison method (manufacturer's measurement procedure): glucose oxidase method (YSI 2300 glucose analyzer) or hexokinase method (Hitachi 917/ cobas 501). Results: Complete assessment according to the International Organization for Standardization (ISO) standard was performed for 34 out of 43 systems, and 27 (79.4%) meet the requirements of the standard, i.e., ≥95% of their results showed at least the minimum acceptable accuracy. For 9 of the 43 systems, complete accuracy assessment was not performed due to an oxygen sensitivity (manufacturer's labeling). The bias (according to Bland and Altman) of all 43 evaluated systems ranged from −14.1% to +12.4%. Conclusions: From the 34 systems completely assessed, 7 systems did not fulfill the minimal accuracy requirements of the ISO standard. The CE mark apparently does not guarantee that all BG systems provide accuracy according to the standard. Because inaccurate systems bear the risk of false therapeutic decisions, regular and standardized evaluation of BG meters and test strips should be requested in order to ensure adherence to quality standards.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“The current requirement of traceability according to ISO 17511 as stated by ISO 15197, without specifying further requirements or performance goals, might be insufficient in light of the considerable systematic differences between measurement procedures observed in this and other studies.21,33”
From Impact of Two Different Reference Measurement Procedures on Apparent System Accuracy of 18 CE-Marked Current-Generation Blood Glucose Monitoring Systems (2022) · cites this paper as [33]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Yet, only a few studies investigated possible systematic differences between different comparison methods (GOD or HK) or measurement procedures up to now, whereas a bias of up to 8% between the different comparison measurement procedures has been reported.21,31-35”
From Impact of Two Different Reference Measurement Procedures on Apparent System Accuracy of 18 CE-Marked Current-Generation Blood Glucose Monitoring Systems (2022) · cites this paper as [33]
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“Nevertheless, several evaluation or surveillance studies within recent years repeatedly revealed that a non-negligible number of CE-marked BGMS, which do not fulfill the minimum accuracy criteria of ISO 15197, had been or currently still are on sale on the European market.12–14”
From System accuracy evaluation of 18 CE-marked current-generation blood glucose monitoring systems based on EN ISO 15197:2015 (2020) · cites this paper as [13]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667
International Organization for Standardization
(2011) · SpringerReference · doi:10.1007/springerreference_16847 · Cited by 12,534
International Organization for Standardization
(2015) · Encyclopedia of Astrobiology · doi:10.1007/978-3-662-44185-5_100574 · Cited by 6,716
(2007) · Endocrine Practice · doi:10.4158/ep.13.s1.1 · Cited by 699
Global guideline for type 2 diabetes: recommendations for standard, comprehensive, and minimal care
IDF Clinical Guidelines Task Force (2006) · Diabetic Medicine · doi:10.1111/j.1464-5491.2006.01918.x · Cited by 433
Polonsky, W. H., Fisher, L., Schikman, C. H., Hinnen, D. A., Parkin, C. G., Jelsovsky, Z., Petersen, B., Schweitzer, M., & Wagner, R. S. (2011) · Diabetes Care · doi:10.2337/dc10-1732 · Cited by 429
Factors affecting blood glucose monitoring: sources of errors in measurement
Ginsberg, B. H. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300438 · Cited by 348
Martin, S., Schneider, B., Heinemann, L., Lodwig, V., Kurth, H.-J., Kolb, H., Scherbaum, W. A., & for the ROSSO Study Group (2006) · Diabetologia · doi:10.1007/s00125-005-0083-5 · Cited by 284
Approved IFCC recommendation on reporting results for blood glucose (abbreviated)
D'Orazio, P. (2005) · Clinical Chemistry · doi:10.1373/clinchem.2005.051979 · Cited by 208
Research works printed but not linked: 3
- Bundesärztekammer, 2008, Dtsch Ärztebl, 105, 341
- Freckmann G, 2011, Diabetes Stoffwechsel Herz, 20, 235
- Essack Y, 2009, JEMDSA, 14, 102
Cited by
Universal mobile electrochemical detector designed for use in resource-limited applications
Nemiroski, A., Christodouleas, D. C., Hennek, J. W., Kumar, A. A., Maxwell, E. J., Fernández-Abedul, M. T., & Whitesides, G. M. (2014) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1405679111 · Cited by 285
“Although useful in a variety of settings, these methods— with the important exception of blood glucose meters (3, 4)—are generally limited to well-resourced laboratories run by skilled personnel.”
MethodWadwa, R. P., Laffel, L. M., Shah, V. N., & Garg, S. K. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0150 · Cited by 198
Hybrid nanostructures for SERS: materials development and chemical detection
Fateixa, S., Nogueira, H. I. S., & Trindade, T. (2015) · Physical Chemistry Chemical Physics · doi:10.1039/c5cp01032b · Cited by 197
Klonoff, D. C., Lias, C., Vigersky, R., Clarke, W., Parkes, J. L., Sacks, D. B., Kirkman, M. S., Kovatchev, B., Panel, T. E. G., Aurand, L., et al. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814539589 · Cited by 174
Directed evolution 2.0: improving and deciphering enzyme properties
Cheng, F., Zhu, L., & Schwaneberg, U. (2015) · Chemical Communications · doi:10.1039/c5cc01594d · Cited by 177
Metadata
- Authors
- Freckmann, G., Schmid, C., Baumstark, A., Pleus, S., Link, M., & Haug, C.
- Affiliations
- Universität Ulm (DE)
- Corresponding author
- Freckmann, Guido
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2012-09
- Volume
- 6
- Issue
- 5
- Pages
- 1060-1075
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC3570840
- OpenAlex
- W2153556745
- Cited by
- 237
- References
- 36
- Citations by year
- 2026: 1 · 2025: 4 · 2024: 9 · 2023: 13 · 2022: 8 · 2021: 11 · 2020: 16 · 2019: 19 · 2018: 30 · 2017: 28 · 2016: 23 · 2015: 22 · 2014: 31 · 2013: 20 · 2012: 1
- FWCI
- 18.99
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Physical Sciences › Physics and Astronomy › Instrumentation › Scientific Measurement and Uncertainty Evaluation
- Keywords
- self-monitoring of blood glucose, blood glucose meters, test strips, measurement accuracy, CE marking, glucose oxidase method, hexokinase method, bias, diabetes management
- MeSH terms
- Adolescent, Adult, Aged, Aged, 80 and over, Blood Glucose, Diabetes Mellitus, Equipment and Supplies, European Union, Female, Humans, Insulin Infusion Systems, Male, Middle Aged, Oxygen, Reference Standards, Blood Glucose Self-Monitoring, Reproducibility of Results, Device Approval, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: System Accuracy Evaluation of 43 Blood Glucose Monitoring Systems for Self-Monitoring of Blood Glucose according to DIN EN ISO 15197. https://app.cochat.ai/papers/10.1177/193229681200600510
- The paper
- Freckmann, G., Schmid, C., Baumstark, A., Pleus, S., Link, M., & Haug, C. (2012). System Accuracy Evaluation of 43 Blood Glucose Monitoring Systems for Self-Monitoring of Blood Glucose according to DIN EN ISO 15197. Journal of Diabetes Science and Technology. https://doi.org/10.1177/193229681200600510
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.